← Product Code [GEX](/productcode/GEX) · K971821

# UNICA 315 M CO2 SURGICAL LASER SYSTEM (K971821)

_Reliant Technologies, Inc. · GEX · Sep 25, 1997 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K971821

## Device Facts

- **Applicant:** Reliant Technologies, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Sep 25, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Unica 315M R/G is indicated for use as a surgical instrument to vaporize, incise, excise, ablate or photocoagulate soft tissue in general surgical procedures.

## Device Story

The Unica 315M R/G is a portable CO2 surgical laser system used by clinicians to perform soft tissue surgery (vaporization, incision, excision, ablation, photocoagulation). The device utilizes a sealed waveguide CO2 laser tube (10,600 nm) to deliver energy via a 4-joint articulated arm. It features an aiming beam (532 nm or 635 nm) to assist the surgeon in targeting. Operation modes include Superpulse, Pulse, and Continuous, with single or repeat pulse trains. The system is air-cooled, battery-backed for 1-5 hours of operation, and can be integrated with microscopes or colposcopes via micromanipulators. The clinician operates the device to deliver precise laser energy to the target tissue, facilitating surgical procedures. The device is designed for portability and robustness, with internal overheat protection and a sealed laser tube requiring no gas refills.

## Clinical Evidence

Bench testing only. No clinical data provided. Equivalence is based on the long-standing history of CO2 laser use and comparison of operational/clinical parameters with legally marketed predicate devices.

## Technological Characteristics

CO2 laser, sealed waveguide tube, TEM00 Gaussian mode. Wavelength: 10,600 nm. Aiming beam: 532 nm or 635 nm. Power: 1-15 Watts. Beam delivery: 4-joint articulated arm. Cooling: Internal air-cooled with overheat protection. Power: 110-120V/220-230V. Battery backup: 1-5 hours. Dimensions: 19.6 x 6.4 x 7.5 inches. Weight: 33 lbs. Complies with 21 CFR 1040 and 1010.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Luxar Corporation LX-20SP Laser System ([K953074](/device/K953074.md))
- Coherent Laser Systems System 451 CO2 Laser System ([K951812](/device/K951812.md))
- Laser Industries Sharplan 1020 CO2 Surgical Laser System ([K860087](/device/K860087.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

SEP 20 2 3

in

Summary of Safety and Effectiveness Information Reliant Technologies, Inc. Unica 315M R/G CO2 Laser System Safe Medical Devices Act of 1990, 21CFR807.92

- 1. Device Name(s)

| Trade Name:           | Unica 315M R/G CO₂ Surgical Laser System                                                                                                                       |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:          | CO₂ Surgical Laser                                                                                                                                             |
| Classification Name:  | Instrument, Surgical, Powered, Laser<br>§ 878.4810 Laser surgical instrument for use in general and plastic surgery and dermatology. Classification: Class II. |
| Classification Panel: | General & Plastic Surgery Devices Panel                                                                                                                        |
| Product Code(s):      | 79-GEX                                                                                                                                                         |

# 2. Establishment Name & Registration Number:

Reliant Technologies, Inc. Name:

Number: 2950711

#### 3. Special Controls:

This device complies with CDRH radiation control regulations as outlined in:

21 CFR §1040, Performance Standards for Light Emitting Products

21 CFR §1010, Performance Standards for Electronic Products, general.

#### 4. Labeling:

The Unica 315M R/G is indicated for use as a surgical instrument to vaporize, incise, excise, ablate or photocoagulate soft tissue in general surgical procedures.

Required laser warning and caution statements are displayed as required for surgical laser devices.

{1}------------------------------------------------

#### 5. Device Description:

| Weight:                       | 33 lb. (15 kg.)                                                                                        |
|-------------------------------|--------------------------------------------------------------------------------------------------------|
| Dimensions:                   | Length:<br>19.6 in. (49 cm.)<br>Height (folded arm):<br>7.5 in. (18 cm.)<br>Width:<br>6.4 in. (16 cm.) |
| Main Laser Source:            | CO2 laser, sealed wave guide tube TEM00.                                                               |
| Treatment Beam<br>Wavelength: | 10,600 nm CO2 laser                                                                                    |
| Aiming Beam<br>Wavelength:    | 532 nm KTP or 635 nm GaAlAs-laser - max. output 5 milliwatt.                                           |
| Modes of Operation*:          | Superpulse/Pulse/Continuous<br>Single or Repeat pulse trains                                           |
| Beam Delivery:                | 4 joint articulated arm that can be partially folded into the laser.                                   |
| Battery backup:               | Equipped with backup battery for 1 to 5 hours of operation<br>depending on output power etc.           |
| Battery charge cycle:         | Full charge is achieved in 5 hours                                                                     |
| Cooling system:               | Internal cooling to instrument case. Overheat protection circuit.                                      |
| Electric requirements:        | Single phase 220-230V 50 Hz, <0.5A, or<br>Single phase 110-120V 60 Hz, <1.0A.                          |

*Reliant Technologies, Inc. continually incorporates innovations into its products. Product specifications may change without notice.

- It contains a sealed laser tube, i.e. no gas filling is needed. The typical life time of the laser . tube is more than 10,000 hours of operation.
- It is mechanically robust and electrically very safe does not contain any higher voltage . than 32 volts.
- It is easy to transport. The 315 M can be hand-carried without difficulty. The articulated . arm can be partially folded into the laser.
- It has a backup battery. The laser output is independent of any line voltage fluctuations or . even total voltage loss for hours.

{2}------------------------------------------------

It can be connected via micromanipulators to microscopes and colposcopes. Special accessories can be employed, e.g. for angular cutting.

#### Attachments/Accessories:

The Unica 315 M CO2 Laser is compatable with the following existing legally marketed Reliant Technologies, Inc. laser attachments and accessories:

- 1. UniMax Model 2000 Color Corrected MicroSpot CO2 Laser Micromanipulator -K920821.
- 2. Acutome 2000 Color Corrected Laparoscopic CO2 Laser Coupler and Sleeves K942868.
- 3. Acutome 100 Color Corrected CO2 Laser Handpiece K942931.
- 4. AccuScan Model C CO2 Laser Scanner K953670.

# 6. Applicant Name & Address:

Reliant Technologies, Inc. 1153 Triton Drive, Suite C Foster City, CA 94404 415.570.6831 - 415.377.0887 (fax)

# 7. Company Contact:

Ms. Lana Black Reliant Technologies, Inc. 1153 Triton Drive, Suite C Foster City, CA 94404 415.570.6831 - 415.377.0887 (fax)

# 8. Substantially Equivalent Device(s):

- 1. Luxar Corporation. LX-20SP Laser System
- 2. Coherent Laser Systems. System 451 CO2 Laser System
- 3. Laser Industries. Sharplan 1020 CO2 Surgical Laser System

{3}------------------------------------------------

# 9. Equivalence:

Based on the long standing history of use of CO2 lasers and the operational and clinical parameters of the Unica 315 M and the referenced comparison devices, Reliant Technologies, Inc. believes that the Unica 315 M laser is substantially equivalent to devices presently legally marketed in the US.

#### 10. Feature Comparison Table:

| SPECIFICATION                 | UNICA 315M  R/G CO2<br>Laser System                               | Luxar - LX-20SP<br>Laser System                                   | Coherent<br>System 451 CO2<br>Laser System                        | Sharplan 1020 CO2<br>Surgical<br>Laser System                     |
|-------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------|
| LASER MEDIUM -<br>WAVELENGTH: | CO2 -10.6 micron                                                  | CO2 -10.6 micron                                                  | CO2 -10.6 micron                                                  | CO2 -10.6 micron                                                  |
| BATTERY BACKUP:               | 1-5 hours depending on<br>laser output power                      | NA                                                                | NA                                                                | NA                                                                |
| OUTPUT Power:                 | 1-15 Watt                                                         | 2 - 20 Watt                                                       | 0.5 - 100 Watt                                                    | 1-20 Watt                                                         |
| MODE:                         | TEMoo, Gaussian                                                   | TEMo1, Multimode                                                  | TEMoo, Gaussian                                                   | TEMoo, Gaussian                                                   |
| Exposure Times:               | Single pulse and variable,<br>0.05 -continuous sec.               | Single pulse and variable,<br>0.05 -continuous sec.               | Single pulse and variable,<br>0.05 -continuous sec.               | Single pulse and variable,<br>0.05 -continuous sec.               |
| POWER<br>REQUIREMENTS:        | 110-120 VAC or<br>220-240 VAC                                     | 110-120 VAC or<br>220-240 VAC                                     | 117-220 VAC                                                       | 120 VAC                                                           |
| AIMING BEAM:                  | 635 nm GaA1As - up to 5<br>mW<br>532 nm up to 3 mW                | NA                                                                | HeNe 0.8 mW- 632.8 nm                                             | HeNe 2.0 mW- 632.8 nm                                             |
| BEAM DELIVERY:                | Articulated arm &<br>accessories                                  | Hollow waveguide                                                  | Articulated arm &<br>accessories                                  | Articulated arm &<br>accessories                                  |
| WEIGHT:                       | 33 lb./15 Kg.                                                     | 40 lb./18.4 kg                                                    | 350 lb./160 kg                                                    | 80 Ib./36 Kg                                                      |
| COOLING:                      | Air Cooled                                                        | Air Cooled                                                        | Water                                                             | Water                                                             |
| INDICATIONS:                  | Cutting, coagulation.<br>excision and ablation of soft<br>tissue. | Cutting, coagulation,<br>excision and ablation of soft<br>tissue. | Cutting, coagulation,<br>excision and ablation of soft<br>tissue. | Cutting, coagulation,<br>excision and ablation of<br>soft tissue. |
| K Numbers:                    | Pending                                                           | K953074                                                           | K951812                                                           | K860087                                                           |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird-like figure with three distinct head profiles layered on top of each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird image.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856

SEP 2 5 1997

Ms. Lana Black .Vice President and Official Correspondent Reliant Technologies, Inc. 1153 Triton Drive, Suite C Foster City, California 94404

Re: K971821 Trade Name: Unica 315M R/G - CO, Surgical Laser System Regulatory Class: II Product Code: GEX Dated: August 21, 1997 Received: August 22, 1997

Dear Ms. Black:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

fccellge

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# 14971821

# Indication for Use

The Unica 315M R/G is indicated for use as a surgical instrument to vaporize, incise, excise, ablate or photocoagulate soft tissue in general surgical procedures

Concurrence of CDRH, Office of Device Evaluation (ODE)

tcoello

(Division Sign-Off) Division of General Re-510(k) Number

Prescription Use V (Per 21CFR§801.109)

OR

Over-The-Counter Use

---

**Source:** [https://fda-staging.innolitics.com/device/K971821](https://fda-staging.innolitics.com/device/K971821)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
