← Product Code [FJL](/productcode/FJL) · K971881

# COMEG RESECTOSCOPE AND LASER-RESECTOSCOPE WORKING ELEMENTS/URETHROTOMES/ELECTRODES/ALBARRAB DEFLECTORS (K971881)

_Comeg Endoscopy · FJL · Aug 12, 1997 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K971881

## Device Facts

- **Applicant:** Comeg Endoscopy
- **Product Code:** [FJL](/productcode/FJL.md)
- **Decision Date:** Aug 12, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Resectoscope and Laser Resectoscope Working Elements are intended to be used in urological procedures to endoscopically remove, cut, coagulate, and/or transect tissue in the bladder, prostate, and/or urethra. The surgeon performs the examination through the urethra. The working elements are devices that house and control the various electrodes and lasers used to remove, cut, coagulate, and transect tissue. The doctor controls the back and forth movement of the electrode using finger controls. The working elements also house a cystoscope for visualization. The high frequency cable transmits electrical current. Albarran deflectors are intended for use in urological procedures to control the flexible instrumentation that is passed through the COMEG cystoscope sheaths. Urethrotomes are intended to be used in urological procedures to endoscopically cut and or transect tissue(strictures) in the urethra. The surgeon performs this function through the urethra and will visually (cystoscope) find the location of the stricture in the urethra. Once the stricture is located a urethrotome knife will be extended an the entire urethrotome will be extracted cutting the affected area. Electrodes are intended to be used endoscopically im conjunction with COMEG resectoscopes, urethrotomes, or cystoscope sheaths. The purpose of the electrodes is for cutting, tissue removal, tissue vaporization, coagulation, and/or transection of tissue. Electrodes are connected to either a resectoscope working element or a high frequency cable. The electrodes are reusable.

## Device Story

System comprises resectoscope/laser-resectoscope working elements, Albarran deflectors, urethrotomes, and electrodes; used in urological procedures. Working elements house cystoscope for visualization and control electrode/laser movement via finger controls; high-frequency cable delivers electrical current for tissue interaction. Surgeon performs procedure endoscopically through urethra. Devices facilitate cutting, coagulation, vaporization, and transection of bladder, prostate, or urethral tissue. Albarran deflectors guide flexible instrumentation. Urethrotomes enable visual identification and mechanical cutting of urethral strictures. Electrodes are reusable and connect to working elements or cables. System provides surgeon with direct visual control and mechanical/electrosurgical tissue management, benefiting patients by enabling minimally invasive urological interventions.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Materials: stainless steel (chrome plated), brass (chrome plated), plastic, silicone, teflon, ultem, tungsten, silver. Components: resectoscope working elements, Albarran deflectors, urethrotomes, electrodes, high-frequency cables. Principle: endoscopic electrosurgical/mechanical tissue manipulation. Connectivity: manual/mechanical connection (Quick-Connection). Sterilization: reusable devices.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Karl Storz (not specified)
- Circon ACMI (not specified)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

/ 97/881 AUG 12 1997

E n d o s c o p y

pg 1 of 2

## SUMMARY OF SAFETY AND EFFECTIVENESS FOR RESECTOSCOPE AND LASER-RESECTOSCOPE WORKING ELEMENTS. ALBARRAN DEFLECTORS, URETHROTOMES, AND ELECTRODES

\$807.92 (a)(1)

Contact Person: Peter Duffy Vice President

May 16, 1997 Date of Summary Preparation:

\$807.92 (a)(2)

Trade Name: COMEG Endoscopy Resectoscope and Laser-Resectoscope Albarran Deflectors. Urethrotomes. Working Elements. and Electrodes Common Name: Endoscopes and accessories Classification Name: Endoscope and accessories (21 CFR \$876.1500) Urethrotome (21 CFR §876.4770) Endoscope ESU unit and accessories (21 CFR \$876.4300)

§807.92 (a)(3)

Legally Marketed Substantially Equivalent Device: Karl Storz and Circon ACMI

\$807.92 (a)(4)

The COMEG devices that we intend to market Description of Device: the following endoscopic electrosurgical instruments and accessories: include resectoscope and laser-resectoscope working elements; Albarran deflectors; urethrotomes (urethrotome bridges, obturators, and knives); high frequency cable; and, electrodes. A "Quick-Connection" feature is used to attach the device to the sheath or working element.

Endoscopic electrosurgical instruments and accessories are described in endoscope and accessories 21 CFR 8876.1500, urethrotomes are described in 21 CFR §876.4770, and endoscope ESU unit and accessories are described in 21 CFR §876.4300.

The Resectoscope and Laser Resectoscope Working Elements are intended to be used in urological procedures to endoscopically remove, cut, coagulate, and/or transect tissue in the bladder, prostate, and/or urethra. The surgeon performs the examination through the urethra. The working elements are devices that house and control the various electrodes and lasers used to remove, cut, coagulate, and transect tissue. The doctor controls the

{1}------------------------------------------------

K971881 Pg 2 of 2

back and forth movement of the electrode using finger controls. The working elements also house a cystoscope for visualization. The high frequency cable transmits electrical current.

These devices are composed of stainless steel chrome plated, plastic, and brass chrome plated. The high frequency cable is composed of silicone.

Albarran deflectors are intended for use in urological procedures to control the flexible instrumentation that is passed through the COMEG cystoscope sheaths.

These devices are composed of stainless steel chrome plated and brass chrome plated.

The intended use for the urethrotomes is in urological procedures to endoscopically cut and/or transect tissue (strictures) in the urethra. The surgeon performs this function through the urethra and will visually (cystoscope) find the location of the stricture in the urethra. Once the stricture is located a urethrotome will be extracted, cutting the affected area.

These devices are composed of stainless steel chrome plated, plastic, and brass chrome plated. The complete working unit is comprised of each of the models below plus a cystoscope.

The electrodes are intended to be used endoscopically in conjunction with COMEG resectoscopes, urethrotomes, or cystoscope sheaths. The purpose of the electrodes is for cutting, tissue removal, tissue vaporization, coagulation, and/or transection of tissue. Electrodes are connected to either a resectoscope working element or a high frequency cable. The electrodes are reusable.

These electrodes are composed of stainless steel, teflon, silicone, ultem (black plastic), tungsten, and silver.

8807.92 (a)(5) Intended Use:

See Device Description above.

\$807.92 (a)(6)

Endosco

Comparison of Technical Characteristics:

The subject devices are similar to devices marketed by Karl Storz and Circon ACMI. The predicate device for the Quick-Connection feature is Circon ACMI. The intended uses for the subject devices are the same for the subject devices and the competitors' products. The materials used to fabricate both the COMEG and the Karl Storz devices and the operational principles and mode of action are similar as well.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its head and wings.

Food and Drug
9200 Corporate
Rockville MD 2

AUG 1 2 1997

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Peter Duffy Vice President COMEG Endoscopy 13790 E. Rice Place Aurora, Colorado 80015

Re: K971881 COMEG Resectoscope and Laser-Resectoscope Working Elements, Urethrotomes (bridges, obturator, and knives), and Electrodes Dated: May 16, 1997 Received: May 20, 1997 Regulatory Class: II 21 CFR §876.1500, 876.4770, and 876.4300 Product Code: FJL, FDC, EZO, and FAS

Dear Mr. Duffy

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitr diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours.

H.D.Liau Yin

Lillian Yin, Ph.D. Director. Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "COMEG" in a bold, sans-serif font, with each letter in white against a black background. Below the word "COMEG", the word "Endoscopy" is written in a smaller, serif font. The word "Endoscopy" is also in white against a black background. The image appears to be a logo or branding for a medical company specializing in endoscopy.

## Indications for Use

510(k) Number (if known) Device Name: Endoscopic electrosurgical instruments and accessories

Indications for Use:

Resectoscope and Laser Resectoscope Working Elements are intended to be used in urological procedures to endoscopically remove, cut, coagulate, and/or transect tissue in the bladder, prostate, and/or urethra. The surgeon performs the examination through the urethra. The working elements are devices that house and control the various electrodes and lasers used to remove, cut, coagulate, and transect tissue. The doctor controls the back and forth movement of the electrode using finger controls. The working elements also house a cystoscope for visualization. The high frequency cable transmits electrical current.

Albarran deflectors are intended for use in urological procedures to control the flexible instrumentation that is passed through the COMEG cystoscope sheaths.

Urethrotomes are intended to be used in urological procedures to endoscopically cut and or transect tissue(strictures) in the urethra. The surgeon performs this function through the urethra and will visually (cystoscope) find the location of the stricture in the urethra. Once the stricture is located a urethrotome knife will be extended an the entire urethrotome will be extracted cutting the affected area.

Electrodes are intended to be used endoscopically im conjunction with COMEG resectoscopes, urethrotomes, or cystoscope sheaths. The purpose of the electrodes is for cutting, tissue removal, tissue vaporization, coagulation, and/or transection of tissue. Electrodes are connected to either a resectoscope working element or a high frequency cable. The electrodes are reusable.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use                          | K        | OR |
|-------------------------------------------|----------|----|
| (Per 21 CFR 801.109)                      |          |    |
| (Division Sign-Off)                       | Dmed VAP |    |
| Division of Reproductive, Abdominal, ENT, |          |    |
| and Radiological Devices                  |          |    |
| 510(k) Number                             | K971881  |    |

-5-Over-the-Counter Use

---

**Source:** [https://fda-staging.innolitics.com/device/K971881](https://fda-staging.innolitics.com/device/K971881)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
