# DENVER PLEURX PLEURAL CATHETER KIT, CATALOG NO. 50-7000, DENVER PLEURX HOME DRAINAGE KIT, CATALOG NO. 50-7500 (K971753)

_Denver Biomaterials, Inc. · DWM · Jun 27, 1997 · Cardiovascular · SESK_

**Canonical URL:** https://fda-staging.innolitics.com/device/K971753

## Device Facts

- **Applicant:** Denver Biomaterials, Inc.
- **Product Code:** [DWM](/productcode/DWM.md)
- **Decision Date:** Jun 27, 1997
- **Decision:** SESK
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5050
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Denver® PLEURX Pleural Catheter Kit and Denver® PLEURX Home Drainage Kit are indicated for intermittent,longterm drainage of symptomatic, recurrent, malignant pleural effusion. The devices are indicated for the palliation of dyspnea due to malignant pleural effusion and for providing pleurodesis.

## Device Story

The Denver® PLEURX Pleural Catheter Kit is a sterile, single-use system for managing malignant pleural effusions. It consists of a flexible, fenestrated silicone catheter with a cuff and valve, placed in the pleural space via the Seldinger technique; the cuff is tunneled subcutaneously. The Home Drainage Kit serves as an accessory for patient-managed drainage, utilizing vacuum bottles to collect pleural fluid. The system allows for intermittent, ambulatory drainage, providing relief from dyspnea and facilitating pleurodesis. The device is used in clinical and home settings; operation involves healthcare providers for placement and patients or caregivers for drainage. By enabling controlled, intermittent removal of fluid, the device improves patient quality of life and daily activity capacity.

## Clinical Evidence

Clinical evaluation under IDE #930085. Subjects showed significant improvement in dyspnea (Friedman test) and quality of life (t-test) compared to baseline. Approximately 50% of subjects achieved pleurodesis; no subjects developed symptomatic recurrence of effusion during the study.

## Technological Characteristics

Materials: Silicone catheter with radiopaque stripe. Design: Fenestrated catheter with cuff and valve. Energy source: Vacuum-based drainage. Connectivity: None. Sterilization: Single-use, sterile. Software: None.

## Regulatory Identification

A patient care suction apparatus is a device used with an intrathoracic catheter to withdraw fluid from the chest during the recovery period following surgery.

## Predicate Devices

- Argyle® Silicone Thoracic Catheter
- Denver® Pleural Effusion Shunt with Externalized Pump Chamber

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K971753

Summary of Safety and Effectiveness Denver® PLEURX Pleural Catheter Kit Denver® PLEURX Home Drainage Kit

JUN 2 7 1997

SUBMITTER INFORMATION

Denver Biomaterials, Inc. 14998 W. 6th Avenue, Bldg. E700 Golden, Colorado 80401 USA

Lynne Leonard Director, Regulatory Affairs/Quality Assurance

DEVICE COMMON NAME

Patient Care Suction Apparatus

DEVICE CLASSIFICATION NAME

21 CFR 870.5050

IDENTIFICATION OF SUBSTANTIALLY EQUIVALENT DEVICES

The devices are substantially equivalent to the Argyle® Silicone Thoracic Catheter in intended use. They differ in that the Argyle® Silicone Thoracic Catheter is intended for continuous drainage, while the PLEURX Catheter is intended for intermittent, ambulatory drainage.

The devices are substantially equivalent to the Denver® Pleural Effusion Shunt with Externalized Pump Chamber in materials and dimensional specifications. They differ in that the Denver® Pleural Effusion Shunt with Externalized Pump Chamber transfers pleural fluid to the peritoneal cavity, while the PLEURX Catheter transfers fluid to a vacuum source.

Differences between the devices and their predicate devices should not affect their safety or effectiveness.

DEVICE DESCRIPTION:

The Denver® PLEURX Pleural Catheter Kit (#50-7000) is a sterile device, for single use only, and is not to be resterilized. The kit contains the PLTURX Catheter; a specifically matched drainage line that can be configured for either wall suction or a vacuum bottle; and the components needed to place the catheter within the pleural space and to dress the catheter exit site after placement .

{1}------------------------------------------------

The PLEURX Catheter contained in the kit is a flexible, fenestrated silicone catheter with a cuff and a valve. The fenestrated portion of the catheter is inserted into the pleural space utilizing a peel-away introducer (Seldinger technique). The portion of the catheter containing the cuff is tunneled subcutaneously, and the remaining portion is left external to the A radiopaque stripe runs the entire length of the body . . catheter.

The Denver® PLEURX Home Drainage Kit (#50-7500), an accessory to the Denver® PLEURX Pleural Catheter Kit, is a sterile, disposable device designed for single use only, and is not to be The kit contains drainage lines, vacuum bottles, resterilized. and the components needed to drain the effusion and to redress the catheter exit site.

### INDICATED USE

The devices are indicated for intermittent, long-term drainage of symptomatic, recurrent, malignant pleural effusion. They are indicated for the palliation of dyspnea due to malignant pleural effusion, and for providing pleurodesis.

BIOCOMPATABILITY TESTING:

The biological safety of the Denver® PLEURX Pleural Catheter Kit has been assured through the selection of materials which are currently used in legally marketed devices.

CLINICAL TESTING:

The Denver® PLEURX Pleural Catheter Kit and Denver® PLEURX Home Drainage Kit were evaluated clinically under IDE #930085.

- Effectiveness of PLEURX Catheter for Relief of Dyspnea 1.
When compared to baseline, subjects showed a significant improvement in their rating of dyspnea after treatment with the PLEURX Catheter, as determined by the Friedman test.

#### Effectiveness of PLEURX Catheter for Improving the 2. Quality of Life

Subjects experienced a significant improvement in quality of life after treatment with the PLEURX Catheter, as determined by the Although the subjects had serious malignant disease, t-test. they experienced an improved ability to carry out daily activities when their pleural effusions were controlled.

#### Effectiveness of PLEURX Catheter for Resulting in 3. Pleurodesis

Approximately half of the subjects treated with the PLEURX Catheter achieved pleurodesis while in the study. No subjects in the PLEURX Catheter arm developed a documented symptomatic recurrence of their effusion.

{2}------------------------------------------------

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Lynne Leonard Director, Regulatory Affairs and Quality Assurance Denver Biomaterials, Inc. 14998 W. 6th Avenue, Bldg. E 700 Golden, Colorado 80401

JUN 27 1397

K971753 Re: Denver® PLEURX Pleural Catheter Kit and Denver® PLEURX Home Drainage Kit Regulatory Class: II (two) Product Code: 74 DWM Dated: May 8, 1997 Received: May 12, 1997

Dear Ms. Leonard:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510 (k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Please note: If you purchase your device components in Amendments. bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510 (k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to additional controls. Existing major regulations affecting your device

{3}------------------------------------------------

# Page 2 - Ms. Lynne Leonard

can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, FDA will verify such assumptions. Failure to comply with the CMP regulation may result in requlatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

FDA notes that your device will contain sutures for which you have provided evidence that the suture characteristics are not altered by the sterilization process used for the device. However, you should be aware of the following additional information regarding the inclusion of a suture as a component of your device:

- The labeling, packaging and method of sterilization of 1. the suture cannot be changed without prior notification, review and clearance by FDA.
- 2. The supplier of the sutures used in your device can not be changed without prior notification, review and clearance by FDA.

In addition, we have determined that your device kit contains lidocaine which is subject to regulation as a drug.

Our substantially equivalent determination does not apply to the drug component of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug component. For information on applicable Agency requirements for marketing this drug, we suggest you contact:

> Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0063

{4}------------------------------------------------

Page 3 - Ms. Lynne Leonard

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Fart 801 and additionally 809 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act, may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Atora J. Callaba

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## INDICATIONS FOR USE

| 510 (K) Number:      | K 971753                                                                    |
|----------------------|-----------------------------------------------------------------------------|
| Device Names:        | Denver® PLEURX Pleural Catheter Kit,<br>Denver® PLEURX Home Drainage Kit    |
| Indications for Use: | The Denver® PLEURX Pleural Catheter Kit<br>and Denver® PLEURX Home Drainage |

and Denver® PLEURX Home Drainage Kit are indicated for intermittent,longterm drainage of symptomatic, recurrent, malignant pleural effusion. The devices are indicated for the palliation of dyspnea due to malignant pleural effusion and for providing pleurodesis.

Concurrence of CDRH, Office of Device Evaluation (ODE)

ui. Daguinal m.D.

6/26/27

Prescription Use: Yes OR
(Per 21 CFR 801.109)

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Over-the-Counter Use: No

----------

---

**Source:** [https://fda-staging.innolitics.com/device/K971753](https://fda-staging.innolitics.com/device/K971753)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).