INTERSEAL ACETABULAR SCREW HOLE PLUG

K971429 · Wrightmedicaltechnologyinc · LPH · Jul 9, 1997 · Orthopedic

Device Facts

Record IDK971429
Device NameINTERSEAL ACETABULAR SCREW HOLE PLUG
ApplicantWrightmedicaltechnologyinc
Product CodeLPH · Orthopedic
Decision DateJul 9, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2

Indications for Use

The INTERSEAL® Acetabular Cup is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2) inflammatory degenerative joint disease such as rheumatoid arthritis; 3) correction of functional deformity; 4) revision procedures where other treatments or devices have failed; and, 5) treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

Device Story

INTERSEAL® Acetabular Screw Hole Plug is a mechanical accessory for INTERSEAL® acetabular shells. Device functions as a filler for unused screw holes to prevent wear debris migration. Machined from ultra-high molecular weight polyethylene (UHMWPE); one-piece, snap-in design. Surgeon inserts plugs into acetabular cup screw holes during total hip arthroplasty implantation. Device provides a physical barrier; no active components or software.

Clinical Evidence

Bench testing only. Tensile properties of EtO-sterilized components were compared to unsterilized UHMWPE bar stock and found to meet or exceed ASTM F 648 requirements.

Technological Characteristics

Material: Ultra-high molecular weight polyethylene (UHMWPE) conforming to ASTM F 648. Design: One-piece, snap-in plug. Sterilization: Ethylene oxide (EtO). No software, electronics, or energy sources.

Indications for Use

Indicated for skeletally mature patients undergoing total hip arthroplasty for non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity correction, revision procedures, or treatment of nonunion, femoral neck, and trochanteric fractures of the proximal femur with head involvement.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a stylized letter "W" logo. The "W" is formed by three thick, black, diagonal lines against a white background. The lines are arranged to create the shape of a "W", with the middle line slightly offset to the right. A registered trademark symbol is located at the bottom right corner of the logo. Image /page/0/Picture/1 description: The image shows the logo for Wright Medical Technology, Inc. The logo features the company name in bold, uppercase letters, with the words "MEDICAL TECHNOLOGY, INC." appearing in a smaller font size below. The address "5677 AIRLINE ROAD ARLINGTON, TN 38002" and phone number "901-867-9971" are printed below the company name. K97.1429 ### 510(k) Summary JUL - 9 1997 Contact Person: Cristie Manuel Date Prepared: April 17, 1997 | Trade/Common Name: | INTERSEAL® Acetabular Screw Hole Plug | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------| | Classification Name: | Hip joint, metal/polymer/metal semi-constrained, porous-coated<br>uncemented prosthesis | | Predicate Devices: | INTERSEAL® Apical Hole Plug manufactured by Wright Medical<br>Technology, Inc.; and Opti-Fix Cement Spacer Pod manufactured<br>by Smith and Nephew. | This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92. ### Description The INTERSEAL® Acetabular Screw Hole Plug provides an optional filler for INTERSEAL® acetabular shell screw holes to prevent a pathway for wear debris through unused screw holes holes during placement of the acetabular shell without screws. The INTERSEAL® Acetabular Screw Hole Plug is machined from ultra high molecular weight polyethylene (UHMWPE) conforming to ASTM F 648. The one-piece, snap-in design will be offered sterile in a pack of up to four plugs for insertion into the screw holes of the INTERSEAL® Acetabular Cup by the surgeon during implantation. ### Indications This device is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1) noninflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia: 2) inflammatory degenerative joint disease such as rheumatoid arthritis: 3) correction of functional deformity; 4) revision procedures where other treatments or devices have failed; and, 5) treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. ### Testing Summary Tensile properties of the EtO-sterilized components and control unsterilized UHMWPE bar stock meet or exceed the minimum requirements of ASTM F 648. {1}------------------------------------------------ -- Image /page/1/Picture/1 description: The image is a black and white circular logo for the Department of Health & Human Services USA. The logo features the department's emblem, which is a stylized representation of a human figure embracing a bird. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged around the emblem in a circular fashion. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Christie Manuel Requlatory Affairs Associate Wright Medical Technology, Inc. 5677 Airline Road 38002 Arlington, Tennessee JUL - 9 1997 ... .. K971429 Re : INTERSEAL® Acetabular Screw Hole Plug Regulatory Class: II Product Code: LPH Dated: April 17, 1997 Received: April 18, 1997 Dear Ms. Manuel: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {2}------------------------------------------------ Fage 2 - Ms. Christie Manuel This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, fo colly Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ | C. | Indications for Use of the Device | |----|-----------------------------------| |----|-----------------------------------| | 510(k) Number (if known): | 9711429 | |---------------------------|---------| |---------------------------|---------| Device Name: INTERSEAL® Acetabular Screw Hole Plugs, Product Line Addition to the INTERSEAL® Acetabular Cup Indications for Use: The INTERSEAL® Acetabular Cup is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: - 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; - 2) inflammatory degenerative joint disease such as rheumatoid arthritis; - correction of functional deformity; 3) - revision procedures where other treatments or devices have failed; 4) and, - ર) treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. (Please do not write below this line-continue on another page if needed) * * * * * ## Concurrence of CDRH, Office of Device Evaluation (ODE) f. colles (Division Sign-Off) Division of General Restorative Devices 510(k) Number. Prescription Use_X (Per 21 CFR 801.109) Or Over-the-Counter Use (Optional Format 1-2-96) # 0000009
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...