K971650 · Data Critical Corp. · DXH · Dec 4, 1997 · Cardiovascular
Device Facts
Record ID
K971650
Device Name
RHYTHMSTAT XL
Applicant
Data Critical Corp.
Product Code
DXH · Cardiovascular
Decision Date
Dec 4, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2920
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The RhythmStat XL System is indicated for use in the diagnostic evaluation and recording of ECG waveforms from a cardiac event recorder. It is intended to provide remote access to ECG data by transtelephonic transmission from an ambulatory cardiac event monitor to the portable PSION palmtop computer with RhythmStat XL software, where the ECG waveforms are demodulated, recorded, displayed and stored.
Device Story
System records/reports ECG data from cardiac event recorders; inputs frequency-modulated acoustic waveforms via telephone; PSION Series 3 palmtop computer with proprietary software demodulates, displays, and stores ECG signals; physician reviews waveforms on screen; data/comments transmitted via modem to Data Critical Reporting Server; server generates printed report faxed to physician office; system enables remote access to ambulatory ECG data; facilitates clinical diagnostic evaluation; allows relaying of acoustic ECG signals between PSION units or to conventional receiving centers.
Clinical Evidence
Bench testing only. Performance compared to Micromedical BIOLOG and Instromedix LifeSigns Receiving Center. ECG recordings obtained via BIOLOG were played back to the PSION/RhythmStat XL system and printed by the Reporting Server. Amplitude and duration of the first ten QRS segments and total QRS duration were measured and compared against printouts from predicate devices. Results showed RhythmStat XL produced waveforms of equivalent quality.
Technological Characteristics
System components: cardiac event recorder (e.g., BIOLOG), PSION Series 3 palmtop computer, and Data Critical Reporting Server. Signal processing: frequency-modulated acoustic waveform demodulation. Connectivity: transtelephonic transmission, modem-based server communication, and fax reporting. Software: proprietary programs for signal reconditioning and transmission.
Indications for Use
Indicated for diagnostic evaluation and recording of ECG waveforms from ambulatory cardiac event monitors. Intended for remote access to ECG data via transtelephonic transmission.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
Predicate Devices
Hewlett-Packard M1490A Wireless Patient Data Communicator (Palm Vue System) (K945277)
{0}------------------------------------------------
K971650
## SECTION 2 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
DEC - 4 1997
Submitter's Name: Data Critical Corporation 100 North Broadway, Suite 2200 Oklahoma City, OK 73102 Telephone: (405) 236-4441 Contact Person: David E. Albert, MD, Chief Scientist
Date of Summary: May 5, 1997
Device Name: RhythmStat XL System
Device Classification: Telephone electrocardiograph transmitter and receiver (74 DXH); 21 CFR § 870.2920
Legally Marketed Devices To Which Equivalence Is Claimed: The legally marketed predicate devices are the Hewlett-Packard M1490A Wireless Patient Data Communicator (Palm Vue System) (K945277); the Instromedix® LifeSigns™ Receiving Center (a preAmendment device); and the Micromedical Industries BIOLOG (K915624).
Device Description: The RhythmStat XL System is a telephone electrocardiograph (ECG) receiving system used for recording and reporting of ECG data from a cardiac event recorder. The System has three main components:
- . A commercially available cardiac event recorder such as the Micromedical Industries BIOLOG or the Instromedix® Heart Card™, that conditions and transmits ECG data in the form of a frequency-modulated acoustic waveform to the PSION Series 3 palmtop computer:
- Two proprietary software programs, contained in the commercially available PSION Series 3 . palmtop computer, that 1) recondition the received signal format and display the signal to the physician on the computer screen, then 2) transmit the ECG data, along with the physician's comments, from the PSION palmtop computer to the Reporting Server via modem; and
- . The Data Critical Reporting Server that accepts the ECG data and commentary and generates a printed report that is automatically transmitted to the physician's office via fax.
Intended Use: The RhythmStat XL System is indicated for use in the diagnostic evaluation and recording of ECG waveforms from a cardiac event recorder. It is intended to provide remote access to ECG data by transtelephonic transmission from an ambulatory cardiac event monitor to the portable PSION palmtop computer with RhythmStat XL software, where the ECG waveforms are demodulated, recorded, displayed and stored.
{1}------------------------------------------------
## Descriptive Summary Of Technological Characteristics And Those Of Predicate
Devices: The RhythmStat XL System is indicated for use in the diagnostic evaluation and recording of ECG waveforms from a cardiac event recorder. It is intended to provide remote access to ECG data by transtelephonic transmission from an ambulatory cardiac event monitor to the portable PSION palmtop computer with RhythmStat XL software, where the ECG waveforms are demodulated, recorded, displayed and stored. The waveforms and accompanying information are then transmitted via modem to the Data Critical Reporting Server for generation of a printed report, which is then faxed to the physician's office. In addition, the PSION palmtop computer with RhythmStat XL software has the capability of relaying the most recently acquired ECG data as an acoustic signal to another PSION with RhythmStat XL software or to a conventional transtelephonic ECG receiving system such as the Instromedix LifeSigns Receiving Center.
The Hewlett-Packard M1490A Wireless Patient Data Communicator (PalmVue System) is intended to provide remote access to data from an HP patient monitoring system by transmitting data from a patient monitor through a Palm Vue dispatch station and paging service provider to a portable handheld Medical Palmtop personal computer. The HP system requires the use of a Hewlett Packard monitoring system and dispatch station. Palm Vue transmission software, an interface card, modem, and a paging service to support the system functions.
The Instromedix® LifeSigns™ Receiving Center is intended to receive, by telephone, ECG data from patients using ambulatory cardiac event monitors or pacer data transmitters. This tabletop device is line-powered and provides a printed record of ECG data; with an optional interface, it allows data to be communicated to a computer.
The Micromedical Industries BIOLOG is intended to acquire, record and store single-lead or 12-lead ECG waveforms via direct patient contact or through a patient cable with electrodes. This portable, hand-held device has an LCD screen, is battery-powered, and is capable of transtelephonic data transmission or receipt. It interfaces directly with the Micromedical Printer to provide a printed record of ECG data.
Performance Data: Testing was conducted by Data Critical Corporation on the RhythmStat XL System to compare its performance to that of two legally marketed predicate devices, the Micromedical Industries BIOLOG recorder and the Instromedix LifeSigns Receiving Center. Subjects were tested using the PSION palmtop computer with RhythmStat XL software and the BIOLOG recorder: additional testing was then conducted with three devices: the PSION with RhythmStat XL software, the BIOLOG recorder, and the Instromedix LifeSigns Receiving Center. ECG recordings were obtained with the BIOLOG, then played back to and recorded by the PSION/RhythmStat XL and printed by the RhythmStat XL Reporting Server. The same recording was then printed out by the predicate devices. the BIOLOG and the LifeSigns Receiving Center. For each ECG printout the amplitude and duration of the first ten QRS segments were analyzed and measured, along with the total duration of all QRS segments. The RhythmStat XL System demonstrated acceptable performance, producing ECG waveforms which were measurably comparable and of equivalent quality to the waveforms of the Micromedical and Instromedix products.
12
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with three human profiles incorporated into its design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Image /page/2/Picture/4 description: The image shows the date "DEC - 4 1997" in bold, black font. The month is abbreviated to "DEC", followed by a hyphen, the number 4, and the year 1997. The text is clear and easily readable against a white background.
Ms. Lisa S. Jones Data Critical Corporation c/o Devices For The Future, LLC 9223 Ilona Lane 77025-4218 Houston, Texas
K971650 Re : RhythmStat XL System Regulatory Class: II (two) 74 DXH Product Code: September 29, 1997 Dated: Received: October 1, 1997
Dear Ms. Jones:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such Existing major regulations affecting your device additional controls. can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. In addition, FDA may publish further announcements regulatory action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Ms. Lisa S. Jones
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K971650
## SECTION 2 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
DEC - 4 1997
Submitter's Name: Data Critical Corporation 100 North Broadway, Suite 2200 Oklahoma City, OK 73102 Telephone: (405) 236-4441 Contact Person: David E. Albert, MD, Chief Scientist
Date of Summary: May 5, 1997
Device Name: RhythmStat XL System
Device Classification: Telephone electrocardiograph transmitter and receiver (74 DXH); 21 CFR § 870.2920
Legally Marketed Devices To Which Equivalence Is Claimed: The legally marketed predicate devices are the Hewlett-Packard M1490A Wireless Patient Data Communicator (Palm Vue System) (K945277); the Instromedix® LifeSigns™ Receiving Center (a preAmendment device); and the Micromedical Industries BIOLOG (K915624).
Device Description: The RhythmStat XL System is a telephone electrocardiograph (ECG) receiving system used for recording and reporting of ECG data from a cardiac event recorder. The System has three main components:
- . A commercially available cardiac event recorder such as the Micromedical Industries BIOLOG or the Instromedix® Heart Card™, that conditions and transmits ECG data in the form of a frequency-modulated acoustic waveform to the PSION Series 3 palmtop computer:
- Two proprietary software programs, contained in the commercially available PSION Series 3 . palmtop computer, that 1) recondition the received signal format and display the signal to the physician on the computer screen, then 2) transmit the ECG data, along with the physician's comments, from the PSION palmtop computer to the Reporting Server via modem; and
- . The Data Critical Reporting Server that accepts the ECG data and commentary and generates a printed report that is automatically transmitted to the physician's office via fax.
Intended Use: The RhythmStat XL System is indicated for use in the diagnostic evaluation and recording of ECG waveforms from a cardiac event recorder. It is intended to provide remote access to ECG data by transtelephonic transmission from an ambulatory cardiac event monitor to the portable PSION palmtop computer with RhythmStat XL software, where the ECG waveforms are demodulated, recorded, displayed and stored.
{5}------------------------------------------------
## Descriptive Summary Of Technological Characteristics And Those Of Predicate
Devices: The RhythmStat XL System is indicated for use in the diagnostic evaluation and recording of ECG waveforms from a cardiac event recorder. It is intended to provide remote access to ECG data by transtelephonic transmission from an ambulatory cardiac event monitor to the portable PSION palmtop computer with RhythmStat XL software, where the ECG waveforms are demodulated, recorded, displayed and stored. The waveforms and accompanying information are then transmitted via modem to the Data Critical Reporting Server for generation of a printed report, which is then faxed to the physician's office. In addition, the PSION palmtop computer with RhythmStat XL software has the capability of relaying the most recently acquired ECG data as an acoustic signal to another PSION with RhythmStat XL software or to a conventional transtelephonic ECG receiving system such as the Instromedix LifeSigns Receiving Center.
The Hewlett-Packard M1490A Wireless Patient Data Communicator (PalmVue System) is intended to provide remote access to data from an HP patient monitoring system by transmitting data from a patient monitor through a Palm Vue dispatch station and paging service provider to a portable handheld Medical Palmtop personal computer. The HP system requires the use of a Hewlett Packard monitoring system and dispatch station. Palm Vue transmission software, an interface card, modem, and a paging service to support the system functions.
The Instromedix® LifeSigns™ Receiving Center is intended to receive, by telephone, ECG data from patients using ambulatory cardiac event monitors or pacer data transmitters. This tabletop device is line-powered and provides a printed record of ECG data; with an optional interface, it allows data to be communicated to a computer.
The Micromedical Industries BIOLOG is intended to acquire, record and store single-lead or 12-lead ECG waveforms via direct patient contact or through a patient cable with electrodes. This portable, hand-held device has an LCD screen, is battery-powered, and is capable of transtelephonic data transmission or receipt. It interfaces directly with the Micromedical Printer to provide a printed record of ECG data.
Performance Data: Testing was conducted by Data Critical Corporation on the RhythmStat XL System to compare its performance to that of two legally marketed predicate devices, the Micromedical Industries BIOLOG recorder and the Instromedix LifeSigns Receiving Center. Subjects were tested using the PSION palmtop computer with RhythmStat XL software and the BIOLOG recorder: additional testing was then conducted with three devices: the PSION with RhythmStat XL software, the BIOLOG recorder, and the Instromedix LifeSigns Receiving Center. ECG recordings were obtained with the BIOLOG, then played back to and recorded by the PSION/RhythmStat XL and printed by the RhythmStat XL Reporting Server. The same recording was then printed out by the predicate devices. the BIOLOG and the LifeSigns Receiving Center. For each ECG printout the amplitude and duration of the first ten QRS segments were analyzed and measured, along with the total duration of all QRS segments. The RhythmStat XL System demonstrated acceptable performance, producing ECG waveforms which were measurably comparable and of equivalent quality to the waveforms of the Micromedical and Instromedix products.
12
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with three human profiles incorporated into its design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Image /page/6/Picture/4 description: The image shows the date "DEC - 4 1997" in bold, black font. The month is abbreviated to "DEC", followed by a hyphen, the number 4, and the year 1997. The text is clear and easily readable against a white background.
Ms. Lisa S. Jones Data Critical Corporation c/o Devices For The Future, LLC 9223 Ilona Lane 77025-4218 Houston, Texas
K971650 Re : RhythmStat XL System Regulatory Class: II (two) 74 DXH Product Code: September 29, 1997 Dated: Received: October 1, 1997
Dear Ms. Jones:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such Existing major regulations affecting your device additional controls. can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. In addition, FDA may publish further announcements regulatory action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{7}------------------------------------------------
Page 2 - Ms. Lisa S. Jones
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
May 5,1997
Page 1 of 1
510(k) Number:
Device Name: Rhythm Stat XL System
Indications for Use: The RhythmStat XL System is indicated for use in the diagnostic evaluation and recording of ECG waveforms from a cardiac event recorder. It is intended to provide remote access to ECG data by transtelephonic transmission from an ambulatory cardiac event monitor to the portable PSION palmtop computer with RhythmStat XL software.
(Concurrence of CDRH, Office of Device Evaluation (ODE)
M. Ruy
(Division Sign-Off) Division of Cardiova and Neurological De 510(k) Number
Prescription Use (Per 21 CFR 801.109)
OR
Over-the-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.