← Product Code [GEX](/productcode/GEX) · K971459

# DERMALASE (K971459)

_Endo Technic Corp. · GEX · Jul 18, 1997 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K971459

## Device Facts

- **Applicant:** Endo Technic Corp.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Jul 18, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Use of the Dermalase ™ laser system is restricted to the incision, excision, ablation, vaporization and hemostasis of soft tissue

## Device Story

DermaLase is a surgical laser system consisting of a laser module and delivery system. The laser module houses the resonator cavity, power supplies, cooling system, and TTCS™ (Target Tissue Cooling System) for tissue irrigation. The system is operated via a control panel with a touch-sensitive membrane keypad and backlit display. Energy delivery options include a fiber optic system (contact or non-contact handpiece) or an articulated arm system (contact or non-contact point delivery handpiece). Used by clinicians in surgical or dermatological settings to perform soft tissue procedures including incision, excision, ablation, vaporization, and hemostasis. The device provides precise tissue interaction through controlled laser energy, with integrated cooling to manage thermal effects on surrounding tissue, facilitating surgical outcomes.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and performance specifications compared to predicate devices.

## Technological Characteristics

Surgical laser system; Er, Cr:YSGG laser source. Features include laser resonator cavity, closed-loop internal cooling, and TTCS™ (Target Tissue Cooling System) for irrigation. Delivery via fiber optic or articulated arm with contact/non-contact handpieces. Control interface uses touch-sensitive membrane keypad and backlit display. Microelectronic circuits for system control.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- CB Erbium/2.94 ([K961748](/device/K961748.md))
- Trilase 2940™ Erbium laser system ([K954013](/device/K954013.md))
- MCL 29 Dermablate™ ([K964128](/device/K964128.md))
- Elmer™ ([K952118](/device/K952118.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K971459

DermaLase™ BioLase Technology, Inc. April 18, 1997

# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION

REGULATORY AUTHORITY:

Safe Medical Devices Act of 1990, 21 CFR 807.92

JUL 1 8 1997

## COMPANY NAME/CONTACT:

- BioLase Technology, Inc. Company: 981 Calle Amanecer San Clemente, California 92673
- Contact: Mr. Andrew Kimmel BioLase Technology, Inc. 981 Calle Amanecer San Clemente, California 92673 (714) 361-1200 (714) 361-0204 Fax

Trade Name:

Common Name:

Surgical Laser

DermaLase™

Classification Name:

Surgical Laser System

Classification 79GEX Code:

EQUIVALENT DEVICES:

| CB Erbium/2.94                    | Continuum Biomedical, Inc. | K961748 |
|-----------------------------------|----------------------------|---------|
| Trilase 2940™ Erbium laser system | Schwartz Electro-Optics    | K954013 |
| MCL 29 Dermablate™                | Aesculap-Meditec           | K964128 |
| Elmer™                            | Biolase Technology, Inc.   | K952118 |

# DESCRIPTION OF THE DEVICE:

The DermaLase ™ laser system consists of two main components: the laser module and the delivery system.

The Laser Module contains the laser resonator cavity, power supplies, closed loop internal cooling system and TTCS™ tissue irrigation and cooling module. Microelectronics circuits and system Control Panel are housed inside the laser console as well.

{1}------------------------------------------------

The Control Panel provides access to all system controls. Actions are initiated through a touch sensitive membrane keypad. The display screen is back lighted for easy viewing.

BioLase currently offers two options for laser energy delivery: a fiber optic based delivery system and an articulated arm based delivery system. The fiber optic system utilizes either a contact or non-contact hand piece. The articulated arm based delivery system utilizes either a contact or non-contact point delivery hand piece.

## Indications for Use:

Use of the Dermalase ™ laser system is restricted to the incision, excision, ablation, vaporization and hemostasis of soft tissue

### SUBSTANTIAL EQUIVALENCE:

This Premarket Notification demonstrates that the laser systems described herein are substantially the same as the DermaLase 7M laser system.

Several different surgical, dental and dermatological laser systems and features are equivalent to those of the DermaLase™ laser unit. The internal components of the DermaLase 14 are almost identical to an earlier, currently 510(k) cleared Biolase erbium laser product, Elmer™ (K952118). The Elmer™ laser system produces laser light at the exact same wavelength, power and frequency as the DermaLase™ and includes Biolase's Target Tissue Cooling System™ for tissue cooling and irrigation.

Other laser manufacturers such as Schwartz Electro-Optics, Continuum Biomedical and Aesculap-Meditec are currently marketing products with similar performance specifications as the DermaLase ™ and for similar indications for use.

### CONCLUSION:

The DermaLase™ is Substantially Equivalent to several available surgical laser systems. As noted above, the laser wavelength and air and water fissue cooling and urrgation spray are equivalent to an earlier Biolase product, Elmer™ (K952118). Several other equivalent erbium laser devices have similar performance specifications, promotional materials and indications for use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract emblem that resembles an eagle or other bird-like figure, composed of several curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# JUL 1 8 1997

Mr. Andrew Kimmel BIOLASE Technology, Inc. ... 981 Calle Amanecer San Clemente, California 92673

Re: K971459

> Trade Name: DermaLase Er, Cr:YSGG Laser with Contact and Non-contact Handpieces Regulatory Class: II Product Code: GEX Dated: April 18, 1997 Received: April 21, 1997

Dear Mr. Kimmel:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801. and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 554-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

t.coellefa

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

..............................................................................................................................................................................

510 (K) Number (if known): K971459

DIGLINE THE TECH

Device Name:

DermaLase

Indications For Use: . . . . . . .

Use of the device is indicated for the incision, excision, ablation, vaporization and coagulation of dermatologic tissues including epidermal nevi, cheilitis, keloids, verrucae, skin tags, keratosis, scar revision, debulking of tumors, cysts, diagnostic biopsy and skin resurfacing.

Use of the device is further indicated for the incision, excision, vaporization and coagulation of soft tissue during general surgical applications where skin incision, tissue disection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors and lesions, tissue ablation and/or vessel coadulation may be indicated

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off)                     |         |
|-----------------------------------------|---------|
| Division of General Restorative Devices | K971459 |
| 510(k) Number                           |         |

Prescription Use (Per 21 CFR 801.109) க

Over-The-Counter Use - -

---

**Source:** [https://fda-staging.innolitics.com/device/K971459](https://fda-staging.innolitics.com/device/K971459)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
