VACUTAINER BRAND PLUS TUBE WITH EDTA ANTICOAGULANT(MULTIPLE)
K971449 · Becton Dickinson Vacutainer Systems · GIM · Jun 17, 1997 · Clinical Chemistry
Device Facts
Record ID
K971449
Device Name
VACUTAINER BRAND PLUS TUBE WITH EDTA ANTICOAGULANT(MULTIPLE)
Applicant
Becton Dickinson Vacutainer Systems
Product Code
GIM · Clinical Chemistry
Decision Date
Jun 17, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
The VACUTAINER® Brand PLUS (plastic) Tube with EDTA is an evacuated blood collection tube which provide a means of collecting, transporting, separating and processing blood in a plastic tube. When the tube is used together with VACUTAINER® Brand Needles and Holders, it is a closed system for the collection of venous blood with the same indications as described herein. Blood collected in a tube containing EDTA Anticoagulant is primarily used for clinical laboratory testing-hematology study using whole blood but may be used for other studies including such studies as testing for lead and FEP analyses, requiring whole blood as determined by the laboratory.
Device Story
Evacuated blood collection tube; PET plastic construction; spray-coated with dipotassium (K2) EDTA anticoagulant to prevent coagulation. Used in clinical laboratory settings by healthcare professionals for venous blood collection. Provides closed system when used with VACUTAINER needles/holders; reduces risk of tube breakage and exposure to blood compared to glass tubes. Facilitates transport and processing of whole blood specimens for hematology, lead, and FEP analysis. Enhances safety during disposal via incineration.
Clinical Evidence
Clinical evaluation compared VACUTAINER® Brand Plus plastic tubes to glass tubes for FEP and blood lead analysis. Results demonstrated analytical and statistical equivalence in performance and sample stability. Bench testing confirmed tube lead levels were <2 ug/L, supporting a 2.5 ug/L limit claim, well below CDC toxicity levels.
Technological Characteristics
PET (Polyethylene terephthalate) plastic evacuated tube; spray-coated with dipotassium (K2) EDTA; rubber stopper or HEMOGARD™ closure assembly. Designed for venous blood collection. No electronic components or software.
Indications for Use
Indicated for collection, transport, and processing of venous whole blood for clinical laboratory hematology studies, lead level testing, and FEP level testing. For professional use in clinical laboratory settings.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
VACUTAINER® Brand Tube with EDTA Anticoagulant (Pre-Amendment Device)
Reference Devices
VACUTAINER® Brand Plus Tube with EDTA Anticoagulant (K901449/A)
Submission Summary (Full Text)
{0}------------------------------------------------
K911449
JUN 17 1997
Image /page/0/Picture/2 description: The image shows the number 100 with a globe in the middle. The globe is black and white and shows the continents of North America, Europe, and Asia. The number 1 is on the left side of the globe, and the two zeros are on the right side of the globe. The numbers are black and white and have a striped pattern.
CELEBRATING THE FIRST ONE HUNDRED: 1897-1997
Becton Dickinson VACUTAINER Systems 1 Becton Drive Franklin Lakes, New Jersey 07417 (201) 847-4500
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
### I. General Information
This Summary of 510(k) Safety and Effectiveness information is being submitted in accordance with the requirements of the SMDA of 1990 and 21 CFR 807.92
Establishment:
- Becton Dickinson VACUTAINER Systems · Address: 1 Becton Drive Franklin Lakes, NJ 07417-1885
- · Registration Number: 2243072
- Contact Person: John A. Schalago . Regulatory Affairs Specialist Telephone no .: 201 - 847 - 6280 Facsimile no .: 201 - 847 - 4858
- April 16, 1997 · Date of Summary:
#### Device Name:
- VACUTAINER® Brand Plus Tube · Trade Name: with EDTA Anticoagulant
- Blood Specimen Collection Device . Classification Name :
- · Classification: Class II
- · Performance Standards: None Established under 514 of the Food, Drug and Cosmetic Act
## II. Safety and Effectiveness Information Supporting the Substantial Equivalence Determination · Device Description
The VACUTAINER® Brand PLUS Tube with EDTA is an evacuated plastic tube for collecting, transmitting and processing blood in a closed plastic tube. The blood collection tube consists of closure assembly, a plastic tube and EDTA coating (dipotassium, K2).
Image /page/0/Picture/20 description: The image shows the logo for Becton Dickinson. The logo is in black and white. The words "BECTON" and "DICKINSON" are stacked on top of each other.
{1}------------------------------------------------
The plastic tube is manufactured from PET (Polyethylene terepthalate Plastic and enhances user safety and disposal because of the reduced risk of tube breakage and incineration as a method of disposal. The standard closure assembly is a basic rubber stopper. The tube is also available with VACUTAINER Systems Hemogard™ Closure assembly which is designed to reduce user exposure to blood. The EDTA anticoagulant tube coating is spray coated in a dipotassium (K2) form. The EDTA prevents specimen coagulation.
### ● Intended Use
The VACUTAINER® Brand PLUS Tube with EDTA is an evacuated blood collection tube which provides a means of collecting, transporting and processing blood in a plastic tube. Blood collected in a tube containing EDTA anticoagulant is used primarily for clinical laboratory hematology studies but may be used for other whole blood specimens, including but not limited to lead level testing and FEP level testing, as determined by the laboratory.
### · Synopsis of Test Methods and Results
The VACUTAINER® Brand Plus Tube with EDTA anticoagulant, as described above, is a plastic tube with closure assembly (rubber stopper or HEMOGARD™ Closure). The plastic tube provides enhanced user safety and disposal because of the reduced risk of tube breakage and the use of incineration as a method of disposal. Breakage reduction and enhanced disposal studies and test results were described 510(k) Premarket Notification K901449/A which received FDA clearance on August 9, 1990.
Clinical evaluation of the VACUTAINER® Brand Plus tube with EDTA Anticoagulant was performed to demonstrate device equivalence for FEP and Blood Lead analyses. The VACUTAINER® Brand Plus Plastic Tube with EDTA was compared to VACUTAINER® Brand Glass Tube with EDTA, at initial time and extended tube storage. The VACUTAINER® Brand PLUS Tube with EDTA demonstrated analytically and statistically equivalent results and sample stability. Therefore, Clinical evaluation supports the Use of the VACUTAINER® Brand PLUS Tube with EDTA for FEP and Blood Lead analyses.
Additionally, bench testing of three different lots of VACUTAINER® Brand Plus Tube with EDTA was performed to determine the tube lead levels. As identified in the draft product labeling, the maximum limit is 2.5 ug/L (ppb). This maximum lead limit for the Plus plastic tube is significantly lower than the current Center for Disease Control and Prevention decision level of 10 ug/dL or 100 ug/L for toxicity in children. Twenty tubes from each of the three lots were tested according to internal test procedures. The Lead levels were evaluated using Atomic Absorption (AA). The calibration standards used to calibrate the AA had lead concentration levels of 2 and 4 ug/L. The results demonstrated that lead level of all tubes from the three different lots were below 2 ug/L and therefore, support the labeling claim of 2.5 ue/L. The test data and results are maintained on file at Becton Dickinson VACUTAINER Systems, Franklin Lakes, NJ.
{2}------------------------------------------------
### · Substantial Equivalence
Becton Dickinson VACUTAINER Systems believes that the VACUTAINER® Brand PLUS Tube with EDTA with the expanded Indications for Use is substantially equivalent to a commercially available blood collection tube. Clinical testing, as described in this premarket notification, demonstrates equivalent performance and effectiveness and supports the determination of substantial equivalence. The predicate devices, manufacturer, K number and clearance date are identified below:
| Manufacturer | Predicate Device | K-Number | Clearance Date |
|-----------------------|------------------------------------------------------|-------------------|-----------------------------------------------------------------------------------------------------------------------------|
| VACUTAINER<br>Systems | VACUTAINER®<br>Brand Tube with<br>EDTA Anticoagulant | Not<br>Applicable | Pre-Amendment Device and,<br>therefore, exempt from premarket<br>notification requirements according to<br>the MDA of 1976. |
Also included in this premarket notification, is a description of an incremental packaging modification which occurred prior to the submission of this premarket notification. VACUTAINER Systems determined that the incremental modification did not significantly affect the safety or efficacy of the VACUTAINER® Brand Tube and Brand PLUS Tube, and therefore, did not require a 510(k) premarket notification.
Oshkoshlago
ohn A. Schalago Regulatory Affairs Specialist Regulatory Affairs Department April 16, 1997 Date
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle's head and neck. The eagle's head is facing to the right.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUN 17
John A. Schalago Requlatory Affairs Specialist Becton Dickinson VACUTIANER Systems 1 Becton Drive Franklin Lakes, New Jersey 07417
Re: K971449 VACUTAINER® Brand PLUS (Plastic) Blood Collection Tube with EDTA Anticoagulant Requlatory Class: II Product Code: GIM Dated: April 16, 1997 Received: April 21, 1997
Dear Mr. Schalaqo:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
* If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: VACUTAINER® Brand PLUS Tube with EDTA Anticoagulant
Indications for Use:
The VACUTAINER® Brand PLUS (plastic) Tube with EDTA is an evacuated blood collection tube which provide a means of collecting, transporting, separating and processing blood in a plastic tube. When the tube is used together with VACUTAINER® Brand Needles and Holders, it is a closed system for the collection of venous blood with the same indications as described herein.
Blood collected in a tube containing EDTA Anticoagulant is primarily used for clinical laboratory testing-hematology study using whole blood but may be used for other studies including such studies as testing for lead and FEP analyses, requiring whole blood as determined by the laboratory.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use <span style="text-decoration: underline;">✓</span> OR | Over-the-Counter Use <span style="text-decoration: underline;">_____</span> |
|------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| (Per 21 CFR 801.109) | (Optional Format 1-2-96) |
| | (Division Sign-Off) |
| | Division of Clinical Laboratory Devices |
| 510(k) Number | K971449 |
510(K) Premarket Notification April 16, 1997 Page 45Page 45
(K) Premarket Notification VACUTAINER® Brand PLUS Tube with EDTA
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.