K971930 · Biotronik, Inc. · DTD · Aug 14, 1997 · Cardiovascular
Device Facts
Record ID
K971930
Device Name
BK-N SEALING CAP
Applicant
Biotronik, Inc.
Product Code
DTD · Cardiovascular
Decision Date
Aug 14, 1997
Decision
PT
Submission Type
Traditional
Regulation
21 CFR 870.3620
Device Class
Class 2
Indications for Use
BK-N Sealing Caps are intended for use as components of cardiac pacing systems. BK-N Sealing Caps provide a permanent barrier of silicone insulation between the electrical conductive elements of the implantable therapeutic device lead(s) or adapter(s) and the physiological BK-N Sealing Caps are designed to accommodate lead diameters of 4F, 5F and environment. 7F.
Device Story
BK-N Sealing Caps are single-component silicone accessories for cardiac pacing systems; designed for chronic implantation. Function as permanent insulation barrier between electrical conductive elements of leads/adapters and physiological environment. Accommodate 4F, 5F, and 7F lead diameters. Manufactured from Silopren LSR-4070 silicone. Used by clinicians during cardiac lead implantation procedures. Benefit: prevents electrical leakage/shorting in pacing systems. Subject to postmarket surveillance and device tracking requirements.
Clinical Evidence
Bench testing only. Testing included ease of use, temperature change resistance, lead extraction force, electrical insulation post-saline storage, ETO residue analysis, packaging/transportation durability, and sterilization validation. No clinical data presented.
Technological Characteristics
Material: Silopren LSR-4070 silicone. Form factor: Single-component sealing cap for 4F, 5F, 7F lead diameters. Sterilization: ETO. No software or electronic components.
Indications for Use
Indicated for use as components of cardiac pacing systems to provide a permanent silicone insulation barrier between electrical conductive elements of implantable leads or adapters and the physiological environment. Accommodates lead diameters of 4F, 5F, and 7F.
Regulatory Classification
Identification
A pacemaker lead adaptor is a device used to adapt a pacemaker lead so that it can be connected to a pacemaker pulse generator produced by a different manufacturer.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for the Submission of Research and Marketing Applications for Permanent Pacemaker Leads and for Pacemaker Lead Adaptor 510(k) Submissions.”
{0}------------------------------------------------
# AUG 1 4 1997
K971930
# 1.2 Safety and Effectiveness Summary
BIOTRONIK BK-N Sealing Caps are single-component silicone devices, designed to be safe and effective through design functionality, simplicity and material biocompatibility.
BK-N Sealing Caps are designed to be chronically implantable. The material of construction is Silopren LSR-4070, which has been extensively used by BIOTRONIK in cardiac pacing lead designs. FDA has reviewed and cleared this material in the following submissions:
BK Series Sealing Caps 510(k) #K970204 Cleared April 29, 1997
Dromos DR and SR Pulse Generators PMA #P950037 Approved October 11, 1996
TIR and TIJ Endocardial Leads 510(k) #K953044 Cleared September 27, 1996
Physios CTM 01 Pulse Generator and ELC XX-UP Epicardial Lead IDE #G960045 Approved August 22, 1996.
Silopren LSR-4070 has been thoroughly tested for biocompatibility, and is commonly used in many market-released Class III products, including pacemaker and defibrillator leads and their accessories. Acute and chronic biocompatibility tests have been performed, as well as long-term implantation studies.
BK-N Sealing Caps are tested for ease of use, ability to withstand temperature change, required extraction force of lead (following ligature application), electrical insulation following storage in warmed, aerated saline (in-vitro testing), and ETO residues. Packaging and transportation durability as well as sterilization validation have been performed to ensure the quality and sterility of the delivered product.
All production and biocompatibility test results were within specifications; therefore, when the currently proposed BK-N Sealing Caps are in use, the patient will be exposed to no risks in excess of those experienced by patients using comparable competitor sealing caps distributed in the United States, or older model BIOTRONIK sealing caps. Complications arising from usage may include body rejection phenomena, fibrosis, displacement, infection, and erosion. BIOTRONIK is not aware of any other adverse safety and effectiveness data on these accessories.
{1}------------------------------------------------
In accordance with the Safe Medical Device Act of 1990, a thorough literature search for adverse safety and effectiveness data on pacemaker lead sealing caps was performed. The search yielded no articles related to the safety of pacemaker lead sealing caps.
#### 1.3 Previous 510(k) Submittals
In a letter dated August 9, 1991 (#K911142), FDA provided notification it had determined the BK Series Sealing Caps were Class III devices substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976. The BK-N Sealing Caps addressed in that 510(k) document were constructed of HTV silicone.
### 1.4 Summary of Proposed Changes
#### 1.4.1 DEVICE CHANGES
BIOTRONIK is proposing the following changes to the BK-N Sealing Cap:
- MOC will be Silopren LSR-4070 .
- MOC change requires slight changes in sealing cap design .
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
AUG 1 4 1997
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856
Mr. Kenneth P. Jensen Regulatory Affairs BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, Oregon 97035-5369
Re : K971930 BK-N Sealing Cap Regulatory Class: III (three) Product Code: DTD Dated: May 23, 1997 Received: May 27, 1997
Dear Mr. Jensen:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (act). The general controls provisions of the act include requirements for registration, listing of devices, good manufacturing practices, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any
{3}------------------------------------------------
#### Page 2 - Mr. Kenneth P. Jensen
obligation you might have under section 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulation.
Under Section 522(a) of the act, manufacturers of certain types of devices identified by the Act or designated by FDA are required to ------------conduct postmarket surveillance studies. FDA has identified under Section 522(a) (1) (A) the device cleared for marketing by this letter as requiring postmarket surveillance.
Within thirty (30) days of first introduction or delivery for introduction of this device into interstate commerce you are required to submit to FDA certification of the date of introduction into interstate commerce, a detailed protocol which describes the postmarket surveillance study, and a detailed profile of the study's principal investigator that clearly establishes the qualifications and experience of the individual to conduct the proposed study. For your information, general guidance on preparing a protocol for a postmarket surveillance study is attached.
Submit five (5) copies to:
Center for Devices and Radiological Health Postmarket Surveillance Studies Document Center Room 3083 (HFZ-544) 1350 Piccard Drive Rockville, Maryland 20850
Within sixty (60) days of receipt of your protocol, FDA will either approve or disapprove it and notify you of the Agency's action in writing. You should not begin your postmarket surveillance study of this device until the protocol has been approved. Data generated under an unapproved protocol may not satisfy your obligation under section 522. Please note that you must continue to collect and report data needed to maintain compliance with Medical Device Reporting regulations (21 CFR 803).
Failure to certify accurately the date or initial introduction of your device into interstate commerce, to submit timely an acceptable protocol, or to undertake and complete and FDA approved postmarket surveillance study consistent with the protocol will be considered violations of section 522. In accordance with the Medical Device Amendments of 1992, failure of a manufacturer to meet its obligations under section 522 is a prohibited act under section 301(q) (1) (C) of the Act (21 U.S.C. 331 (q)(1)(C). Further, under section 502(t) (3) of the act (21 U.S.C. 352(t)(3)), a device is misbranded if there is a failure or refusal to comply with any requirement under section 522 of
{4}------------------------------------------------
Page 3 - Mr. Kenneth P. Jensen
the act. Violations of sections 301 or 502 may lead to regulatory actions including seizure of your product, injunction, prosecution, or civil money penalties.
If you have questions specifically concerning postmarket surveillance study requirements, contact the Postmarket Surveillance Studies Branch-- ---at (301) 594-0639. "
In addition, on August 16, 1993, the Final Rule for Device Tracking was published in the Federal Register, pages 43442-43455 (copy enclosed). Be advised that under Section 519(e) of the Act as amended by the Safe Medical Devices Act of 1990, FDA has identified the above device as a device which requires tracking. Because the device is subject to tracking, you are required to adopt a method of tracking that follows the devices through the distribution chain and then identifies and follows the patients who receive them. The specific requirement of the regulation are found in 21 CFR 821 as described in the August 16, 1993 Federal Register beginning on page 43447.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Walker.
Thomas JV Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
#### 1.1.1 INDICATIONS FOR USE
BK-N Sealing Caps are intended for use as components of cardiac pacing systems. BK-N Sealing Caps provide a permanent barrier of silicone insulation between the electrical conductive elements of the implantable therapeutic device lead(s) or adapter(s) and the physiological BK-N Sealing Caps are designed to accommodate lead diameters of 4F, 5F and environment. 7F.
> (Division Sign-Off) Division of Cardiovascular, and Neurological Device 510(k) Number .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.