K971719 · Therox, Inc. · DQX · Nov 3, 1997 · Cardiovascular
Device Facts
Record ID
K971719
Device Name
THEROX INFUSION GUIDEWIRES
Applicant
Therox, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Nov 3, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TherOx Infusion Guidewires are intended for use in facilitating the placement of interventional devices in the cardiovascular and neurovascular system. TherOx Infusion Guidewires are designed for small vessel use. TherOx Infusion Guidewires are also intended to be used for the controlled selective infusion of physician-specified fluid agents before, during or after interventional procedures in the cardiovascular or neurovascular system. TherOx Infusion Guidewires are not intended for delivery of contrast agents. Fluid agents utilized with TherOx Infusion Guidewires should be fully prepared and used according to their manufacturer's instructions for use. TherOx Infusion Guidewires are not intended for neonatal or pediatric use.
Device Story
TherOx Infusion Guidewire is a 0.014-inch diameter guidewire (180 cm or 300 cm lengths) designed for small vessel navigation in cardiovascular and neurovascular systems. Device facilitates placement of interventional devices and enables controlled selective infusion of physician-specified fluid agents. Proximal end connects to a Tuohy-Borst adapter for fluid delivery; torquing knob allows distal tip orientation. Used by physicians in clinical settings under fluoroscopic guidance. Device provides a pathway for interventional tools and a conduit for therapeutic fluids, potentially improving procedural outcomes by allowing targeted delivery of agents. Does not deliver contrast agents.
Clinical Evidence
Bench testing (in vitro) evaluated tensile strength, torque response, tip deflection, joint integrity, and fluid infusion characteristics per FDA PTCA Catheter System Testing Guideline. Dimensional conformance verified. Animal studies (3 dogs, 2 pigs) assessed maneuverability, tip-shaping, radiopacity, and compatibility with PTCA balloons/ancillary products under fluoroscopy. Results showed performance comparable to commercially available guidewires with no observed vascular trauma or complications. Clinical studies confirmed the device functions effectively as an accessory to interventional devices.
Technological Characteristics
Materials include stainless steel and platinum coils, polyimide, and hypotube. Dimensions: 0.014" OD, 180 cm or 300 cm lengths. Features a distal diffuser section and proximal Tuohy-Borst connection. Sterilized via Ethylene Oxide (EtO). Biocompatibility compliant with ISO 10993-1 and Tripartite Agreement. Mechanical design allows torque-controlled tip orientation.
Indications for Use
Indicated for adult patients requiring placement of interventional devices or controlled selective infusion of physician-specified fluid agents in the cardiovascular and neurovascular systems. Not for neonatal or pediatric use. Not for delivery of contrast agents.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Advanced Cardiovascular Systems 0.014" Hi Torque Floppy Guidewire
Advanced Cardiovascular Systems Hi Torque Exchange Guidewire
{0}------------------------------------------------
# 510(k) SUMMARY
TherOx. Inc. 2025 Newport Blvd., Suite 200 Costa Mesa, CA 92627 Phone: (714) 645-4271 Fax: (714) 645-0208
NOV - 3 1997
Contact Paul Zalesky, Ph.D. Chief Operating Officer
Classification
Wire, Guide Percutaneous Catheter
#### DESCRIPTION
TherOx Infusion Guidewires are intended for use in facilitating the placement of interventional devices in the cardiovascular and neurovascular system. TherOx Infusion Guidewires are designed for small vessel use.
TherOx Infusion Guidewires are also intended to be used for the controlled selective infusion of physician-specified fluid agents before, during or after interventional procedures in the cardiovascular or neurovascular system.
TherOx Infusion Guidewires are not intended for delivery of contrast agents. Filuid agents utilized with TherOx Infusion Guidewires should be fully prepared and used according to their manufacturer's instructions for use.
TherOx Infusion Guidewires are not intended for neonatal or pediatric use.
The TherOx Infusion Guidewire has an outside diameter of .014 in. and is available 180 cm and 300 cm in length, and is intended to pass freely thru standard catheters that have a thru lumen of 016 in. or larger.
The 300 cm TherOx Exchange Guidewire differs only in its overall length from the 180 cm TherOx Infusion Guidewire. The distal 180 cm of the TherOx Exchange Guidewire is identical to the 180 cm TherOx Infusion Guidewire. Only the proximal hypotube has been extended by 120 cm to achieve the 300 cm length. This extended proximal portion always remains outside the body and facilitates the replacement of one interventional device for another.
{1}------------------------------------------------
A proximal fluid connection is secured by connecting the proximal open end to a simple Tuohy-Borst connection. The infusion guidewire may be torqued by rotation of the torquing knob to direct the orientation of the distal tip.
# TECHNOLOGICAL CHARACTERISTICS
This product is equivalent in intended use as well as design, composition and function to the legally marketed Advanced Cardiovascular Systems 0.014" Hi Torque Floppy Guidewire and Hi Torque Exchange Guidewire and the legally marketed MediTech Katzen Infusion Wire (K883880).
## SUMMARY OF STUDIES
### Performance Data
### In Vitro Tests:
Sample 2X sterilized (EtO) product was tested in vitro to assess all aspects of functionality and durability, evaluating each of the product's components. The product performance was also evaluated under simulated in vivo conditions via an aorta test fixture with indwelling PTCA catheter (see schematics enclosed in Exhibit 2). All bench testing was accomplished in accordance with the FDA's PTCA Catheter System Testing Guideline (Feb., 1989). Performance analysis assessed tensile testing, guidewire torque response, tip deflection and strength, joint integrity and strength, handling characteristics, fluid infusion characteristics, and product failure modes. Test results indicate that the Infusion Wire meets or exceeds its functional specification and is safe for its intended use.
Complete Infusion Wires were fabricated and inspected for dimensional conformance with the product specification. Length and O.D. measurements were made for the stainless steel coil, platinum coil, coil-to-coil screw ioint, distal tip, core-to-hypotube joint, overall length, hypotube length and O.D., diffuser section, bond length and O.D., and polyimide-tohypotube joint length and O.D. All dimensions were within the design specifications.
# RADIOPACITY
Radiopacity tests were performed during animal studies under fluoroscopic guidance. The TherOx Infusion Guidewire demonstrated adequate radiopacity when compared with commercially available guidewires and vascular catheters.
# BIOCOMPATIBILITY OF TEST MATERIALS
Tests for biocompatibility of materials for the TherOx Infusion Guidewire were performed to establish that the materials used in the device met the qualifications for short-term use in the vascular system to comply with the standards set forth in the guidance documents in Tripartite Agreement on Biocompatibility and ISO 10993-1.
{2}------------------------------------------------
In determining biocompatibility test design, testing was selected as deemed appropriate for the type of tissue/device interface and the duration of patient exposure.
Biocompatability testing was performed on sterile product.
TherOx feels that the biocompatibility test program used is appropriate and sufficient for a device that has less than 24 hours of tissue exposure.
TherOx feels the biocompatibility testing which was completed and passed, adequately addresses the issue of toxicological safety for this device's intended use.
### In Vivo Tests
In vivo animal studies were performed on 3 closed-chest dogs and 2 closed-chest pigs at 3 research centers to assess torque characteristics, maneuverability, tip-shaping, radiopacity, and compatibility with PTCA balloons and ancillary PTCA products.
The guidewire was directed to the heart by passing it through a guide catheter. Selected branches of the left and right coronary arteries were investigated. All studies were performed under fluoroscopic visualization.
A total of four commercially available guide catheters with multiple-tip configurations were utilized. Commercially distributed PTCA catheters manufactured by two companies, and of varying dimensions, were utilized to demonstrate compatibility with the Ther Infusion Guidewire.
During the animal studies the TherOx Infusion Guidewire was tested against commercially available guidewires. The TherOx Infusion Guidewire performed comparable to commercially available guidewires when assessing maneuverability, compatibility, radiopacity and tip-shaping properties.
No complications were observed as a result of utilizing the TherOx Infusion Guidewire. The investigators concluded that the TherOx Infusion Guidewire could be advanced into the coronary arteries and small vessels. The TherOx Infusion Guidewire performed with relative ease and no arteriographic or physiologic evidence of vascular trauma was observed.
### CLINICAL STUDY
Clinical studies demonstrated the TherOx Infusion Guidewire performed comparably as an accessory device, for commercially available interventional devices.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 3 1997
Paul J. Zalesky, Ph.D. TherOx, Inc. 2025 Newport Boulevard Costa Mesa, California 92627
K971719 Re : TherOx Infusion Guidewires Regulatory Class: II (two) Product Code: 74 DQX Dated: August 30, 1997 September 4, 1997 Received:
Dear Dr. Zalesky:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Paul J. Zalesky, Ph.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial in your of (x) pror device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callehan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number: K971719
Device Name: TherOx Infusion Guidewires
Indications for Use:
TherOx Infusion Guidewires are intended for use in facilitating the placement of interventional devices in the cardiovascular and neurovascular system. TherOx Infusion Guidewires are designed for small vessel use.
TherOx Infusion Guidewires are also intended to be used for the controlled selective infusion of physician-specified fluid agents before, during or after interventional procedures in the cardiovascular or neurovascular system.
TherOx Infusion Guidewires are not intended for delivery of contrast agents. Fluid agents utilized with TherOx Infusion Guidewires should be fully prepared and used according to their manufacturer's instructions for use.
TherOx Infusion Guidewires are not intended for neonatal or pediatric use.
Tua A. Rpp
(Division Sign-Off)
Division of Cardiovascular, Respiratory, co Neurological De 510(k) Number
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
### Concurrence of DCRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
✓
C
Over the Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.