← Product Code [DKB](/productcode/DKB) · K971600

# EMIT CALIBRATOR B LEVEL 1 (CUTOFF)/EMIT CALIBRATOR B LEVEL 2 (HIGH) (K971600)

_Behring Diagnostics, Inc. · DKB · Jun 25, 1997 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K971600

## Device Facts

- **Applicant:** Behring Diagnostics, Inc.
- **Product Code:** [DKB](/productcode/DKB.md)
- **Decision Date:** Jun 25, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3200
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

The Emit® Calibrator B Level 1 and Calibrator B Level 2 are used in the calibration of the Emite d.a.u.™ Amphetamine Class, Barbiturate (300 ng/mL cutoff), Benzodiazepine (300 ng/ml cutoff), Cocaine Metabolite (150 ng/mL cutoff), Methadone, Methaqualone, and Propoxyphen. A caws. The Emit Calibrator B Level 1 and Calibrator B Level 2 are also used in the calibration of the Emite II Barbiturate (300 ng/mL cutoff), Benzodiazepine (300 ng/mL cutoff), Methadone, Methaquelone, Opiates 300/2000, and Propoxyphene Assays.

## Device Story

Emit® Calibrator B Level 1 and Level 2 are liquid calibrators used in clinical laboratories to calibrate Emit® d.a.u.™ and Emit® II drug assays. These calibrators provide known concentrations of specific drugs (benzoylecgonine, d-amphetamine, methadone, methaqualone, morphine, oxazepam, propoxyphene, secobarbital) to establish a reference curve for the assay. By comparing patient sample signals against this calibrated curve, the assay quantifies drug levels. This process enables healthcare providers to detect drug presence or concentration, aiding in clinical decision-making regarding drug screening or monitoring. The modified version adds morphine at 2000 ng/mL (Level 1) and 4000 ng/mL (Level 2) to the existing formulation.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Liquid calibrator containing specific concentrations of drugs of abuse (benzoylecgonine, d-amphetamine, methadone, methaqualone, morphine, oxazepam, propoxyphene, secobarbital). Used for calibration of Emit® d.a.u.™ and Emit® II assays. No specific materials, energy sources, or software algorithms described.

## Regulatory Identification

A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Emit® Calibrator B Level 1 (cutoff) and Emit® Calibrator B Level 2 (high) ([K912729](/device/K912729.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a handwritten sequence of characters, which appears to be 'K971600'. The characters are written in a simple, somewhat stylized manner, with the numbers being more rounded. A horizontal line underlines the entire sequence, possibly indicating that it is a label or identifier.

Image /page/0/Picture/1 description: The image shows a logo with the word "BEHRING" in bold, uppercase letters. Below the word, there is a signature that appears to read "L. Behring". The logo is contained within a rectangular border.

JUN 2 5 1997

3403 Yerba Buena Road San Jose. CA 95135

## SUMMARY OF SAFETY AND EFFECTIVENESS Emit® Calibrator B Level 1 (cutoff) Emit® Calibrator B Level 2 (high)

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of SMDA 1990.

Behring Diagnostics Inc. is submitting this Premarket Notification, 510(k) to convey our intention to manufacture for commercial distribution a modified Emit® Calibrator B Level 1 (cutoff) and a modified Emit® Calibrator B Level 2 (high) used in the calibration of the Emit® d.a.u. ™ Amphetamine Class, Barbiturate (300 ng/mL cutoff), Benzodiazepine (300 ng/mL cutoff), Cocaine Metabolite (150 ng/mL cutoff), Methadone, Methaqualone, and Propoxyphene Assays. Emit® Calibrator B Level 1 (cutoff) and Emit® Calibrator B Level 2 (high) are also used in the calibration of the Emit® II Barbiturate (300 ng/mL cutoff), Benzodiazepine (300 ng/mL cutoff), Methadone, Methaqualone, Opiates 300/2000, and Propoxyphene Assays.

The modified Emit® Calibrator B Level 1 (cutoff) and Emit® Calibrator B Level 2 (high) are substantially equivalent to the currently marketed Emit® Calibrator B Level 1 (cutoff) and Emit® Calibrator B Level 2 (high) (K912729) with regard to intended use and overall performance characteristics. The most significant difference between the modified Emit® Calibrator B Level 1 (cutoff) and Emit® Calibrator B Level 2 (high) is the addition of morphine at 2000 ng/mL in the Calibrator B Level 1 (cutoff) and morphine at 4000 ng/mL in the Calibrator B Level 2 (high).

The modified Calibrators B contain the following drug concentrations:

|                 | Level 1<br>(ng/mL) | Level 2<br>(ng/mL) |
|-----------------|--------------------|--------------------|
| Benzoylecgonine | 150                | 3000               |
| d-Amphetamine   | 300                | 2000               |
| Methadone       | 300                | 1000               |
| Methaqualone    | 300                | 1500               |
| Morphine        | 2000               | 4000               |
| Oxazepam        | 300                | 1000               |
| Propoxyphene    | 300                | 1000               |
| Secobarbital    | 300                | 1000               |

In conclusion, Behring Diagnostics Inc. considers the modified Emit® Calibrator B Level 1 (cutoff) and the modified Emit® Calibrator B Level 2 (high) to be substantially equivalent to the existing Emit® Calibrator B Level 1 (cutoff) and Emit® Calibrator B Level 2 (high).

Syva® and Behring diagnostic products

ﻨﻤﺴﯿﺴﻪ

{1}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features an abstract eagle design with three tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUN 25 097

David Kolesar Administrator, Requlatory Affairs Behring Diagnostics ....................... P.O. Box 49013 San Jose, California 95161-9013

Re: K971600 Emit® Calibrator B Level 1 (Cutoff)/Emit® Calibrator B Level 2 (High) Requlatory Class: II Product Code: DKB Dated: April 29, 1997 Received: May 1, 1997

Dear Mr. Kolesar:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at t (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

sincerely yours,
Steven Sitman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known): _

Device Name: _ Emit® Calibrator B Level 1 (cuttof)

Emit® Calibrator B Level 2 (high)

Indications For Use:

The Emit® Calibrator B Level 1 and Calibrator B Level 2 are used in the calibration of the Emite d.a.u.™ Amphetamine Class, Barbiturate (300 ng/mL cutoff), Benzodiazepine (300 ng/ml cutoff), Cocaine Metabolite (150 ng/mL cutoff), Methadone, Methaqualone, and Propoxyphen. A caws. The Emit Calibrator B Level 1 and Calibrator B Level 2 are also used in the calibration of the Emite II Barbiturate (300 ng/mL cutoff), Benzodiazepine (300 ng/mL cutoff), Methadone, Methaquelone, Opiates 300/2000, and Propoxyphene Assays.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Clinical Laboratory I. svices 510(k) Number .

Prescription Use_

OR

Over-The-Counter Use_

(Division Sign-Off)
Division of Clinical Laboratory Levices
510(k) Number K97/600

---

**Source:** [https://fda-staging.innolitics.com/device/K971600](https://fda-staging.innolitics.com/device/K971600)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com