← Product Code [GCJ](/productcode/GCJ) · K971475

# SHIELDED SURGICAL TROCAR (K971475)

_Ethicon Endo-Surgery, Inc. · GCJ · Aug 21, 1997 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K971475

## Device Facts

- **Applicant:** Ethicon Endo-Surgery, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Aug 21, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The intended use of the New Device is to establish a path of entry for minimally invasive instruments. Instruments provided with a stopcock are intended for insufflation of the operative space when the trocar is in place.

## Device Story

ENDOPATH® Dilating Tip Trocar consists of obturator and sleeve sub-assemblies. Obturator features sharp flat blade or pyramidal tip with spring-loaded shield to protect internal structures post-entry. Sleeve includes inner gasket/flapper door and outer gasket to maintain pneumoperitoneum; some models include stopcock for insufflation and lever for desufflation. Integral threads on cannula exterior provide tissue retention. Used in OR by surgeons during minimally invasive procedures to provide access for instruments. Device facilitates entry and maintains operative space pressure, benefiting patient by enabling minimally invasive access.

## Clinical Evidence

Bench testing only. Pre-clinical laboratory evaluations demonstrated acceptable performance in device functionality, including insertion, tissue security, and maintenance of pneumoperitoneum.

## Technological Characteristics

Trocar sizes 3mm to 12mm diameter, 65mm to 150mm length. Features spring-loaded shield, gasket/flapper door seals, and external cannula threads for retention. Stopcock included on select models for insufflation. Mechanical device; no software or energy source.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

AUG 21

K97/475

## 510(k) Summary of Safety and Effectiveness Appendix A

| Statement                | Information supporting claims of substantial equivalence, as defined under the Federal<br>Food, Drug and Cosmetic Act, respecting safety and effectiveness is summarized<br>below. For the convenience of the Reviewer, this summary is formatted in accordance<br>with the Agency's final rule "...510(k) Summaries and 510(k) Statements..." (21 CFR<br>§807) and can be used to provide a substantial equivalence summary to anyone<br>requesting it from the Agency.                                                                |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>description    | The ENDOPATH® Dilating Tip Trocar consists of two main sub-assemblies:<br>an obturator sub-assembly and a sleeve sub-assembly.                                                                                                                                                                                                                                                                                                                                                                                                          |
|                          | The obturator consists of a sharp flat blade tip or pyramidal tip and a spring<br>loaded shield. The shield is designed to cover the flat blade or pyramidal tip to<br>protect internal structures from puncture or laceration once the abdominal or<br>thoracic cavity has been entered.                                                                                                                                                                                                                                               |
|                          | The sleeve subassembly has an inner gasket seal or flapper door seal and an<br>outer gasket seal to maintain pneumoperitoneum when instrumentation is<br>inserted and withdrawn through the cannula during a surgical procedure.<br>Some instruments are provided with a stopcock for insufflating the operative<br>space and a flapper door lever to allow quick desufflation. Integral threads<br>along the outside diameter of the cannula portion of the sleeve provide a<br>retention mechanism to stabilize the sleeve in tissue. |
|                          | The ENDOPATH® Dilating Tip Trocar shall be provided in a variety of sizes<br>from 3mm to 12mm in diameter and 65mm to 150mm in length.                                                                                                                                                                                                                                                                                                                                                                                                  |
| Intended use             | The intended use of the New Device is to establish a path of entry for<br>minimally invasive instruments. Instruments provided with a stopcock are<br>intended for insufflation of the operative space when the trocar is in place.                                                                                                                                                                                                                                                                                                     |
| Indications<br>statement | The ENDOPATH® Dilating Tip Trocar has application in general, thoracic,<br>gynecologic, or other minimally invasive surgical procedures to establish a<br>path of entry for minimally invasive instruments.                                                                                                                                                                                                                                                                                                                             |

Continued on next page

l

|

{1}------------------------------------------------

| Appendix A<br>Continued          | 510(k) Summary of Safety and Effectiveness,                                                                                                                                                                                                                                                                                                                                                                     |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>characteristics | The technological characteristics of the New Device are the same as the<br>Predicate Device.                                                                                                                                                                                                                                                                                                                    |
| Performance<br>data              | Pre-clinical laboratory evaluations were performed to ensure that the device<br>can be used as designed. The studies demonstrated acceptable performance to<br>the Predicate Device in mating the obturator with the sleeve, insertion into the<br>operative cavity, removal of the obturator from the sleeve, security of the<br>sleeve in tissue, and maintenance of pneumoperitoneum of the operative space. |
| Conclusion                       | Based on the 510(k) summaries and 510(k) statements (21 CFR §807) and the<br>information provided herein, we conclude that the New Device is substantially<br>equivalent to the Predicate Device under the Federal Food, Drug and<br>Cosmetic Act.                                                                                                                                                              |
| Contact                          | Ivan S. Placko<br>Project Manager<br>Regulatory Affairs Department<br>Ethicon Endo-Surgery, Inc.<br>4545 Creek Road<br>Cincinnati, Ohio 45242                                                                                                                                                                                                                                                                   |
| Date                             | April 21, 1997                                                                                                                                                                                                                                                                                                                                                                                                  |

-

:

1

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Ed Billips, RAC Project Manager, Regulatory Affairs Ethicon Endo-Surgery, Inc. 4545 Creek Road Cincinnati, Ohio 45242-2839

AUG 21 1997

Re: K971475 Trade Name: ENDOPATH® Dilating Tip Trocar Regulatory Class: II Product Code: GCJ Dated: July 30, 1997 Received: July 31, 1997

Dear Mr. Billips:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Ed Billips, RAC

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. . _ _ _ _ _ ..

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Sincerely yours,

coope

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Appendix B Indications for Use Statement

Following is the Indications for Use Statement: Statement

510(k) Number: K 971475

Device Name: ENDOPATH® Dilating Tip Trocar

## Indications for Use:

The ENDOPATH® Dilating Tip Trocar has application in abdominal, thoracic, and gynecologic minimally invasive surgical procedures to establish a path of entry for minimally invasive instruments.

Aciolo

on Sign-Off) IDivis Division of General Restorative Devi 510(k) Number

Prescription Use
(Per 21 CFR 801.109)

49

---

**Source:** [https://fda-staging.innolitics.com/device/K971475](https://fda-staging.innolitics.com/device/K971475)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
