K971548 · Kinetic Concepts, Inc. · OMP · May 22, 1997 · General, Plastic Surgery
Device Facts
Record ID
K971548
Device Name
AMBUVAC
Applicant
Kinetic Concepts, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
May 22, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ambulatory V.A.C. device is indicated for patients who whould benefit from a suction device, particularly as the device use may promote wound healing, including for patients who would benefit from vacuum-assisted drainage and removal of infectious material or other fluids from wounds under the influence of continuous and/or alternating suction pressures.
Device Story
Ambulatory V.A.C. is a portable, battery-powered suction pump for wound therapy. Device applies continuous or intermittent negative pressure to wounds via a disposable collection canister and multi-lumen tubing. System includes a pump housing, vacuum pump, control system, and sterile disposables (canister with hydrophobic filter, wound dressing pad, adhesive film). Operated by clinicians or patients; pump carried in a pouch or placed on a surface. User selects therapy parameters (pressure, mode) via LCD and keypad. Microprocessor monitors negative pressure; provides feedback and alarms for canister full, dressing leak, or low battery. Output is negative pressure at the wound site, facilitating fluid drainage and wound healing. Device stores previous settings for continuity of care.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and design comparisons to the predicate device.
Technological Characteristics
Powered suction pump; oil-less diaphragm pump mechanism. Materials: translucent plastic canister, USP Class VI flexible tubing, polyurethane/PVA foam dressing, plastic adhesive drape. Energy: battery or low voltage adapter. Dimensions: 15cm x 11cm x 6cm; weight ~500g. Features: microprocessor control, LCD/keypad interface, 0.2 micron hydrophobic filter, negative pressure feedback loop. Sterilization: sterile single-use disposables.
Indications for Use
Indicated for patients requiring vacuum-assisted wound drainage and removal of infectious material or fluids to promote healing. Contraindicated for fistulas (excluding blind/incomplete), necrotic tissue (including bone osteomyelitis), and skin cancer.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
## 510 (k) Summary K971548
| I. Name of Device: | Ambulatory V.A.C |
|-------------------------------|--------------------------------------------|
| II. Classification Name: | Powered Suction Pump (per 21 CFR 878.4780) |
| | MAY 22 1997 |
| III. Substantial Equivalence: | V.A.C. Plus, 510(k) No. 945062 |
#### IV. Device Description:
The Ambulatory V.A.C. is a portable suction device that can be powered by a separate battery pack or a low voltage power adapter. When activated, it applies continuous or intermittent negative pressure to a wound, promoting wound healing and drainage of fluids and/or infectious materials from the wound into a disposable collection canister.
The Ambulatory V.A.C. consists of a small housing that contains a vacuum pump and control system with a containment chamber for a disposable wound fluid collection canister, and sterile, single-use disposables. Accessories included in the Ambulatory V.A.C. system are separate battery packs, a battery charger, an optional modem and a carrying belt and pouch. The disposables consists of the wound fluid collection canister with an integral hydrophobic filter and an attached length of multi-lumen tubing, a wound dressing pad, and an overlying adhesive film that adheres to the skin surrounding the wound.
The Ambulatory V.A.C. pump housing comprises of a compact plastic molding that weighs approximately 500 grams and measures 15 cm x 11 cm x 6 cm. When in use, the pump can be placed inside the pouch and carried on the belt or, when the patient is stationary, the pump can be placed on a table or flat surface.
The pump is turned ON by applying power either by the battery pack or the low voltage adapter. An LCD and membrane keypad on the top of the unit becomes active when power is applied. Operating parameters are selected using the LCD and keypad. The LCD is also used during operation to display the current negative pressure being applied and any alam condition messages.
The Ambulatory V.A.C. automatically stores the operating parameters from the previous therapy session. When the Ambulatory V.A.C. unit is initialized, the LCD displays the phrase "NEW PATIENT" at the top the display with "NO" and "YES" beneath it. "NO" is displayed over the left-arrow key and "YES" is displayed over the right-arrow key. Both arrow keys are located directly below the LCD. If the left-arrow key is pressed, for NO, the phrase "Will operate at previous settings" is displayed on the LCD. The operating parameters for the current therapy session remain as they were for the previous session. The THERAPY ON/OFF key, located to the left of the LCD, is then pressed to begin therapy.
{1}------------------------------------------------
If the right-arrow key is pressed for YES, indicating a new patient, the phrase "USING DEFAULTS" is displayed on the LCD. The user must specify whether default or user-selected parameters are to be used for the current therapy session. To use default parameters and begin operation, the THERAPY ON/OFF key is pressed and the unit will operate at default settings of 125mmHg with continuous therapy. To specify new parameters, the SELECT OPTIONS key, located to the right of the LCD, is pressed. The LCD displays the words "VAC TARGET =" followed by a number representing the current vacuum pressure setting. Pressing the right-arrow key increases the current setting in 25mmHg increments, while pressing the left-arrow key When the desired vacuum pressure is displayed, the decreases it in 25mmHg increments. SELECT OPTIONS key is again pressed to save the setting.
The LCD next displays a choice of two operation modes: 1) continuous and 2) intermittent. The word "CONTIN." appears over the left-arrow key and the word "INTERMIT." appears over the right-arrow key. The currently selected mode flashes on and off. To select continuous mode (if it is not already the currently selected mode), the user presses the left-arrow key and then the THERAPY ON/OFF key, to begin operation. If the continuous mode is already selected, the user simply presses the THERAPY ON/OFF key. To select the intermittent mode (if it is not already the currently selected mode), the user presses the right-arrow key and then the THERAPY ON/OFF key to being operation of therapy ON for 5 minutes and OFF for 2 minutes. The Ambulatory V.A.C. runs continuously until the THERAPY ON/OFF key is again pressed.
The Ambulatory V.A.C. pump has the following Alarm functions:
- · Canister Full
- · Dressing Leak
- · Battery Low
Each alarm condition is accompanied by an appropriate LCD message and an audible alarm that can be silenced by correcting the condition or pressing the therapy button.
#### SINGLE USE DISPOSABLES
- · wound fluid collection canister (50 ml) with integral filter and attached multilumen tube ( flexible medical tubing, meets USP Class VI criteria)
- polyurethane foam pad .
- plastic adhesive overlay .
- · Flexible tube to dressing connector
The wound fluid collection canister is a disposable chamber made of translucent plastic used for receiving fluids emitted from a wound. The canister includes a plastic cap that firmly locates the canister inside the pump housing. The canister is also graduated to allow the user to easily view and measure the volume of wound fluids within the canister.
Page 2
510(k) Summary Ambulatory V.A.C. Date Prepared: April 24, 1997
ﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
{2}------------------------------------------------
The collection canister has two parts which locate on spigots inside the pump housing. One spigot is used for applying a vacuum down the central bore of the multilumen tube. The other spigot locates in the second port which in turn is connected to the outer lumen of the tube and is used for sensing negative pressure at the wound site. A 0.2 micron hydrophobic membrane filter is mounted over the canister outlet. Fluid is drawn up through the central bore of the tube and into the canister. When the canister becomes full, the face of the filter is occluded and the negative pressure is reduced at the wound site which triggers the canister full alarm.
The disposable wound pad is fabricated from either polyurethane or PVA foam. In order for wound fluids to be communicated into the canister, the pad must be secured over the wound. This is accomplished by applying a plastic adhesive drape over the wound pad that gently adheres to the skin surrounding the wound. With the wound area sealed, vacuum pressure is transmitted to the wound when the suction pump is activated.
#### INDICATIONS FOR USE
The Ambulatory V.A.C. is a compact, portable device indicated for patients who would benefit from a suction device, particularly as the device use may promote wound healing, including for patients who would benefit from vacuum-assisted drainage and removal of infectious material or other fluids from wounds under the influence of continuous and/or alternating suction pressures.
## CONTRAINDICATIONS
The Ambulatory V.A.C. suction device is contraindicated for the following types of wounds:
- · Fistulas, excluding blind or incomplete fistulas,
- · Presence of necrotic tissue (including bone osteomyelitis),
- · Skin Cancer.
### PRECAUTIONS
- · Patients on anticoagulant, or difficult hemostatis
Page 3
510(k) Summary Ambulatory V.A.C. Prepared by: J. Harbour Kinetic Concepts, Inc. P.O. Box 659508 San Antonio, TX 78265
Date Prepared: April 24, 1997
.
A
1 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 -
{3}------------------------------------------------
| DESCRIPTION | Pump unit carried on belt with<br>separate battery pack | Hand carry, portable pump unit |
|---------------------------------------------------|-----------------------------------------------------------------------|----------------------------------------------------------------------|
| WEIGHT | 1.1 pounds | 10 pounds |
| POWER | Battery powered or low voltage<br>adapter, 115V/60Hz | 115V/60Hz |
| THERAPY SETTINGS | Range 0-125mmHg | Range 0-125mmHg |
| CONTINUOUS AND<br>INTERMITTENT<br>THERAPY SETTING | Custom settings range from 30<br>secs. to 10 min. for ON/OFF<br>times | Custom settings range from 30<br>secs. to 10 min for ON/OFF<br>times |
| EXHAUST FILTER | YES | YES |
| SETTINGS LOCKOUT | YES | YES |
| PUMP TYPE | Oil-less Diaphragm | Oil-less Diaphragm |
| CANISTER CAPACITY | 50 ml | 300 ml |
| MICROPROCESSOR<br>CONTROL | YES | YES |
| NEGATIVE PRESSURE<br>FEEDBACK | YES | YES |
| | | |
## DIFFERENCES BETWEEN CURRENT AND PREDICATED DEVICE
Page 4
510(k) Summary Ambulatory V.A.C. Prepared by: J. Harbour Kinetic Concepts, Inc. P.O. Box 659508 San Antonio, TX 78265 Date Prepared: April 24, 1997
{4}------------------------------------------------
# Ambulatory V.A.C. SPECIFICATIONS COMPARISON CHART
| | AmbuVAC | VACPLUS | CARE E-VAC |
|------------------------------|---------------------------------------|---------------------------------------|--------------------|
| MANUFACTURER/<br>DISTRIBUTOR | Kinetic Concepts, Inc. | Kinetic Concepts, Inc. | Aeros Instruments |
| MAX. PRESSURE | | 380mm/15" Hg | 550mm/22" Hg |
| INDICATOR | Gauge | Gauge | Gauge |
| WEIGHT | 2 lbs | 10 lbs | 9.7 lbs |
| FREE AIR<br>DISPLACEMENT | 2 liters/min | 9 liters/min | 27 liters/min |
| CYCLE | Intermittent and<br>Continuous Option | Intermittent and<br>Continuous Option | Continous |
| VACUUM<br>VARIABILITY | Yes, Manual | Yes, Manual | Yes, Manual |
| TIME METER | Yes | Yes | No |
| EXHAUST<br>FILTER | Yes | Yes | Unavailable |
| PUMP TYPE | Oil-less Diaphragm | Oil-less Diaphragm | Oil-less Diaphragm |
| POWER | 115V/60Hz | 115V/60Hz | 110V/60Hz |
| DIMENSIONS | 6"Wx4.3"Hx3"D | 11"Wx12"Hx7"D | 19"Wx11.5"Hx16.5"D |
| IN-TAKE HOSE | 0.25" ID | 0.25" ID | Unavailable |
| ILLUMINATED<br>POWER SWITCH | Yes | Yes | Yes |
Page 5
510(k) Summary Ambulatory V.A.C. Prepared by: J. Harbour Kinetic Concepts, Inc. P.O. Box 659508 San Antonio, TX 78265 Date Prepared: April 24, 1997
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The eagle is composed of three curved lines that form the shape of the bird's body and wings. The text is in all caps and is evenly spaced around the circle.
Public Health Service
APR - 7 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
KCI USA, Inc. % Ms. Christy Oviatt 6203 Farinon Drive San Antonio, Texas 78230
Re: K971548
Trade/Device Name: AmbuVAC Device Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: II Product Code: OMP Dated: April 24, 1997 Received: April 28, 1997
Dear Ms. Oviatt:
This letter corrects our substantially equivalent letter of May 22, 1997.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not
{6}------------------------------------------------
Page 2 - Ms. Christy Oviatt
limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
for Rob Romer
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
- - - - - - - - - - - - - - - - - -. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
: 【
Page / of /
510(k) Number (if known): K971548
Device Name: Ambulatory V.A.C.
Indications For Use:
・しと
The Ambulatory V.A.C. device is indicated for patients who whould benefit from a suction device, particularly as the device use may promote wound healing, including for patients who would benefit from vacuum-assisted drainage and removal of infectious material or other fluids from wounds under the influence of continuous and/or alternating suction pressures.
NOTE: The Indications for Use of the Ambulatory V.A.C. are identical to the predicate device, VAC PLUS, 510(k)945062.
IPLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of General Restorative Devices | |
| 510(k) Number | K971548 |
Prescription Use (Per 21 CFR 801. 109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
い
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.