HEMODIALYSIS VENOUS BLOOD TUBING SET OR SIMILAR TRADE NAME
Device Facts
| Record ID | K971687 |
|---|---|
| Device Name | HEMODIALYSIS VENOUS BLOOD TUBING SET OR SIMILAR TRADE NAME |
| Applicant | National Medical Care, Inc. |
| Product Code | FJK · Gastroenterology, Urology |
| Decision Date | Jul 29, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5820 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Hemodialysis Venous Blood Tubing Set is intended for use as the extracorporeal blood circuit during hemodialysis. It is intended for single use only. The Hemodialysis Venous Blood Tubing Set is indicated for use with conventional and high flux negative pressure hemodialyzer equipment.
Device Story
Hemodialysis venous blood tubing set functions as extracorporeal blood circuit during hemodialysis. Device connects patient to hemodialyzer equipment; facilitates blood flow during treatment. Used in clinical settings by healthcare professionals. Single-use disposable component. Output is the physical transport of blood within the circuit. Benefits patient by enabling safe extracorporeal blood circulation during dialysis therapy.
Clinical Evidence
Bench testing only. Data provided for functional performance, biocompatibility of blood-contacting materials, and integrity of bonded connections.
Technological Characteristics
Extracorporeal blood tubing set. Materials tested for biocompatibility. Bonded connections tested for integrity. Sterilization method and packaging consistent with predicate devices. No electronic, software, or algorithmic components.
Indications for Use
Indicated for use as an extracorporeal blood circuit during hemodialysis in patients requiring conventional or high flux negative pressure hemodialysis treatment. For single use only.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.