K971681 · Signal Medical Corp. · JDI · Dec 4, 1997 · Orthopedic
Device Facts
Record ID
K971681
Device Name
SMC FEMORAL HIP COMPONENT
Applicant
Signal Medical Corp.
Product Code
JDI · Orthopedic
Decision Date
Dec 4, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This femoral component of the hip is indicated for use in patients with: 1 - Osteoarthritis 2 Rheumatoid Arthritis 3 Traumatic Arthritis and when the use of a more conservative procedure has failed or is unacceptable.
Device Story
Signal Medical Corporation Femoral Component is a hip implant manufactured from cobalt chromium alloy (ASTM F-75). Available in four sizes (10.5mm to 15mm) with two collar angular settings (20 and 35 degrees). Features a circumferential collar for cortical arc loading and proximal compressive load distribution. Offered with or without a sintered beaded surface; non-sintered versions include a distal hole for a cement centralizer. Non-porous coated version is for cemented use only; porous sintered beaded version is for cemented or non-cemented press-fit use. Implanted by surgeons in clinical settings to replace the femoral head/neck, restoring joint function and reducing pain in patients with degenerative or traumatic arthritis.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design, material, and geometric comparison to legally marketed predicate devices.
Technological Characteristics
Material: Cobalt chromium alloy (ASTM F-75). Design: Femoral hip component, 10.5mm-15mm sizes. Features: Circumferential collar, 5.708-degree trunnion taper, optional sintered beaded surface. Cemented or press-fit application depending on surface coating.
Indications for Use
Indicated for patients with osteoarthritis, rheumatoid arthritis, or traumatic arthritis where conservative procedures have failed or are unacceptable.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
### 12971681 DEC =4 p
# 510K SUMMARY OF SAFETY AND EFFECTIVENESS
The Signal Medical Corporation Femoral Component of the Hip is manufactured with cobalt chromium alloy (ASTM F-75). The design is made available in four (4) sizes ranging from 10.5mm to 15mm. The stem portion of the femoral component is essentially identical to the DePuv AML Femoral Component (K852198) and the BioPro Horizon Femoral Component (K962769). The collar will be available in two different angular settings of 20 degrees to the horizontal and 35 degrees to the horizontal. The device will be offered with and without a sintered beaded surface, as depicted on the engineering drawings. The taper on the trunnion is a standard 5.708 degree angle as found on the Whiteside Biomechanics (K961625) and the StelKast Proform Cemented Femoral Component (K951083). The device has a circumfrential collar to load the bone on the cortical arc and distribute the compressive load of the implant as proximally as possible. The stems distal to the collar are essentially identical in geometry and size as the DePuy AML (K852198) and the BioPro Horizon (K962769). The non-sintered beaded surface device will have a 5mm hole in the distal tip of the implant to accommodate a cement centralizer from the StelKast Corporation (K951083). If further information is required please contact Dr. Louis Serafin, Signal Medical Corporation, 3315 Berry Drive, Lakeport, MI 48059, phone - 810-385-8302. This device is designed for cemented use only in the non-porous coated version. With the porous sintered beaded surface, the device is designed for both cemented and non-cemented press-fit use. This femoral component of the hip is indicated for use in patients with:
1 - Osteoarthritis
- 2 Rheumatoid Arthritis
- 3 Traumatic Arthritis and when the use of a more conservative procedure has failed or is unacceptable.
-3-
{1}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 4 1997
Louis A. Serafin, Jr., M.D. ·President Signal Medical Corporation 3315 Berry Drive Lakeport, Michigan 48059
K971681 Re : SMC Femoral Component Trade Name: Regulatory Class: II Product Codes: JDI and LPH Dated: September 5, 1997 Received: September 12, 1997
Dear Dr. Serafin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 - Louis A. Serafin, Jr., M.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D.
a M. Witten, Ph.D., M.D. Director
Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### 12971681 DEC =4 p
# 510K SUMMARY OF SAFETY AND EFFECTIVENESS
The Signal Medical Corporation Femoral Component of the Hip is manufactured with cobalt chromium alloy (ASTM F-75). The design is made available in four (4) sizes ranging from 10.5mm to 15mm. The stem portion of the femoral component is essentially identical to the DePuv AML Femoral Component (K852198) and the BioPro Horizon Femoral Component (K962769). The collar will be available in two different angular settings of 20 degrees to the horizontal and 35 degrees to the horizontal. The device will be offered with and without a sintered beaded surface, as depicted on the engineering drawings. The taper on the trunnion is a standard 5.708 degree angle as found on the Whiteside Biomechanics (K961625) and the StelKast Proform Cemented Femoral Component (K951083). The device has a circumfrential collar to load the bone on the cortical arc and distribute the compressive load of the implant as proximally as possible. The stems distal to the collar are essentially identical in geometry and size as the DePuy AML (K852198) and the BioPro Horizon (K962769). The non-sintered beaded surface device will have a 5mm hole in the distal tip of the implant to accommodate a cement centralizer from the StelKast Corporation (K951083). If further information is required please contact Dr. Louis Serafin, Signal Medical Corporation, 3315 Berry Drive, Lakeport, MI 48059, phone - 810-385-8302. This device is designed for cemented use only in the non-porous coated version. With the porous sintered beaded surface, the device is designed for both cemented and non-cemented press-fit use. This femoral component of the hip is indicated for use in patients with:
1 - Osteoarthritis
- 2 Rheumatoid Arthritis
- 3 Traumatic Arthritis and when the use of a more conservative procedure has failed or is unacceptable.
-3-
{4}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 4 1997
Louis A. Serafin, Jr., M.D. ·President Signal Medical Corporation 3315 Berry Drive Lakeport, Michigan 48059
K971681 Re : SMC Femoral Component Trade Name: Regulatory Class: II Product Codes: JDI and LPH Dated: September 5, 1997 Received: September 12, 1997
Dear Dr. Serafin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Louis A. Serafin, Jr., M.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D.
a M. Witten, Ph.D., M.D. Director
Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
510(k)Number(ifknown):________________________________________________________________________________________________________________________________________________________
SMC FEMORAL COMPONENT Device Name: Indications For Use:
> Osteoarthritis Rheumatoid Arthritis Traumatic Arthritis Where the use of a more conservative procedure has failed or is unacceptable
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) |
|-----------------------------------------|
| Division of General Restorative Devices |
| 510(k) Number | K971681 |
|---------------|---------|
|---------------|---------|
| Prescription Use | X | OR | Over-The- |
|----------------------|---|----|-----------|
| Counter Use | | | |
| (Per 21 CFR 801.109) | | | |
ﺟﺔ ﻳﺘﺴﻌﺪ ﻳﺎ ﻳ
(Optional Format
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.