The IMMULITE® Free T4 (2-Step) assay is designed for the quantitative measurement of non-protein-bound thyroxine (free T4) levels in serum. It is intended strictly for in vitro use as an aid in the clinical assessment of thyroid status.
Device Story
IMMULITE® Free T4 (2-Step) is a solid-phase, chemiluminescent enzyme immunoassay performed on the IMMULITE® Automated Immunoassay Analyzer. Input: patient serum sample. Process: sample incubated with monoclonal antibody-coated polystyrene bead; unbound material removed via centrifugal wash; alkaline phosphatase-labeled triiodothyronine added; second incubation and wash; chemiluminescent substrate (phosphate ester of adamantyl dioxetane) added. Operation: substrate hydrolysis by alkaline phosphatase produces light; photon output measured by luminometer is inversely proportional to free T4 concentration. Used in clinical laboratories by technicians/lab personnel. Output: quantitative free T4 concentration. Clinical utility: aids assessment of thyroid status, particularly when carrier protein (TBG) levels are altered, helping distinguish between thyroid dysfunction and physiological/pathological carrier protein variations.
Clinical Evidence
Method comparison study performed on 172 patient samples comparing IMMULITE® Free T4 (2-Step) to the predicate Coat-A-Count® Free T4. Results: linear regression slope 1.01, intercept -0.02 ng/dL, correlation coefficient r=0.946. Mean values for both assays were 1.6 ng/dL.
Technological Characteristics
Solid-phase chemiluminescent enzyme immunoassay. Components: polystyrene bead coated with monoclonal anti-T4 antibody. Energy source: automated analyzer luminometer. Detection: chemiluminescent substrate (phosphate ester of adamantyl dioxetane) and alkaline phosphatase label. Connectivity: integrated with IMMULITE® Automated Immunoassay Analyzer. Form factor: reagent system for automated analyzer.
Indications for Use
Indicated for quantitative measurement of free T4 in serum as an aid in clinical assessment of thyroid status. Useful for assessing thyroid status in patients with TBG elevations (e.g., pregnancy, oral contraceptive use, estrogen therapy) and detecting TBG level alterations that mask abnormal thyroid function in hypothyroid and hyperthyroid patients.
Regulatory Classification
Identification
A free thyroxine test system is a device intended to measure free (not protein bound) thyroxine (thyroid hormone) in serum or plasma. Levels of free thyroxine in plasma are thought to reflect the amount of thyroxine hormone available to the cells and may therefore determine the clinical metabolic status of thyroxine. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or other bird in flight, composed of three curved lines.
JUL 25 1997
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Edward M. Levine, Ph.D. Director of Clinical Affairs Diagnostic Products Corporation 5700 West 96th Street Los Angeles, CA 90045-5597
K971792 Re: IMMULITE® Free T4 (2-Step) Regulatory Class: II Product Code: CEC Dated: May 14, 1997 Received: May 15, 1997
Dear Dr. Levine:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{1}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as …… described in your 510 (k) premarket notification. "The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# 510 (k) Summary Safety and Effectiveness
JUL 2 5 1997 5971792
This summary of safety and effectiveness information has been prepared in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.
Name: Address:
Telephone Number: Facsimile Number:
Contact Person:
Date of Preparation:
Device Name Trade: Common:
Catalog Number:
Classification: CLIA Complexity Category:
Manufacturer:
Establishment Registration Number:
Substantially Equivalent Predicate Device:
Description of Device:
Intended Use of the Device:
Diagnostic Products Corporation 5700 West 96th Street Los Angeles, California 90045-5597
(213) 776-0180 (213) 776-0204
Edward M. Levine, Ph.D. Director of Clinical Affairs
May 14, 1997
IMMULITE® Free T4 (2-Step) Reagent system for the determination of free T4 in serum
LKFT1 (100 tests), LKFT5 (500 tests)
Class II device, 75-CEC (862.1695) Moderate, based on previous classification of analogous tests.
Diagnostic Products Corporation 5700 West 96th Street Los Angeles, California 90045-5597
DPC's Registration Number is 2017183
DPC's Coat-A-Count® Free T4 (K822882)
IMMULITE® Free T4 (2-Step) is a solid-phase, Chemiluminescent enzyme immunoassay for use with the IMMULITE® Automated Immunoassay Analyzer.
The IMMULITE® Free T4 (2-Step) assay is designed for the quantitative measurement of non-protein-bound thyroxine (free T4) levels in serum. It is intended strictly for in vitro use as an aid in the clinical assessment of thyroid status.
## Summary and Explanation of the Test:
The principal thyroid hormone, thyroxine (T4) circulates almost entirely bound to carrier proteins, chief of which is thyroxine-binding globulin (TBG), in an equilibrium that tends to reassert itself in the face of altered levels of the carrier proteins by inducing a corresponding alteration in the total level of T4 in
{3}------------------------------------------------
Diagnostic Products Corporation IMMULITE®Free T4 (2-Step) May 14, 1997
## Summary and Explanation of the Test (continued):
circulation, while leaving the free T4 concentration level relatively unchanged. Hence, the free T4 concentration may be expected to correlate more closely than the total T4 concentration with clinical thyroid status, for an abnormal total T4 result may signify either an abnormality in thyroid function or simply a variation (physiological or pathological) in the carrier proteins.
# Technological Comparison to Predicate:
DPC's IMMULITE® Free T4 (2-Step) is a solid-phase, 2-step chemiluminescent immunoassay. The solid phase, a polystyrene bead enclosed within an IMMULITE® Test Unit, is coated with a monoclonal antibody for T4.
The patient sample is introduced into the Test Unit, and incubated for approximately 30 minutes at 37°C with intermittent agitation. During this time, free T4 in the sample binds to antibody binding sites on the bead. Unbound material is then removed by a centrifugal wash.
An alkaline phosphatase-labeled triiodothyronine is introduced, and the Test Unit is incubated for approximately another 30-minute cycle. The unbound enzyme conjugate is removed by a centrifugal wash. Substrate is then added, and the Test Unit is incubated for a further 10 minutes.
The chemiluminescent substrate, a phosphate ester of adamantyl dioxetane, undergoes hydrolysis in the presence of alkaline phosphatase to yield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light, thus improving precision by providing a window for multiple readings. The bound complex -- and thus also the photon output, as measured by the luminometer -- is inversely proportional to the concentration of free T4 in the sample.
The IMMULITE® Free T4 (2-step) system has been optimized to leave essentially undisturbed the original equilibrium between free and protein-bound T4 in the patient sample.
The DPC's Coat-A-Count® Free T4 procedure is a solid-phase radioimmunoassay, wherein 127-labeled T4 analog competes for a fixed time with free T4 in the patient sample for sites on T4-specific antibody. Because the antibody is immobilized to the wall of a polypropylene tube, simply decanting the supernatant suffices to terminate the competition and to isolate the antibody-bound fraction of the radiolabeled free T4. Counting the tube in a gamma counter then yields a number, which converts by way of calibration curve to measure of the free T4 present in the patient sample.
The Coat-A-Count® Free T4 system has been optimized to eliminate all binding of the T4 analog tracer to endogenous proteins, while leaving essentially undisturbed the original equilibrium between free and protein-bound T4 in the patient sample. The tracer itself has no measurable affinity for thyroxin-binding globulin (TBG), the principal thyroid hormone transport protein. To prevent binding of the tracer to albumin, blocking agents are present, at a concentration carefully adjusted to avoid displacement of native T4 from endogenous carrier proteins. To further minimize the risk of "stripping", the system employs an antibody, at low concentration, with an affinity for T4 slightly less than that of TBG, and operates under physiological conditions of temperature, pH and ionic strength.
{4}------------------------------------------------
Diagnostic Products Corporation IMMULITE® Free T4 (2-Step) May 14, 1997
## Method Comparison:
The IMMULITE® Free T4 (2-Step) procedure was compared to Coat-A-Count® Free T4 assay on 172 patient samples, with free T4 concentrations ranging from approximately 0.3 to 4.8 ng/dl. Linear regression analysis yielded the following statistics:
(IMMULITE® Free T4 2-Step) = 1.01 (Coat-A-Count® Free T4) - 0.02 ng/dL r = 0.946
1.6 ng/dL (IMMULITE® Free T4 2-Step) Mean Values: 1.6 ng/dL (Coat-A-Count® Free T4)
## Performance Equivalence:
Diagnostic Products Corporation asserts that IMMULITE® Free T4 (2-Step) is substantially equivalent to other commercially marketed free T4 assays, such as Coat-A-Count® Free T4. Each product is intended strictly for in vitro diagnostic use as an aid in the clinical assessment of thyroid status.
### Conclusion:
The data presented in this summary of safety and effectiveness is the data that the Food and Drug Administration used in granting DPC substantial equivalence for IMMULITE® Free T4 (2-Step).
Edward L. Renie
Edward M. Levine, Ph.D. Director, Clinical Affairs
sty/ez
Date
{5}------------------------------------------------
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name: IMMULITE* Free T4 (2-Step) Indications For Use:
DPC's IMMULITE® Free T4 is intended for use with the IMMULITE® Automated Analyzer. It is a solid-phase, 2-step chemiluminescent enzyme immunoassay designed for the ouantitative measurement of non-protein-bound thyroxine (free T4) levels in serum. It is intended strictly for in vitro use as an aid in the clinical assessment of thyroid status related to abnormality in thyroid function or simply a variation (physiological or pathological) in the carrier proteins, chief of which is thyroxine-binding globulin (TBG). This assay is useful in assessment of thyroid status due to TBG elevations typical of pregnancy, oral contraceptive use and estrogen therapy. Additionally, this test system is for use as an aid in detecting alterations in the TBG level which mask the effects of abnormal thyroid function in hypothyroid and hyperthyroid patients.
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number k971797
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.