LASERSCOPE VELA ERBIUM:YAG LASER SYSTEM AND ACCESSORIES
K971843 · Laserscope · GEX · Jul 23, 1997 · General, Plastic Surgery
Device Facts
Record ID
K971843
Device Name
LASERSCOPE VELA ERBIUM:YAG LASER SYSTEM AND ACCESSORIES
Applicant
Laserscope
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 23, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Laserscope VELA Erbium: YAG Laser System and Accessories are intended for the surgical incision. vaporization and congulation of soft tissue. All soft tissue is included such as skin, subcutaneous tissue, striated and smooth tissue, muscle, cartilage menus membranc, lymph vessels and nodes, organs and glands. DermatologyPlastic Surgery: Indications include, cpidcrmal nevi, teleagiestasia, soider veins, scinic chailitis, keloids, verrucee, skin togs, keratoses, sear revision, debulking banign tumors, decubitis ulcers and skin resurfacing. General Surgery: The Erbium: YAG lass is intended for the surgical incision, vaportuation and coagulation of soft tissue during any general surgery application where skin incision, tissue of external lumors and lesions, complete or partial resection of internal organs, tissue abletion and/or vessel coagulation may be indicated. Genitourinary: Indications include lesions of the external genitalia, welcon and anus, penis, scrown and urchire (includes condyloma acuminata, gian perined condyloma und verrucous carcinoma), valvar lesions, polyps and familial polyps of the colon. Gynecology: Indications include cervical intraepithelial neoplusia (CDV), herpes simplex, endometrial adhesions, cysts and condyloms. ENT: Indications include car, nose and throat lessions, polyps, cysts, hyperkeratosis; excision of carcinogenic lissue, oral leukoplakia. Oral/Maxillofacial: Indications include benign oral tunnors, oral and glossel lessions and gingivectory. Ophthalmology: Indications include soft tissue surrounding the eye and orbit and unterior capsulotomy. Podiatry: Indications include warts, plantar verrucae, large mossic verrucae and matrixectory .
Device Story
Laserscope Erbium:YAG Laser System consists of a moveable console with power supplies, aiming laser, treatment laser, and cooling mechanism. User controls parameters via softouch keypad and LCD display. Aiming beam is coincident with Erbium:YAG beam, delivered through an articulating arm to a handpiece. Focusing adapters (1-8 mm spot sizes) attach to handpiece. Device used by clinicians for cutting, vaporizing, and coagulating soft tissue. Benefits include precise tissue ablation and coagulation across various surgical specialties.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use compared to the predicate device.
Technological Characteristics
Erbium:YAG laser system; 2.94 micron wavelength; articulating arm delivery; 1-8 mm spot size adapters; cooling mechanism for heat dissipation; conforms to 21 CFR 1040.10 and 1040.11.
Indications for Use
Indicated for surgical incision, vaporization, and coagulation of soft tissue (skin, subcutaneous, muscle, cartilage, mucous membrane, lymph, organs, glands) across dermatology, plastic surgery, general surgery, genitourinary, gynecology, ENT, oral/maxillofacial, ophthalmology, and podiatry specialties. Prescription use only.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Continuum Biomedical CB Erbium/2.94 Er:YAG Laser System
Submission Summary (Full Text)
{0}------------------------------------------------
# K971843
JUL 23 1997
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS LASERSCOPE ERBIUM:YAG LASER SYSTEM AND ACCESSORIES
## REGULATORY AUTHORITY:
Safe Medical Devices Act of 1990, 21 CFR 807.92
#### COMPANY NAME/CONTACT:
Lisa McGrath Laserscope 3052 Orchard Drive San Jose, CA 95134-2011 Phone: 408 943-0636 408 943-1454 FAX:
### DEVICE TRADE NAME:
Laserscope Erbium:YAG Laser System and Accessories
#### DEVICE COMMON NAME:
Erbium:YAG Laser System
## DEVICE DESCRIPTION:
10
The Laserscope Erbium:YAG Laser System and Accessories consists of a moveable console containing power supplies, aiming and treatment lasers on a solid optical deck and a cooling mechanism to dissipate the heat generated by the system. A softouch keypad control panel with LCD displays enables the user to control the laser system operating parameters. The aiming beam is coincident with the Erbium:YAG beam and passes through an articulating arm. The aiming beam emerges at the distal end of a handpiece. Focusing handpiece adapters, which attach to the handpiece, are available in spot sizes of 1 - 8 mm.
{1}------------------------------------------------
## SUMMARY OF SAFETY AND EFFECTIVENESS, PAGE 2
#### DEVICE CLASSIFICATION:
Erbium:YAG Laser Systems and Accessories have not been specifically classified; however Nd:YAG, CO2, and Argon Surgical Lasers have been classified as Class II medical devices by the OB/GYN, General, Plastic Surgery and ENT Device Advisory Panels.
#### PERFORMANCE STANDARDS:
All Laserscope Erbium:YAG Surgical Laser Systems and Accessories conform with federal regulations and the performance standards 21 CFR 1040.10 and 1040.11 for medical laser systems.
#### INDICATION FOR USE STATEMENT:
The Laserscope Erbium: YAG Laser System and Accessories are indicated for use in procedures involving cutting (incision/excision), vaporizing and coagulating soft tissue.
All soft tissue is included such as skin, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs and glands.
Clinical Applications:
Dermatology/Plastic Surgery General Surgery (Soft Tissue) Genitourinary Gynecology ENT Soft Tissue Procedures Oral and Maxillofacial Surgery Ophthalmology Podiatry
{2}------------------------------------------------
## SUMMARY OF SAFETY AND EFFECTIVENESS, PAGE 3
## COMPARISON WITH PREDICATE DEVICE:
The Laserscope Erbium:YAG Laser System and Accessories is substantially equivalent to the Continuum Biomedical CB Erbium/2.94 Er:YAG Laser System.
The risks and benefits for the Laserscope Erbium:YAG Laser System and Accessories are comparable to the predicate device when used for similar clinical applications.
Since the Laserscope Erbium:YAG Laser System and Accessories are substantially equivalent with respect to indications for use, materials, method of operation and physical construction to the predicate device, we believe it clearly meets the requirements for substantial equivalence according to Section 510(k) guidelines. Safety and effectiveness are reasonably assured, therefore justifying 510(k) clearance for commercial sale.
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Lisa McGrath Sr. Regulatory Affairs Specialist Laserscope . . .. .. . . . . . . . . . . 3052 Orchard Drive San Jose, California 95134-2011
JUL 23 1997
Re: K971843
> Trade Name: Laserscope VELA Erbium: YAG Laser System and Accessories Regulatory Class: II Product Code: GEX Dated: May 12, 1997 Received: May 14, 1997
Dear Ms. McGrath:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
fo collafe
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT page I
843.
510(k) Number:
971843
Device Name: . . . . . . _ .. .
Laserscope Erbium: YAG Laser. System and Accessorics . . . . . . . . . . . . .
Indications for Use:
The Laserseope VELA Erbium: YAG Laser System and Accessories are intended for the surgical incision. vaporization and congulation of soft tissue. All soft tissue is included such as skin, subcutaneous tissue, striated and smooth tissue, muscle, cartilage menus membranc, lymph vessels and nodes, organs and glands.
DermatologyPlastic Surgery: Indications include, cpidcrmal nevi, teleagiestasia, soider veins, scinic chailitis, keloids, verrucee, skin togs, keratoses, sear revision, debulking banign tumors, decubitis ulcers and skin resurfacing.
General Surgery: The Erbium: YAG lass is intended for the surgical incision, vaportuation and coagulation of soft tissue during any general surgery application where skin incision, tissue of external lumors and lesions, complete or partial resection of internal organs, tissue abletion and/or vessel coagulation may be indicated.
Genitourinary: Indications include lesions of the external genitalia, welcon and anus, penis, scrown and urchire (includes condyloma acuminata, gian perined condyloma und verrucous carcinoma), valvar lesions, polyps and familial polyps of the colon.
Gynecology: Indications include cervical intraepithelial neoplusia (CDV), herpes simplex, endometrial adhesions, cysts and condyloms.
ENT: Indications include car, nose and throat lessions, polyps, cysts, hyperkeratosis; excision of carcinogenic lissue, oral leukoplakia.
Oral/Maxillofacial: Indications include benign oral tunnors, oral and glossel lessions and gingivectory.
Ophthalmology: Indications include soft tissue surrounding the eye and orbit and unterior capsulotomy.
Podiatry: Indications include warts, plantar verrucae, large mossic verrucae and matrixectory .
Prescription Use
(Per 21 CFR 801.109)
bcsolles
gn-Off)
General Restorative Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.