The Osteonics® Acetabular Wedges will provide the orthopedic surgeon with a prosthetic alternative to structural allografting in cases of segmental acetabular deficiencies. The Osteonics® Acetabular Wedges, because they are accessory components to an existing total hip replacement system, are subject to the general indications and contraindications of their mating acetabular shell components, as follows: General Indications for Total Hip Replacement Components: Painful, disabling joint disease of the hip resulting from degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis. Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure. Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results. Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum. Indications Specific to the Acetabular Wedges: As an alternative to structural allograft in cases of superior and superior/posterior segmental acetabular deficiencies.
Device Story
Prosthetic acetabular wedge system; accessory component for total hip replacement. Wedge-shaped implants (8, 12, 16mm thickness) with machined interior steps and AD-HA coating. Intraoperatively assembled to mating acetabular shells using PMMA bone cement. Provides alternative to structural allograft for segmental acetabular deficiencies. Used by orthopedic surgeons in clinical/OR settings. Output is a reconstructed acetabular construct, affixed to the acetabulum via cemented or cementless methods (per mating shell specifications). Benefits patient by addressing bone stock deficiencies during hip arthroplasty.
Clinical Evidence
Bench testing only. Specimens successfully endured 10 million cycles of physiologically relevant loading with no signs of interface failure.
Technological Characteristics
Materials: Commercially pure titanium substrate with AD-HA coating. Design: Wedge/crescent shape, 8-16mm thickness, machined interior steps, 6.5mm bone screw holes. Fixation: PMMA bone cement for wedge-to-shell assembly; shell-to-acetabulum fixation via cemented or cementless methods. No software or electronic components.
Indications for Use
Indicated for patients with painful, disabling hip joint disease (degenerative, rheumatoid, post-traumatic arthritis, or avascular necrosis), revision of failed hip procedures, or cases requiring reconstruction due to poor bone stock or acetabular deficiencies. Specific use as an alternative to structural allograft for superior and superior/posterior segmental acetabular deficiencies.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
MARS (Modular Acetabular Reconstruction System)
Osteonics® Secur-Fit™ HA Acetabular Shells
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 2 2 1997
Osteonics® Acetabular Wedge System
510(k) Summary
5971422
期間
# 510(k) Summary Osteonics® Acetabular Wedge System
# Submission Information
Name and Address of the Sponsor:
Contact Person:
Date of Summary Preparation:
Device Identification
Proprietary Name:
Common Name:
Classification Name and Reference:
Allendale, NJ 07401-1677
Osteonics Corporation
Terry Sheridan Regulatory Affairs Specialist
April 15, 1997
59 Route 17
Osteonics® Acetabular Wedge System
Acetabular augmentation devices for total hip replacement acetabular components
Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented or Non-Porous Uncemented Prosthesis 21 CFR §888.3353
# Predicate Device Identification
The Osteonics® Acetabular Wedges are substantially equivalent to the following competitive and/or Osteonics devices, which have previously been determined substantially equivalent by FDA:
- The MARS (Modular Acetabular Reconstruction System) Device .
- The Osteonics® Secur-Fit™ HA Acetabular Shells ●
ﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
{1}------------------------------------------------
4 7
เ
# Device Description
The Osteonics® Acetabular Wedges are characterized by the following design features.
- Basic Wedge Shape: The Osteonics® Acetabular Wedges feature a basic wedge (or . crescent) shape that conforms to the outer geometry of the mating acetabular shell.
- Size range: The Osteonics Acetabular Wedges are available in a wide range of outer . diameter sizes.
- Thickness Options: The Osteonics® Acetabular Wedges come in thicknesses of 8, 12, . or 16mm.
- Interior Normalizations: The interior face of the acetabular wedges features machined . steps that correspond to the normalizations featured on the exterior of the mating shells.
- Screw Hole Availability: The Osteonics® Acetabular Wedges feature 6.5mm bone . screw holes that are compatible with commercially available Osteonics 6.5mm Cancellous Bone Screws.
- Surface Treatments: The exterior surface of the Osteonics® Acetabular Wedges . features Osteonics' AD-HA Coating.
## Intended Use:
### General
The Osteonics® Acetabular Wedges will provide the orthopedic surgeon with a prosthetic alternative to structural allografting in cases of segmental acetabular deficiencies.
### Indications
The Osteonics® Acetabular Wedges, because they are accessory components to an existing total hip replacement system, are subject to the general indications and contraindications of their mating acetabular shell components, as follows:
# General Indications for Total Hip Replacement Components
- Painful, disabling joint disease of the hip resulting from degenerative arthritis, rheumatoid . arthritis, post-traumatic arthritis or late stage avascular necrosis.
- . Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques ● are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as . indicated by deficiencies of the acetabulum.
Indications Specific to the Acetabular Wedges
- As an alternative to structural allograft in cases of superior and superior/posterior . segmental acetabular deficiencies.
{2}------------------------------------------------
11 11:4
2 2
# Assembly
The Osteonics® Acetabular Wedges are intended for intraoperative assembly to mating acetabular shells. The operating room staff will use polymethyl methacrylate (PMMA) bone cement to affix a selected acetabular wedge to its mating acetabular shell.
### Fixation
The assembled acetabular wedge-to-shell construct can be affixed to the acetabulum either with or without bone cement, according to the specific fixation methods already determined substantially equivalent by FDA for the commercially available, mating acetabular shell components.
# Statement of Technological Comparison:
The substantial equivalence of the Osteonics Acetabular Wedges is supported by a comparison of the subject device to the above-cited predicate devices.
A comparison of the subject and predicate devices in terms of intended use, materials, and design follows.
### Intended Use
Both the subject devices and the predicate MARS devices are intended to provide an alternative to structural allograft in cases of segmental acetabular deficiencies. Use of the subject wedges in an acetabular construct that can be affixed to the acetabulum either with or without bone cement is predicated by the mating acetabular shells, which have already been determined Substantially Equivalent by FDA for either cemented or cementless fixation.
### Materials
#### Substrate materials
The subject device and the predicate devices feature substrates manufactured from commercially pure titanium.
#### Coating materials
The Osteonics® Acetabular Wedges feature Osteonics' AD-HA coating. This coating has already been extensively characterized by Osteonics and has already been thoroughly reviewed by FDA on several previous occasions.
#### Design
The subject devices, like the predicate MARS devices, are modular wedges which come in a variety of widths and thicknesses. Both device systems are intended for intraoperative assembly to an otherwise hemi-spherical shell design. The subject devices are assembled to their mating shells via PMMA bone cement, while the predicate MARS devices are assembled
- 144. 2
{3}------------------------------------------------
#### Osteonics® Acetabular Wedge System
42
1 - 18
to their mating shells via a taper and screw mechanism. This difference, however, is not significant because the assembly method of the subject devices has been shown to withstand anticipated physiologic loading conditions.
# Performance Data:
The Osteonics® Acetabular Wedges have been tested to ensure that the wedge-to-shell assembly is strong enough to withstand anticipated in-vivo loading conditions. All specimens successfully endured 10 million cycles of physiologically relevant loading, with no signs of impending wedge/shell interface failure.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract eagle design with three wing-like shapes. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 2 1997
Robert A. Koch, J.D. Director, Requlatory/Legal Affairs Osteonics Corporation ……… 59 Route 17 Allendale, New Jersey 07401-1677
K971422 Re : Acetabular Wedge System Requlatory Class: II Product Codes: MEH and LPH Dated: April 15, 1997 Received: April 17, 1997
Dear Mr. Koch:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). This decision is based on consideration of the specific design of stem and coating composition detailed in this application. You may, therefore, market the device, subject to the general controls provisions of the Act and the following limitation:
You may not label or in any way promote these devices . I for biological attachment, enhanced clinical or radiographic performance, enhanced fixation and/or longterm stable fixation. The data presented support equivalence with no additional claims over a conventional press-fit hip prosthesis (i.e., mechanical interlock, only).
Additional limitations for more specific claims of safety and effectiveness may be forthcoming. Should additional limitations be applied you will be contacted in writing to inform you of the additional labeling limitations.
{5}------------------------------------------------
#### Page 2 - Robert A. Koch, J.D.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
You may market your device under the above limitations as class II devices. These devices would be considered not substantially equivalent to a legally marketed predicate device if labeled with other intended uses and/or claims of safety or effectiveness. Any other intended uses or claims may cause the device to be classified into Class III under Section 513 (f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing.
Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to A substantially equivalent determination assumes 895. compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact
{6}------------------------------------------------
#### Page 3 - Robert A. Koch, J.D.
the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance …… at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# 510(k) Number (if known): K ヨニ | 4 2 2
### Device Name: Osteonics® Acetabular Wedge System
#### Indications For Use : . . . . . . . . .
The Osteonics® Acetabular Wedges, because they are accessory components to an existing total hip replacement system, are subject to the general indications of their mating acetabular shell components, as follows:
### General Indications for Total Hip Replacement Components
- Painful, disabling joint disease of the hip resulting from degenerative arthritis, rheumatoid . arthritis, post-traumatic arthritis or late stage avascular necrosis.
- . Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- . Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- . Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
### Indications Specific to the Acetabular Wedges
- · As an alternative to structural allograft in cases of superior and superior/posterior segmental acetabular deficiencies.
The assembled acetabular wedge-to-shell construct can be affixed to the acetabulum either with or without bone cement, according to the specific fixation methods already determined substantially equivalent by FDA for the commercially available, mating acetabular shell components.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|----------------------------------------------------------------|--------------------------|
| (Division Sign-Off)<br>Division of General Restorative Devices | |
| 510(k) Number | K971422 |
| Prescription Use | OR Over-The-Counter Use |
| (Per 21 CFR 801.109) | (Optional Format 1-2-96) |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.