MEDICAL Z, S.A. MEDIGEL Z SCAR MANAGEMENT GEL

K971916 · Medical Z, S.A. · MDA · Jul 29, 1997 · General, Plastic Surgery

Device Facts

Record IDK971916
Device NameMEDICAL Z, S.A. MEDIGEL Z SCAR MANAGEMENT GEL
ApplicantMedical Z, S.A.
Product CodeMDA · General, Plastic Surgery
Decision DateJul 29, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4025
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The Medigel Z products are indicated for temporary use in the management of hypertrophic scars and keloids resulting from wounds, trauma, or burns. The Medigel Z products are not wound dressings and are not intended to be used on open wounds. It is recommended that the products be used after the closing of a wound (8th to 10th day), or just after the first sign of hypertrophic scars.

Device Story

Medigel Z is a scar management device composed of tri-block polymer and mineral oil; available as sheets (3mm or 5mm thickness), elastic sleeves, finger-tubes, sternal straps, gloves, and adhesive-backed pads/patches (e.g., Mammopatch, Abdopatch). Device functions as a topical gel interface for hypertrophic scars and keloids; adhesive versions eliminate need for secondary compression bandages. Used in clinical or home settings to manage scar tissue post-wound closure. Provides physical barrier/contact to scar site; intended to assist in scar management process.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and material composition compared to legally marketed predicate devices.

Technological Characteristics

Tri-block polymer and mineral oil composition. Available in various form factors: sheets (3-5mm), elastic sleeves, finger-tubes, straps, and adhesive-backed pads. Some versions include fabric backing. Non-sterile, topical application.

Indications for Use

Indicated for temporary management of hypertrophic scars and keloids in patients with scars resulting from wounds, trauma, or burns. Not for use on open wounds; recommended for use after wound closure (8th-10th day) or upon first sign of hypertrophic scarring.

Regulatory Classification

Identification

Silicone sheeting is intended for use in the management of closed hyperproliferative (hypertrophic and keloid) scars.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected and form a single, abstract shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the image. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Medical Z. S.A. c/o Ms. Emalee G. Murphy McKenna & Cuneo, L.L.P ... . 1900 K Street, N.W. Washington, District of Columbia 20006-1108 JUL 2 9 1997 Re: K971916 > Trade Name: Medigel Z™ Scar Management Gel Regulatory Class: Unclassified Product Code: MDA Dated: May 23, 1997 Received: May 23, 1997 Dear Ms. Murphy: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the {1}------------------------------------------------ Page 2 - Ms. Emalee G. Murphy Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ K971916 ## Appendix C ## 510(k) Summary Medical Z, S.A. Medigel Z™ JUL 2 9 1997 I. General Information on Submitter: | Name: | Medical Z, S.A. | |------------------------|-----------------------------------------------------------| | Address: | BP 39 - 55, rue de l'Église<br>F-61110 Rémalard<br>France | | Telephone: | (33) 02 33 73 77 21 | | Fax: | (33) 02 33 73 78 88 | | Contact Person: | André Zagamé | | Date Summary Prepared: | May 23, 1997 | II. General Information on Device ្រីកុំប៉ុ | Name: | Medigel Z <sup>TM</sup> Scar Management Gel | |-------|---------------------------------------------| |-------|---------------------------------------------| | Classification Name: | Elastomer, Silicone, for Scar Management | |----------------------|------------------------------------------| |----------------------|------------------------------------------| III. Predicate Devices: Silipos® Tri Block Polymer Gel (K942695) Smith & Nephew Rolyan Cica-Care™ Silicone Gel Sheet (K935803) IV. Description of the Device: The Medigel Z is a combination of a tri-block polymer and mineral oil. Medigel Z sheets are available in two thicknesses, 3 mm and 5 mm. The 3 mm sheets are 4 inches by 4 inches. The 5 mm sheets are available in three sizes: 8 by 8 inches, 12 by 16 inches, and 16 by 20 inches. The 5 mm sheets are lined with a fabric backing. The Medigel Z is also available in four different sized elastic sleeves. The sleeves are 9.5 inches long and range from 12 to 55 inches in circumference. The Medigel Z is also available as an unlined finger-tube, a fabric lined sternal strap, and an elastic lined glove (open finger tips) and chin strap. {3}------------------------------------------------ The Medigel Z also is available with an adhesive which eliminates the need for a compression bandage in most cases. The adhesive Medigel Z is available with or without a fabric backing and comes in sheets or circular pads. The "Mammopatch Gel Z" and "Abdopatch Gel Z" are fabric backed versions of the adhesive Medigel Z that are shaped specifically for use on the breast and abdomen. ## V. Intended Use: The Medigel Z products are indicated for temporary use in the management of hypertrophic scars and keloids resulting from wounds, trauma, or burns. The Medigel Z products are not wound dressings and are not intended to be used on open wounds. It is recommended that the products be used after the closing of a wound (8th to 10th day), or just after the first sign of hypertrophic scars. VI. Technological Characteristics of Device Compared to Predicate Device: The Medigel Z™ shares the same technological characteristics, and is manufactured with the same materials as the Silipos® Tri Block Polymer Gel, with the exception of the use of an adhesive to attach the device to the patient's skin. The Cica-Care™ gel, like the Medigel Z, uses an adhesive to attach the gel the body. {4}------------------------------------------------ K971916 510(k) Number (if known): Device Name: Medigel Z Scar Management Gel Indications For Use: The Medigel Z products are indicated for temporary use in the management of hypertrophic scars and Keloids resulting from . wounds, trauma, or burns. The Medigel Z products are not wound dressings and are not intended to be used on open wounds. It is recommended that the products be used after the closing of a wound (8th to 10th day), or just after the first sign of hypertrophic scars. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Divi Off) enerai Restorative Devices, 6971916 51VUItiber Image /page/4/Picture/6 description: The image shows the words "Prescription Use" in bold font. Below that, the text "(Per 21 CFR 801.109)" is shown in a smaller font. There is a line drawn through the words "Prescription Use". OR Over-The-Counter Use
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...