MEDISYSTEMS DIALYSIS PRIMING SETS
Device Facts
| Record ID | K971860 |
|---|---|
| Device Name | MEDISYSTEMS DIALYSIS PRIMING SETS |
| Applicant | Medisystems Corp. |
| Product Code | KOC · Gastroenterology, Urology |
| Decision Date | Aug 18, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5820 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The intended use of the Medisystems Dialysis Priming Sets is to provide a means to deliver fluids and medicaments from a collapsible container or vented bottle into a patient's vascular system during dialysis procedures.
Device Story
Medisystems Dialysis Priming Set; fluid delivery system for dialysis. Components: spike connector for solution containers, flexible drip chamber, adjustable flow-regulating clamp, flexible delivery tubing, two-port connector. Connects to extracorporeal tubing or IV needle sets via luer connectors. Used in clinical dialysis settings to prime or rinse-back vascular access. Manual operation by healthcare providers. Benefits: provides controlled fluid/medicament administration into patient vascular system.
Clinical Evidence
No clinical data. Safety and effectiveness supported by biocompatibility and performance testing, including sterility, pyrogenicity, and physical testing per device master records.
Technological Characteristics
Components: spike connector, flexible drip chamber, adjustable clamp, flexible tubing, luer connectors. Materials of construction differ from predicate but validated via biocompatibility and performance testing. Manual flow control. Non-powered, mechanical device.
Indications for Use
Indicated for delivery of fluids and medicaments from a collapsible container or vented bottle into a patient's vascular system during dialysis procedures.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.