K971459 · Endo Technic Corp. · GEX · Jul 18, 1997 · General, Plastic Surgery
Device Facts
Record ID
K971459
Device Name
DERMALASE
Applicant
Endo Technic Corp.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 18, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Use of the Dermalase ™ laser system is restricted to the incision, excision, ablation, vaporization and hemostasis of soft tissue
Device Story
DermaLase is a surgical laser system consisting of a laser module and delivery system. The laser module houses the resonator cavity, power supplies, cooling system, and TTCS™ (Target Tissue Cooling System) for tissue irrigation. The system is operated via a control panel with a touch-sensitive membrane keypad and backlit display. Energy delivery options include a fiber optic system (contact or non-contact handpiece) or an articulated arm system (contact or non-contact point delivery handpiece). Used by clinicians in surgical or dermatological settings to perform soft tissue procedures including incision, excision, ablation, vaporization, and hemostasis. The device provides precise tissue interaction through controlled laser energy, with integrated cooling to manage thermal effects on surrounding tissue, facilitating surgical outcomes.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and performance specifications compared to predicate devices.
Technological Characteristics
Surgical laser system; Er, Cr:YSGG laser source. Features include laser resonator cavity, closed-loop internal cooling, and TTCS™ (Target Tissue Cooling System) for irrigation. Delivery via fiber optic or articulated arm with contact/non-contact handpieces. Control interface uses touch-sensitive membrane keypad and backlit display. Microelectronic circuits for system control.
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of dermatologic tissues (e.g., epidermal nevi, cheilitis, keloids, verrucae, skin tags, keratosis, scar revision, tumor debulking, cysts, biopsy, skin resurfacing) and general surgical soft tissue applications (e.g., skin incision, tissue dissection, excision of external/internal tumors and lesions, partial/complete resection of internal organs, vessel coagulation).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K971459
DermaLase™ BioLase Technology, Inc. April 18, 1997
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
REGULATORY AUTHORITY:
Safe Medical Devices Act of 1990, 21 CFR 807.92
JUL 1 8 1997
## COMPANY NAME/CONTACT:
- BioLase Technology, Inc. Company: 981 Calle Amanecer San Clemente, California 92673
- Contact: Mr. Andrew Kimmel BioLase Technology, Inc. 981 Calle Amanecer San Clemente, California 92673 (714) 361-1200 (714) 361-0204 Fax
Trade Name:
Common Name:
Surgical Laser
DermaLase™
Classification Name:
Surgical Laser System
Classification 79GEX Code:
EQUIVALENT DEVICES:
| CB Erbium/2.94 | Continuum Biomedical, Inc. | K961748 |
|-----------------------------------|----------------------------|---------|
| Trilase 2940™ Erbium laser system | Schwartz Electro-Optics | K954013 |
| MCL 29 Dermablate™ | Aesculap-Meditec | K964128 |
| Elmer™ | Biolase Technology, Inc. | K952118 |
# DESCRIPTION OF THE DEVICE:
The DermaLase ™ laser system consists of two main components: the laser module and the delivery system.
The Laser Module contains the laser resonator cavity, power supplies, closed loop internal cooling system and TTCS™ tissue irrigation and cooling module. Microelectronics circuits and system Control Panel are housed inside the laser console as well.
{1}------------------------------------------------
The Control Panel provides access to all system controls. Actions are initiated through a touch sensitive membrane keypad. The display screen is back lighted for easy viewing.
BioLase currently offers two options for laser energy delivery: a fiber optic based delivery system and an articulated arm based delivery system. The fiber optic system utilizes either a contact or non-contact hand piece. The articulated arm based delivery system utilizes either a contact or non-contact point delivery hand piece.
## Indications for Use:
Use of the Dermalase ™ laser system is restricted to the incision, excision, ablation, vaporization and hemostasis of soft tissue
### SUBSTANTIAL EQUIVALENCE:
This Premarket Notification demonstrates that the laser systems described herein are substantially the same as the DermaLase 7M laser system.
Several different surgical, dental and dermatological laser systems and features are equivalent to those of the DermaLase™ laser unit. The internal components of the DermaLase 14 are almost identical to an earlier, currently 510(k) cleared Biolase erbium laser product, Elmer™ (K952118). The Elmer™ laser system produces laser light at the exact same wavelength, power and frequency as the DermaLase™ and includes Biolase's Target Tissue Cooling System™ for tissue cooling and irrigation.
Other laser manufacturers such as Schwartz Electro-Optics, Continuum Biomedical and Aesculap-Meditec are currently marketing products with similar performance specifications as the DermaLase ™ and for similar indications for use.
### CONCLUSION:
The DermaLase™ is Substantially Equivalent to several available surgical laser systems. As noted above, the laser wavelength and air and water fissue cooling and urrgation spray are equivalent to an earlier Biolase product, Elmer™ (K952118). Several other equivalent erbium laser devices have similar performance specifications, promotional materials and indications for use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract emblem that resembles an eagle or other bird-like figure, composed of several curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUL 1 8 1997
Mr. Andrew Kimmel BIOLASE Technology, Inc. ... 981 Calle Amanecer San Clemente, California 92673
Re: K971459
> Trade Name: DermaLase Er, Cr:YSGG Laser with Contact and Non-contact Handpieces Regulatory Class: II Product Code: GEX Dated: April 18, 1997 Received: April 21, 1997
Dear Mr. Kimmel:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801. and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 554-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
t.coellefa
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
..............................................................................................................................................................................
510 (K) Number (if known): K971459
DIGLINE THE TECH
Device Name:
DermaLase
Indications For Use: . . . . . . .
Use of the device is indicated for the incision, excision, ablation, vaporization and coagulation of dermatologic tissues including epidermal nevi, cheilitis, keloids, verrucae, skin tags, keratosis, scar revision, debulking of tumors, cysts, diagnostic biopsy and skin resurfacing.
Use of the device is further indicated for the incision, excision, vaporization and coagulation of soft tissue during general surgical applications where skin incision, tissue disection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors and lesions, tissue ablation and/or vessel coadulation may be indicated
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------|---------|
| Division of General Restorative Devices | K971459 |
| 510(k) Number | |
Prescription Use (Per 21 CFR 801.109) க
Over-The-Counter Use - -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.