NIHON KOHDEN TEC-7531 CARDIOLIFE PORTABLE DEFIBRILLATOR AND ACCESSORIES
K971355 · Nihon Kohden America, Inc. · LDD · Jul 15, 1997 · Cardiovascular
Device Facts
Record ID
K971355
Device Name
NIHON KOHDEN TEC-7531 CARDIOLIFE PORTABLE DEFIBRILLATOR AND ACCESSORIES
Applicant
Nihon Kohden America, Inc.
Product Code
LDD · Cardiovascular
Decision Date
Jul 15, 1997
Decision
ST
Submission Type
Traditional
Regulation
21 CFR 870.5300
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Nihon Kohden's TEC-7531A cardiolife is intended for medical purposes. The device will deliver an electrical shock of a maximum of 360 joules of energy used for defibrillating (restoring normal heart rhythm) the atria or ventricles of the heart or to terminate other cardiac arrhythmias. The electrical shock is delivered through paddles attached to the main unit. This device will supply a periodic electrical pulse intended to pace the heart when applied to the chest surface through electrodes. This device serves as a cardiac monitor, with the ability to measure heart rate and to sound an alarm when the heart rate falls outside preset upper and lower limits. This device will also condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g. a central monitoring station, and will recondition a physiological signal to its original format when received from another device, e.g. transmitter. The TEC-7531A cardiolife will be available for use by a physician, or under the supervision of a physician, for adult and pediatric patients within a medical facility and in remote environment.
Device Story
TEC-7531A cardiolife is a portable, multi-functional cardiac device; inputs include physiological signals via electrodes/paddles; device performs defibrillation (up to 360J), external transcutaneous pacing, and cardiac monitoring; outputs include electrical shocks, pacing pulses, heart rate data, and RF-transmitted physiological signals; used in medical facilities and remote environments by physicians or under physician supervision; provides real-time monitoring with alarm functionality for heart rate limits; benefits patients by restoring normal heart rhythm, terminating arrhythmias, and providing life-sustaining pacing and continuous monitoring.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including electromagnetic, environmental, safety, and performance testing, alongside software validation for signal acquisition, processing, display, and recording functions.
Technological Characteristics
Multi-functional device: defibrillator, external transcutaneous pacemaker, cardiac monitor, and RF physiological signal transmitter/receiver. Energy source: electrical. Connectivity: RF signal transmission. Software: performs signal acquisition, processing, display, and recording. Form factor: portable unit with external paddles/electrodes.
Indications for Use
Indicated for adult and pediatric patients requiring defibrillation to restore normal heart rhythm, termination of cardiac arrhythmias, external transcutaneous cardiac pacing, and cardiac monitoring with heart rate alarms.
Regulatory Classification
Identification
Low-energy DC-defibrillator: A device that delivers into a 50 ohm test load an electrical shock of a maximum of 360 joules of energy used for defibrillating (restoring normal heart rhythm) the atria or ventricles of the heart or to terminate other cardiac arrhythmias. This generic type of device includes low energy defibrillators with a maximum electrical output of less than 360 joules of energy that are used in pediatric defibrillation or in cardiac surgery. The device may either synchronize the shock with the proper phase of the electrocardiogram or may operate asynchronously. The device delivers the electrical shock through paddles placed either directly across the heart or on the surface of the body. High-energy DC-defibrillator: A device that delivers into a 50 ohm test load an electrical shock of greater than 360 joules of energy used for defibrillating the atria or ventricles of the heart or to terminate other cardiac arrhythmias. The device may either synchronize the shock with the proper phase of the electrocardiogram or may operate asynchronously. The device delivers the electrical shock through paddles placed either directly across the heart or on the surface of the body.
Predicate Devices
Nihon Kohden cardiolife, TEC-7200A Portable Defibrillator (K863405)
Nihon Kohden cardiolife, TEC-8251A Portable Defibrillator (K913085)
Nihon Kohden Life Scope 14, BSM-8800A Bedside Monitor (K920154)
Nihon Kohden Life Scope L, BSM-2101A Bedside Monitor (K914092)
Submission Summary (Full Text)
{0}------------------------------------------------
K971355
NIHON KOHDEN AMERICA, INC. April 9, 1997
JUL 1 5 1997
510(k) NOTIFICATION TEC-7531A cardiolife
## SECTION 2 - 510(k) SUMMARY
Name and Address of Applicant Nihon Kohden America, Inc. 2601 Campus Drive Irvine, California 92612-1601
Contact Person and Telephone Mr. Gary Reasoner Director of Product Operations (714) 250-3959 ext. 3387
The TEC-7531A cardiolife is classified as Class III by the Division of Cardiovascular, Respiratory and Neurological Devices and the Cardiovascular Device Classification Panel under 21 CFR Part 870.5550 External Transcutaneous Cardiac Pacemaker (Non-invasive) as per Product Classification Code 74 DRO. Additional device classification is Class II by the Division of Cardiovascular, Respiratory and Neurological Devices and the Cardiovascular Device Classification Panel under 21 CFR Part 870.5300 DC-Defibrillator, Low-energy (including paddles) as per Product Classification Code 74 LDD, under 21 CFR Part 870.2300 Monitor, Cardiac Monitor (including Cardiotachometer and Rate Alam) as per Product Classification Code 74 DRT and under 21 CFR Part 870.2910 Radiofrequency Physiological Signal Transmitter and Receiver as per Product Classification Code 74 DRG.
Common names for the TEC-7531A cardiolife are Low-energy DC-defibrillator, Cardiac Monitor, External Transcutaneous Cardiac Pacemaker, Radiofrequency Physiological Signal Receiver and Radiofrequency Physiological Signal Transmitter.
The predicate marketed devices are the Nihon Kohden cardiolife, TEC-7200A Portable Defibrillator per 510(k) #K863405, commercial distribution certification dated October 31, 1986, the Nihon Kohden cardiolife. TEC-8251A Portable Defibrillator per 510(k) #K913085, commercial distribution dated September 25, 1991, the Nihon Kohden Life Scope 14, BSM-8800A Bedside Monitor per 510(k) #K920154, commercial distribution dated December 18, 1992 and the Nihon Kohden Life Scope L, BSM-2101A Bedside Monitor per 510(k) #K914092. commercial distribution dated May 28. 1992.
Nihon Kohden's TEC-7531A cardiolife is intended for medical purposes. The device will deliver an electrical shock of a maximum of 360 ioules of energy used for defibrillating (restoring normal heart mythm) the atria or ventricles of the heart or to terminate other cardiac arrhythmias. The electrical shock is delivered through paddles attached to the main unit. This device will supply a periodic electrical pulse intended to pace the heart when applied to the chest surface through electrodes. This device serves as a cardiac monitor, with the ability to measure heart rate and to sound an alarm when the falls outside preset upper and lower limits. This device will also condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g. a central monitoring station, and will recondition a physiological signal to its original format when received from another device, e.g. transmitter. The TEC-7531A cardiolife will be available for use by a physician, or under the supervision of a physician, within a medical facility and in remote environment.
To date, no performance standards or special controls are known or established for this device as required by Section 514 of the Food, Drug and Cosmetic Act and implemented by 21 CFR Part 861.
The TEC-7531A cardiolife was subject to electromagnetic, environmental, safety and performance testing procedures. These tests verified the operation of the device. Software validation tested the operation of the software functions of acquiring, processing, displaying and recording of the results confirmed that the devices performed within specifications.
Therefore based on the preceding information, Nihon Kohden Corporation believes that the TEC-7531A cardiolife Portable Defibrillator is substantially equivalent to the predicate devices: the Nihon Kohden TEC-7200A cardiolife Portable Defibrillator, the Nihon Kohden TEC-8250A cardiolife Portable Defibrillator, the Nihon Kohden BSM-8800A Life Scope 14 Bedside Monitor and the Nihon Kohden BSM-2101A Life Scope L Bedside Monitor.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles facing to the right, with a bird-like shape above them.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 5 1997
Mr. Gary Reasoner Nihon Kohden America, Inc. 2601 Campus Drive Irvine, California 92612-1601
Re: K971355 Nihon Kohden TEC-7531A Cardiolife Portable Defibrillator and Accessories Requlatory Class: III (three) Product Code: 74 DRD Dated: April 9, 1997 Received: April 11, 1997
Dear Mr. Reasoner:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 - Mr. Gary Reasoner
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
NIHON KOHDEN AMERICA, INC. April 9, 1997
510(k) NOTIFICATION TEC-7531A cardiolife
G. Indications for Use Statement
K971355 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: TEC-7531A cardiolife Portable Defibrillator
Indications for Use:
Nihon Kohden's TEC-7531A cardiolife is intended for medical purposes. The device will deliver an electrical shock of a maximum of 360 joules of energy used for defibrillating (restoring normal heart rhythm) the atria or ventricles of the heart or to terminate other cardiac arrhythmias. The electrical shock is delivered through paddles attached to the main unit. This device will supply a periodic electrical pulse intended to pace the heart when applied to the chest surface through electrodes. This device serves as a cardiac monitor, with the ability to measure heart rate and to sound an alarm when the heart rate falls outside preset upper and lower limits. This device will also condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g. a central monitoring station, and will recondition a physiological signal to its original format when received from another device, e.g. transmitter.
The TEC-7531A cardiolife will be available for use by a physician, or under the supervision of a physician, for adult and pediatric patients within a medical facility and in remote environment.
Prescription Device <
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.