K971853 · American Medical Devices, Inc. · HJK · Aug 6, 1997 · Ophthalmic
Device Facts
Record ID
K971853
Device Name
ENDOVIEW SPPHIRE LENS SET
Applicant
American Medical Devices, Inc.
Product Code
HJK · Ophthalmic
Decision Date
Aug 6, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1385
Device Class
Class 2
Indications for Use
For magnification during vitreoretinal surgery.
Device Story
Endoview™ Sapphire Lens Set provides visualization of the posterior segment of the eye during vitreoretinal surgery. The system comprises a sterilization container, a lens ring (with or without infusion port), and various sapphire lenses (Plano Concave, Symmetric Concave-Concave, Asymmetric Concave-Concave, Prism 15°/30° Concave, and Prism Concave-Concave). During surgery, the surgeon sutures the ring onto the sclera to stabilize the lens on the cornea. The infusion ring option allows for continuous irrigation, eliminating manual application. The surgeon selects and swaps lenses during the procedure to achieve required visualization angles. Sapphire material is utilized for its hardness, chemical resistance, and wide transmission band, offering superior durability compared to quartz or plastic alternatives. The device is intended for use by ophthalmic surgeons in a clinical or surgical setting.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing and material safety comparisons. Sapphire material has a history of use in ophthalmic devices since the mid-1980s.
Technological Characteristics
Lenses: Single-crystal sapphire. Lens Ring: Aluminum. Sterilization Container: High-temperature plastic. Connectivity: None (manual). Sterilization: ETO or steam (AAMI/USP standards). Form factor: Set of interchangeable lenses and rings for scleral mounting.
Indications for Use
Indicated for magnification during vitreoretinal surgery for patients undergoing such procedures.
Regulatory Classification
Identification
A polymethylmethacrylate (PMMA) diagnostic contact lens is a device that is a curved shell of PMMA intended to be applied for a short period of time directly on the globe or cornea of the eye for diagnosis or therapy of intraocular abnormalities.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for American Medical Devices, Inc. The logo consists of the letters "A MDI" in a bold, sans-serif font. The "A" is in a white box, while the "MDI" is in a black box. To the right of the logo, the words "AMERICAN MEDICAL DEVICES, Inc." are stacked vertically.
AUG - 6 1997
K971853
May 16, 1997
# Premarket Notification [510(k)] Summary
Submitter: American Medical Devices, Inc 1100 Northside Drive Atlanta, GA 30318 Phone: (404) 815-5233 (404) 815-5235 Fax:
Official Correspondent: Frank J. Tighe
Trade Name: Endoview™ Sapphire Lens Set
Common Name: Surgical Contact Lens Set
Registration Number: We have registered but have not received our application back as of this date.
Class: Class II
Class Name: Unknown
Panel: Ophthalmic
Product Code: 86HDQ
#### Device Description
The Endoview™ Sapphire Contact Lens Set consists of one sterilization container, one lens ring (either infusion or non-infusion), and one each of the following lenses: Plano Concave, Symmetric Concave-Concave, Asymmetric Concave, Prism 15° - Concave, Prism 30° Concave, and a Prism Concave-Concave Lens. The lenses, rings and sterilization container may be sold separately or as a set.
During vitreoretinal surgery, the surgeon requires the aid of a contact lens in order to visualize the posterior segment of the eye. First, the ring is sewn onto the sclera to hold the lens in place on the cornea.
Page 1 Summary
| Toll-Free | Direct | Facsimile | Internet |
|--------------|--------------|--------------|-----------------------------------|
| 800-793-1473 | 404-815-5233 | 404-815-5235 | ftighe@americanmedicaldevices.com |
{1}------------------------------------------------
The customer has the option of choosing a ring with an infusion port, which eliminates the need for the surgeon to constantly place irrigant on the corneal. This type of ring device is currently available from Ocular Instruments (510(k) 902496. Once the ring is in place, the surgeon places one of the lenses in the ring in order to visualize the posterior segment of the eye. The surgeon changes the lenses during the case depending on where visualization is required. Most lenses on the market have been manufactured from quartz and high refractive index plastic because of their low cost and good refractive properties. However, these materials can be easily scratched, chipped and damaged during normal cleaning, handling and sterilization. Single Crystal Sapphire, on the other hand, is an ideal material for ophthalmic surgery because of its wide transmission band, extreme hardness and excellent chemical resistance.
| Substantial Equivalence Comparison | | | | |
|------------------------------------------------------|------|------------|-----------------------|---|
| American Medical<br>Devices, Inc. | DORC | Grieshaber | Ocular<br>Instruments | |
| Flat Lens for<br>Central viewing | X | X | X | X |
| 15 degree lens for mid-<br>Peripheral viewing | X | X | X | X |
| 30 degree lens for wide<br>field viewing | X | X | X | X |
| Bi-concave lens<br>For air filled phakik<br>Patients | X | X | X | X |
| Shipped Non Sterile | X | X | X | X |
| Sterilize with either ETO<br>Or steam | X | X | X | X |
| Sterilization Container | X | X | X | X |
| Infusion Lens Ring | X | X | X | X |
| Lens Ring** | X | X | X | X |
Statement of indications for use. For magnification during vitreoretinal surgery.
** All lens sets have lens rings. American Medical Devices, Inc and Ocular Instruments offer a lens ring with an infusion port.
Page 2 Summary
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for American Medical Devices, Inc. The logo is divided into two sections. The left section has a white background with the letter "A" in black, and a black background with the letters "MDI" in white. The right section has the words "AMERICAN MEDICAL DEVICES, Inc." stacked on top of each other in black.
#### Sterility
The Device is shipped non-sterile to the user. The user may either ETO or steam sterilize the lenses according AAMI or USP Standards. ETO and Steam Sterilization of Sapphire in Ophthalmic Devices has been performed since the mid 1980's A user instruction sheet is included with each lens set (See Page 4 of this summary), which instructs the user to clean the lenses immediately after used with warm soapy water. In addition, the instructions clearly state to only sterilize thoroughly cleaned instruments and to Steam or ETO sterilize the lenses prior to use according to AAMI or USP standards.
#### Packaging
The device is shipped in a high temperature plastic sterilization container, which contains a custom-molded cavity to secure the lenses and ring during sterilization.
#### Materials Used in Manufacture
Lenses - The lenses are manufactured from sapphire. Sapphire has been approved for use in ophthalmic devices since the mid-1980s with product code 86HNN and with the following 510(k)s: K853129 Rudolph Beaver Inc. Beaver Multipurpose Sapphire Knife. K860852 Med-Tech Development Corp. Sapphire Blade Surgical Knife. K861744 Keeler Instruments. Inc., Sapphire Knife Blade. These lenses are non-invasive and only contact the cornea during the surgical procedure whereas the sapphire knives are invasive and are used to make an entry into the cornea and actually enter the anterior chamber of the eve.
Sterilzation Container: Manufactured from high temperature plastic. The Ocular Instruments and DORC sterilization containers are also manufactured from high temperature plastic.
Lens Ring: Aluminum. Ocular Instruments, Grieshaber, and DORC all use aluminum for their lens rings
| | | | Page 3 Summary |
|--------------|--------------|--------------|-----------------------------------|
| Toll-Free | Direct | Facsimile | Internet |
| 800-793-1473 | 404-815-5233 | 404-815-5235 | ftighe@americanmedicaldevices.com |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for American Medical Devices, Inc. The logo consists of two parts: a square divided into two halves, with the left half being white and the right half being black, and the letters "AMDI" in bold black font on the square. To the right of the square is the text "AMERICAN MEDICAL DEVICES, Inc." in a bold, sans-serif font.
## ENDOVIEW™ SAPPHIRE CONTACT LENS SET
# USER, CLEANING AND STERILIZATION INSTRUCTIONS
## THIS DEVICE IS NON STERILE
#### Device Description
The Endoview™ Sapphire Contact Lens Set consists of one sterilization container, one lens ring (either infusion or non-infusion), and one each of the following lenses: Plano Concave, Symmetric Concave-Concave, Asymmetric Concave-Concave, Prism 15° - Concave, Prism 30° Concave, and a Prism Concave-Concave Lens. The lenses, rings and sterilization container may be sold separately or as a set.
During vitreoretinal surgery, the surgeon requires the aid of a contact lens in order to visualize the posterior segment of the eye. First, the ring is sewn onto the sclera to hold the lens in place on the cornea. The customer has the option of choosing a ring with an infusion port, which eliminates the need for the surgeon to constantly place irrigant on the corneal. Once the ring is in place, the surgeon places one of the lenses in the ring in order to visualize the posterior segment of the eye. The surgeon changes the lenses during the case depending on where visualization is required. Most lenses on the market have been manufactured from quartz and high refractive index plastic because of their low cost and good refractive properties. However, these materials can be easily scratched, chipped and damaged during normal cleaning, handling and sterilization. Single Crystal Sapphire, on the other hand, is an ideal material for ophthalmic surgery because of its wide transmission band, extreme hardness and excellent chemical resistance.
Statement of indications for use. For magnification during vitreoretinal surgery.
#### CLEANING
Clean the lenses immediately after use with warm soapy water.
#### STERILIZATION THIS DEVICE IS NON STERILE.
Only sterilize the product after it has been thoroughly cleaned. ETO or steam sterilize the product prior to use according AAMI or USP Standards.
| | | Summary Page 4 | |
|--------------|--------------|----------------|-----------------------------------|
| Toll-Free | Direct | Facsimile | Internet |
| 800-793-1473 | 404-815-5233 | 404-815-5235 | ftighe@americanmedicaldevices.com |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for American Medical Devices, Inc. On the left side of the logo, the letter "A" is in white on a black square, and the letters "MDI" are in black on a white square. To the right of the square logo, the words "AMERICAN MEDICAL DEVICES, Inc." are stacked on top of each other.
### ENDOVIEW™ SAPPHIRE LENS SET CONTENTS
#### Trade or proprietary or model name
ﺴ
#### Model Number
| 1. | EndoviewTM Sapphire Lens Set (Consisting of 1 each of the following)** | 600.00 |
|-----|------------------------------------------------------------------------|--------|
| 2. | EndoviewTM Flat Sapphire Lens (Machemer) | 600.01 |
| 3. | EndoviewTM 15° Prism Sapphire Lens (Tolentino) | 600.02 |
| 4. | EndoviewTM 30° Prism Sapphire Lens (Tolentino) | 600.03 |
| 5. | EndoviewTM Biconcave Sapphire Lens (Landers) | 600.04 |
| 6. | EndoviewTM Asymmetric Biconcave Sapphire Lens (de Juan) | 600.05 |
| 7. | EndoviewTM 30° Biconcave Sapphire Lens (Woldoff,Tano) | 600.06 |
| 8. | EndoviewTM Sapphire Lens Set Sterilization Container | 600.07 |
| 9. | EndoviewTM Infusion Lens Ring** | 600.08 |
| 10. | EndoviewTM Lens Ring** | 600.09 |
**The customer has the option of choosing the lens ring with or without infusion.
| Summary Page 5 | | | |
|----------------|--------------|--------------|-----------------------------------|
| Toll-Free | Direct | Facsimile | Internet |
| 800-793-1473 | 404-815-5233 | 404-815-5235 | ftighe@americanmedicaldevices.com |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure with three flowing lines representing the body, head, and arm. The figure is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper portion of the circle. The logo is rendered in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 6 1997
American Medical Devices, Inc. c/o Mr. Frank |. Tighe 1100 Northside Drive Atlanta, GA 30318
Re: K971853 Trade Name: Endoview™ Sapphire Lens Set Regulatory Class: II Product Code: 86 HJK Dated: May 16, 1997 Received: May 20, 1997
Dear Mr. Tighe:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS)inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Flectronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your-device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A Rosyl Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo for American Medical Devices, Inc. The logo is divided into three sections. The first section is a white square with a large black letter A. The second section is a black square with the letters MDI in white. The third section has the words AMERICAN MEDICAL DEVICES, Inc. in black.
(510(k) Number (if known):
Device Name: Endoview™ Sapphire Lens Set
Indications for Use:
For magnification during vitreoretinal surgery.
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel W. C. Brown, Ph.D.
Prescription Use ズ (Per 21 CFR 801.109) (Division Sign-Off) Division Ophthalmic Devices 510(k) Number
Over-The-Counter Use
(Optional Format 1-2-96)
| 1 011-1 rec<br>STATIS AND AND PERSONAL PROPERTY OF CHARACT PUBLICATION OF CHEARTH | |
|-----------------------------------------------------------------------------------|--|
| 800-793-1473 | |
Direct 404-815-5233
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.