ASSERACHROM TPA TEST KIT

K971519 · American Bioproducts Co. · GGP · Aug 19, 1997 · Hematology

Device Facts

Record IDK971519
Device NameASSERACHROM TPA TEST KIT
ApplicantAmerican Bioproducts Co.
Product CodeGGP · Hematology
Decision DateAug 19, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7290
Device ClassClass 2

Indications for Use

The ASSERACHROM® tPA test kit is intended for the quantitative determination of tissue-type plasminogen activator (tPA) antigen in citrated plasma by the sandwich technique of enzyme immunoassay (EIA) known as enzyme-linked immunosorbent assay (ELISA). The quantitative determination of tPA is generally performed during the assessment of the fibrinolytic system of a patient, for instance, following a thrombotic episode. In addition, it has been reported that a high antigenic level of tPA is associated with a greater risk of myocardial infarction or vascular cerebral accident. Finally, quantitative tPA assays are useful in the detection of hypofibrinolytic and hyperfibrinolytic states.

Device Story

ASSERACHROM® tPA test kit is an ELISA-based in vitro diagnostic assay for quantitative measurement of tPA antigen in citrated plasma. Principle: sandwich enzyme immunoassay using mouse monoclonal anti-tPA antibodies. Input: patient citrated plasma sample. Process: plasma incubated in antibody-coated microwells; captured tPA binds to second peroxidase-conjugated anti-tPA antibody; enzymatic reaction with orthophenylenediamine and hydrogen peroxide produces yellow color proportional to tPA concentration; reaction stopped with strong acid. Output: colorimetric intensity measured to determine tPA level. Used in clinical laboratory settings by trained personnel. Results assist clinicians in evaluating fibrinolytic system status, identifying hypo/hyperfibrinolytic states, and assessing risk for myocardial infarction or stroke.

Clinical Evidence

Bench testing only. Reported performance metrics: detection limit of 1 ng/ml; working range 1-50 ng/ml; intra-assay reproducibility < 6.5%; inter-assay reproducibility < 8%. No clinical trial data provided.

Technological Characteristics

Sandwich ELISA format. Components: mouse monoclonal anti-tPA antibody-coated plastic microwells, peroxidase-conjugated anti-tPA antibody, orthophenylenediamine substrate, hydrogen peroxide, and stop solution. Detection: colorimetric. Storage: 2°-8°C. Reagent stability: 18 months unopened.

Indications for Use

Indicated for quantitative determination of tPA antigen in citrated plasma to assess patient fibrinolytic system, particularly following thrombotic episodes. Useful for detecting hypofibrinolytic and hyperfibrinolytic states and assessing risk of myocardial infarction or vascular cerebral accident.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K97151.9 Page 33 of 33 ## AMERICAN BIOPRODUCTS COMPANY Premarket 510(k) Notification ASSERACHROM® tPA Test Kit ## AUG / 9 1997 #### VIII. Safety and Effectiveness Summary The ASSERACHROM® tPA test kit is intended for the quantitative determination of tissue-type plasminogen activator (tPA) antigen in citrated plasma by the sandwich technique of enzyme immunoassay (EIA) known as enzyme-linked immunosorbent assay (ELISA). Mouse monoclonal anti-tPA antibody is coated on the inside wall of a plastic microwell. The plasma to be tested is allowed to incubate in the microwell for two hours at room temperature, during which any tPA present is captured by the monoclonal antibody. Next, a second (and different) mouse monoclonal anti-tPA antibody coupled with peroxidase is added to the microwell for another 2-hour incubation at room temperature; this antibody-enzyme conjugate binds to another antigenic determinant of the tPA molecule that is already bound to the microwell in the first incubation step. The bound enzyme is revealed by its action on orthophenylenediamine in the presence of hydrogen peroxide to produce a yellow color; after addition of a strong acid to stop the enzymatic action, the intensity of the color developed bears a direct relationship to the tPA level initially present in the test plasma. The kit provides sufficient reagents to perform 96 tests in micro-ELISA plate format. Reagents in intact (unopened) kits remain stable for 18 months after their date of manufacture, when stored at 2°-8°C. Reconstituted reagent stabilities are as follows: Reagent ● (antibody-coated microwell strips) must be used immediately after opening of its package; Reagent @ (anti-tPA-Peroxidase), 24 hours at 2°-8°C; Reagent Oa (ortho-Phenylenediamine) dissolved together with Reagent Ob (Urea Peroxide), 1 hour at room temperature (18°-25°C); 1:10 diluted Reagent © (Dilution Buffer) and 1:20 diluted Reagent ® (Washing Solution), 15 days at 2°-8°C, when free of any contamination; both the Reagent ® (tPA Calibrator) and Reagent ® (tPA Control), 4 hours at 20°C. The ASSERACHROM® tPA reagent system has a detection limit of 1 ng/ml; a working range from 1 ng/ml up to the tPA level of Reagent ® (tPA Calibrator, up to 50 ng/ml); intra-assay reproducibility of < 6.5%; inter-assay reproducibility of < 8%. The ASSERACHROM® tPA test kit is claimed to be substantially equivalent to the predicate device TintElize® tPA test kit which is also an ELISA procedure for tPA available from Biopool (K934314). {1}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" surrounding it. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Andrew Loc B. Le, Ph.D. Director, Requlatory Affairs and Quality Assurance American Bioproducts Company Five Century Drive Parsippany; New Jersey 07054 - -------- AUG 1 9 1997 K971519/S2 Re : Trade Name: ASSERACHROM® tPA Test Kit Requlatory Class: II Product Code: GGP Dated: August 5, 1997 Received: August 7, 1997 Dear Dr. Le: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Fart 320) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {2}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Putman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page 1 of 1 ### Attachment A K971519 510(k) Number (if known Device Name: ___ ASSERACHROM® tPA Test Kit Indications for Use: The ASSERACHROM® tPA test kit is intended for the quantitative determination of tissue-type plasminogen activator (tPA) antigen in citrated plasma by the sandwich technique of enzyme immunoassay (EIA) known as enzyme-linked immunosorbent assay (ELISA). The quantitative determination of tPA is generally performed during the assessment of the fibrinolytic system of a patient, for instance, following a thrombotic episode. In addition, it has been reported that a high antigenic level of tPA is associated with a greater risk of myocardial infarction or vascular cerebral accident. Finally, quantitative tPA assays are useful in the detection of hypofibrinolytic and hyperfibrinolytic states. Umorre La. D. Montemayor Division Sign-Off (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number 971519 (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) (Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Clinical Laboratory Devices Prescription Use (Per 21 CFR 801.109) 11 OR 510(k) Number Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...