K971735 · Schick Technologies, Inc. · KGI · Dec 2, 1997 · Radiology
Device Facts
Record ID
K971735
Device Name
ACCUDEXA BONE DENSITOMETER
Applicant
Schick Technologies, Inc.
Product Code
KGI · Radiology
Decision Date
Dec 2, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1170
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The accuDEXA is a dual-energy x-ray device indicated for use in measuring an index of the bone density of the middle finger of the nondominant hand (BMD). This BMD value is a relative indicator of bone density elsewhere in the body. The measurement is compared to similar measurements from patients of the same gender and ethnicity to provide t-scores (number of standard deviations above or below the mean for a reference group of young healthy individuals) and z-scores (number of standard deviations above or below the mean for a cohort of the same age). A t-score or a z-score may be used by a physician as one factor, in conjunction with other clinical indicators, to diagnose osteoporosis and other bone disorders. When a normative database of the same ethnicity and gender is not available, the BMD value may still be used to compare to the patient's own baseline value, for example in following the patient's response to treatment for osteoporosis. Below normal bone density can be associated with a variety of conditions or disorders of bone. For example, the Society of Nuclear Medicine and American College of Nuclear Physicians has identified specific medical indications for measurement of a patient's bone mass: for patients with premenopausal oophorectomy, spontaneous menopause, or estrogen deficiency conditions; for treatment-related osteopenia; when the diagnosis of osteopenia is suggested or established by other means, such as x-ray; during long-term immobilization; for endocrinopathies known to be associated with osteopenia; for post-gastrectomy and other malabsorption states leading to osteopenia: during long-term corticosteroid therapy; for chronic renal disease, particularly in childhood or adolescence; and to monitor treatment programs for osteoporosis.
Device Story
AccuDEXA is a Dual Energy X-Ray Absorptiometer (DEXA) for bone mineral density (BMD) measurement. Device captures images of the middle finger of the non-dominant hand using two distinct X-ray energy levels generated by varying X-ray tube high voltage. Algorithms analyze these images to calculate BMD and bone mineral content (BMC). Results are compared against a normative database to generate t-scores and z-scores. Used by clinicians to assist in diagnosing osteoporosis and monitoring treatment response. Provides a relative indicator of systemic bone density. Benefits include non-invasive assessment of bone mass in patients with risk factors for osteopenia.
Clinical Evidence
Bench testing only. Study on cadavers evaluated precision, accuracy, and repeatability of measurements. Results showed high repeatability with repositioning and high correlation with ashed bone weights. Performance compared favorably to predicate devices. A clinical study was conducted to develop a normative reference database for t-score and z-score calculations.
Technological Characteristics
Dual Energy X-Ray Absorptiometer (DEXA). Uses digital imaging array. Operates by modulating X-ray tube high voltage to produce two energy levels. Computes BMD and BMC via algorithmic analysis of finger images. Includes normative database for score calculation.
Indications for Use
Indicated for patients requiring bone mineral density (BMD) assessment of the middle finger of the non-dominant hand to diagnose osteoporosis or monitor bone disorders. Applicable to patients with conditions associated with osteopenia, including premenopausal oophorectomy, menopause, estrogen deficiency, treatment-related osteopenia, long-term immobilization, endocrinopathies, malabsorption states, long-term corticosteroid therapy, and chronic renal disease.
Regulatory Classification
Identification
A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
The Osteogram Radiographic Absorptiometry (RA) device
Schick CDR Digital Dental X-ray Receptor Array (K933455)
Submission Summary (Full Text)
{0}------------------------------------------------
KIMDS
DEC - 2 1997
## 510(k) Summary
# Schick accuDEXA Bone Densitometer
Common/Classification Name: Bone Densitometer 21 CFR 892.1170
Schick Technologies, Inc. 31-00 47th Avenue Long Island City, NY 11101 718-937-5765, 718-937-5962 (FAX) Contact: Enrico Renzi, Prepared: May 8, 1997
#### LEGALLY MARKETED PREDICATE DEVICES A.
The accuDEXA is substantially equivalent to bone densitometers The Osteogram Radiographic currently marketed in the U.S. Absorptiometry (RA) device, a pre-Amendment device originally from Compumed (now distributed by Merck), is similar in that it uses the three middle fingers of the hand to compute the bone density, though it uses X-ray film for recording. The Hologic QDR 2000+ DXA bone densitometer (K913321) is similar in that it uses the same dual-energy x-ray absorptiometry (DXA) technology, though it uses the forearm as the measurement site. The Schick CDR Digital Dental X-ray Receptor Array (K933455) currently marketed by Schick Technologies uses a similar digital imaging array.
### DEVICE DESCRIPTION B.
The accuDEXA device is a Dual Energy X-Ray Absorptiometer (DEXA) device. The device is intended to calculate an index of bone mineral density in the middle finger of the non-dominant hand. By changing the high voltage on the X-ray tube, two energies are produced. Each of the two settings produces an image of the finger and each image is analyzed using various algorithms to produce a value of bone mineral density (BMD) and bone mineral content (BMC). These values are compared with a normative database, yielding a t-score and a z-score. The t-score is the number of standard deviations that the patient is above or below the mean of a reference sample of young healthy individuals. The zscore is the number of standard deviations that the patient is above or below the mean of a reference sample of individuals of the same age as the patient.
{1}------------------------------------------------
### INDICATIONS FOR USE C.
The accuDEXA is a dual-energy x-ray device indicated for use in measuring an index of the bone density of the middle finger of the nondominant hand (BMD). This BMD value is a relative indicator of bone density elsewhere in the body. The measurement is compared to similar measurements from patients of the same gender and ethnicity to provide t-scores (number of standard deviations above or below the mean for a reference group of young healthy individuals) and z-scores (number of standard deviations above or below the mean for a cohort of the same age). A t-score or a z-score may be used by a physician as one factor, in conjunction with other clinical indicators, to diagnose osteoporosis and other bone disorders. When a normative database of the same ethnicity and gender is not available, the BMD value may still be used to compare to the patient's own baseline value, for example in following the patient's response to treatment for osteoporosis.
Below normal bone density can be associated with a variety of conditions or disorders of bone. For example, the Society of Nuclear Medicine and American College of Nuclear Physicians has identified specific medical indications for measurement of a patient's bone mass: for patients with premenopausal oophorectomy, spontaneous menopause, or estrogen deficiency conditions; for treatment-related osteopenia; when the diagnosis of osteopenia is suggested or established by other means, such as x-ray; during long-term immobilization; for endocrinopathies known to be associated with osteopenia; for post-gastrectomy and other malabsorption states leading to osteopenia: during long-term corticosteroid therapy; for chronic renal disease, particularly in childhood or adolescence; and to monitor treatment programs for osteoporosis.
### D. SUBSTANTIAL EQUIVALENCE SUMMARY
The intended use for the accuDEXA and the predicate devices is the same. The accuDEXA Bone Densitometer has the same technological characteristics as the predicate devices. However, the descriptive characteristics may not be sufficiently precise to assure substantial equivalence. Therefore, Schick Technologies sponsored performance tests on the device and the predicate devices (see Section F, below), and a comparison of these data demonstrate that the accuDEXA Bone Densitometer is substantially equivalent to the predicate devices.
### E. TECHNOLOGICAL CHARACTERISTICS
See Device Description, above.
{2}------------------------------------------------
### TESTING F.
Schick Technologies has sponsored a study on cadavers that examines the precision and accuracy of the accuDEXA device. The study also directly compared the bone density values obtained with the predicate devices. This study showed that the accuDEXA measurements are very repeatable, even with repositioning. The very high correlation of the accuDEXA measurements with the ashed bone weights shows that the accuDEXA measurements are very accurate. The good correlation of the accuDEXA measurement with measurements with the two predicate devices demonstrates that the accuDEXA results for the middle finger are at least as good an indicator of BMD as the currently marketed predicate devices.
A reference data base has been developed in a clinical study which allows the calculation of t-scores and z-scores for each patient.
### G. CONCLUSIONS
Schick Technologies has demonstrated through its comparison of characteristics with the predicate devices and comparison of performance testing with the predicate devices that the accuDEXA Bone Densitometer is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human profiles facing to the right.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC = 2 1997
Schick Technologies, Inc. c/o T. Whit Athey, Ph.D. Senior Consultant C.L. McIntosh & Associates, Inc. 12300 Twinbrook Parkway, Suite 625 Rockville, Maryland 20852
Re:
K971735 AccuDEXA Bone Densitometer Dated: September 11, 1997 Received: September 11, 1997 Regulatory class: II 21 CFR 892.1170/Procode: 90 KGI
Dear Dr. Athey:
::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::
We have reviewed your Section 510(k) notification of intent to market the device and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
W.Liau Yu
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
KIMDS
DEC - 2 1997
## 510(k) Summary
# Schick accuDEXA Bone Densitometer
Common/Classification Name: Bone Densitometer 21 CFR 892.1170
Schick Technologies, Inc. 31-00 47th Avenue Long Island City, NY 11101 718-937-5765, 718-937-5962 (FAX) Contact: Enrico Renzi, Prepared: May 8, 1997
#### LEGALLY MARKETED PREDICATE DEVICES A.
The accuDEXA is substantially equivalent to bone densitometers The Osteogram Radiographic currently marketed in the U.S. Absorptiometry (RA) device, a pre-Amendment device originally from Compumed (now distributed by Merck), is similar in that it uses the three middle fingers of the hand to compute the bone density, though it uses X-ray film for recording. The Hologic QDR 2000+ DXA bone densitometer (K913321) is similar in that it uses the same dual-energy x-ray absorptiometry (DXA) technology, though it uses the forearm as the measurement site. The Schick CDR Digital Dental X-ray Receptor Array (K933455) currently marketed by Schick Technologies uses a similar digital imaging array.
### DEVICE DESCRIPTION B.
The accuDEXA device is a Dual Energy X-Ray Absorptiometer (DEXA) device. The device is intended to calculate an index of bone mineral density in the middle finger of the non-dominant hand. By changing the high voltage on the X-ray tube, two energies are produced. Each of the two settings produces an image of the finger and each image is analyzed using various algorithms to produce a value of bone mineral density (BMD) and bone mineral content (BMC). These values are compared with a normative database, yielding a t-score and a z-score. The t-score is the number of standard deviations that the patient is above or below the mean of a reference sample of young healthy individuals. The zscore is the number of standard deviations that the patient is above or below the mean of a reference sample of individuals of the same age as the patient.
{5}------------------------------------------------
### INDICATIONS FOR USE C.
The accuDEXA is a dual-energy x-ray device indicated for use in measuring an index of the bone density of the middle finger of the nondominant hand (BMD). This BMD value is a relative indicator of bone density elsewhere in the body. The measurement is compared to similar measurements from patients of the same gender and ethnicity to provide t-scores (number of standard deviations above or below the mean for a reference group of young healthy individuals) and z-scores (number of standard deviations above or below the mean for a cohort of the same age). A t-score or a z-score may be used by a physician as one factor, in conjunction with other clinical indicators, to diagnose osteoporosis and other bone disorders. When a normative database of the same ethnicity and gender is not available, the BMD value may still be used to compare to the patient's own baseline value, for example in following the patient's response to treatment for osteoporosis.
Below normal bone density can be associated with a variety of conditions or disorders of bone. For example, the Society of Nuclear Medicine and American College of Nuclear Physicians has identified specific medical indications for measurement of a patient's bone mass: for patients with premenopausal oophorectomy, spontaneous menopause, or estrogen deficiency conditions; for treatment-related osteopenia; when the diagnosis of osteopenia is suggested or established by other means, such as x-ray; during long-term immobilization; for endocrinopathies known to be associated with osteopenia; for post-gastrectomy and other malabsorption states leading to osteopenia: during long-term corticosteroid therapy; for chronic renal disease, particularly in childhood or adolescence; and to monitor treatment programs for osteoporosis.
### D. SUBSTANTIAL EQUIVALENCE SUMMARY
The intended use for the accuDEXA and the predicate devices is the same. The accuDEXA Bone Densitometer has the same technological characteristics as the predicate devices. However, the descriptive characteristics may not be sufficiently precise to assure substantial equivalence. Therefore, Schick Technologies sponsored performance tests on the device and the predicate devices (see Section F, below), and a comparison of these data demonstrate that the accuDEXA Bone Densitometer is substantially equivalent to the predicate devices.
### E. TECHNOLOGICAL CHARACTERISTICS
See Device Description, above.
{6}------------------------------------------------
### TESTING F.
Schick Technologies has sponsored a study on cadavers that examines the precision and accuracy of the accuDEXA device. The study also directly compared the bone density values obtained with the predicate devices. This study showed that the accuDEXA measurements are very repeatable, even with repositioning. The very high correlation of the accuDEXA measurements with the ashed bone weights shows that the accuDEXA measurements are very accurate. The good correlation of the accuDEXA measurement with measurements with the two predicate devices demonstrates that the accuDEXA results for the middle finger are at least as good an indicator of BMD as the currently marketed predicate devices.
A reference data base has been developed in a clinical study which allows the calculation of t-scores and z-scores for each patient.
### G. CONCLUSIONS
Schick Technologies has demonstrated through its comparison of characteristics with the predicate devices and comparison of performance testing with the predicate devices that the accuDEXA Bone Densitometer is substantially equivalent to the predicate devices.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human profiles facing to the right.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC = 2 1997
Schick Technologies, Inc. c/o T. Whit Athey, Ph.D. Senior Consultant C.L. McIntosh & Associates, Inc. 12300 Twinbrook Parkway, Suite 625 Rockville, Maryland 20852
Re:
K971735 AccuDEXA Bone Densitometer Dated: September 11, 1997 Received: September 11, 1997 Regulatory class: II 21 CFR 892.1170/Procode: 90 KGI
Dear Dr. Athey:
::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::
We have reviewed your Section 510(k) notification of intent to market the device and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
W.Liau Yu
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# STATEMENT OF INDICATIONS FOR USE
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: __Schick accuDEXA Bone Densitometer______________________________________________________________________________________________________________________________
Indications For Use:
The accuDEXA is a dual-energy x-ray device indicated for use in estimating the bone density of the middle finger of the non-dominant hand (BMD). This BMD value is a relative indicator of bone density elsewhere in the body.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Siegman
Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological Devi 510(k) Number
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.