← Product Code [MBI](/productcode/MBI) · K971332

# AMS MAINSTAY UROLOGIC BONE ANCHOR (K971332)

_American Medical Systems, Inc. · MBI · Jul 9, 1997 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K971332

## Device Facts

- **Applicant:** American Medical Systems, Inc.
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Jul 9, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The AMS Mainstay Urologic Bone Anchor is intended only for the fixation of surgical suture material to the pelvis for the purpose of bladder neck suspension for female urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency.

## Device Story

AMS Mainstay Urologic Bone Anchor is a titanium alloy implant designed for bladder neck suspension procedures. Device functions by anchoring surgical suture material into cortical or cancellous bone of the pelvis, allowing secure attachment of soft tissue to bone. Implant features a proximal hole for suture attachment and a cancellous screw thread design for self-drilling/self-tapping installation. Used by surgeons in clinical settings to treat female urinary incontinence. By providing a stable bone-suture interface, the device facilitates the suspension of the bladder neck, addressing urethral hypermobility or intrinsic sphincter deficiency. Device is supplied as a sterile implant.

## Clinical Evidence

Bench testing only. Testing evaluated the Maximum Suture to Bone Anchor Interface Break Force for monofilament suture. Two sizes of non-absorbable monofilament suture (#3-0 and #2) were tested. Results showed all sutures exceeded USP minimum knot-pull strength requirements for Class I non-absorbable suture by more than a factor of four.

## Technological Characteristics

Material: Ti-6Al-4V titanium alloy. Design: Cancellous screw thread, self-drilling, self-tapping, proximal hole for suture attachment. Dimensions: 3.5 mm anchor. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Howmedica Mainstay Soft Tissue Anchor ([K953531](/device/K953531.md))

## Reference Devices

- Zimmer Statak Anchors ([K926384](/device/K926384.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

American Medical Sys Pfizer Hospital Products C 700 Bren Road West linnetonka, MN 55343 Tel 612 933 4666 Fax 612 930 Toll Free 800 328 3881

JUL - 9 1997

Image /page/0/Picture/3 description: The image shows the logo for Pfizer, a pharmaceutical company. The logo consists of the word "Pfizer" in a stylized, italicized font, with the "P" and "f" connected. The wordmark is placed inside a black oval shape. The logo is simple and recognizable, and it is often used in the company's marketing and advertising materials.

American Medical Systems

# SUMMARY OF SAFETY AND EFFEC TIVENESS

# SUBMITTER I.

American Medical Systems, Inc. Name and Address: 10700 Bren Road West Minnetonka, MN 55343 USA John M. Otto Contact Person: April 9, 1997 Date of Summary Preparation: June 9, 1997 Date Summary Amended: 2183959 Establishment Registration Number:

## DEVICE NAME II.

Device Common or Usual Name:

Staple, Fixation, Bone Appliance for Reconstruction of Bone to Soft Tissue

Device Trade Name:

AMS Mainstay™ Urologic Bone Anchor

## PREDICATE DEVICE III.

Howmedica® Mainstay™ Soft Tissue Anchor K953531

### DEVICE DESCRIPTION IV.

The AMS Mainstay™ Urologic Bone Anchor for bladder neck suspension procedures are titanium alloy implants. They are used to anchor suture material within cortical or cancellous bone sites, thereby providing a means for firmly securing soft tissue to bone. Two (2) Mainstay™ anchors are packaged in one box.

The recommended nonabsorbable suture size for the 3.5 mm Mainstay™ anchor is up to a maximum size of #0 (USP) when using two (2) strands and up to a maximum size of #2 when using a single strand.

## V. INDICATIONS FOR USE

The AMS Mainstay Urologic Bone Anchor is intended only for the fixation of surgical suture material to the pelvis for the purpose of bladder neck suspension for female urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency.

{1}------------------------------------------------

#### VI. COMPARISON TO PREDICATE DEVICE

The AMS Mainstay™ Urologic Bone Anchor with the proposed modifications is substantially equivalent to the currently marketed Howmedica® Mainstay™ Soft Tissue Anchor.

#### Intended Use a.

The AMS Mainstay™ Urologic Bone Anchor with the proposed modifications has the same intended use as one of the indications cleared for the currently marketed Howmedica® Mainstay™ Soft Tissue Anchor (Bladder Neck Suspension).

#### b. Device Performance

The AMS Mainstay™ Urologic Bone Anchor with the proposed modifications is comparable with respect to intended use and technological characteristics to the Howmedica® Mainstay™ Soft Tissue Anchor which is in commercial distribution in the United States. American Medical Systems has provided a table on the following page which compares the similarities and differences of the devices.

#### Bench Testing C.

The only new test performed for the AMS Mainstay™ Urologic Bone Anchor was for the Maximum Suture to Bone Anchor Interface Break Force for Monofilament Suture.

The objective of the test was to measure the maximum force required to break monofilament suture that has been attached to a bone anchor, and to compare these data to the minimum knot-pull strength requirements in the USP for the same size and class of suture. Two (2) sizes of a nonabsorbable monofilament suture were evaluated in this test.

All of the sutures tested exceeded the USP recommended minimum knot-pull strength for their respective class and size of suture (non-absorbable, Class I). The average suture break force for both the #3-0 and the #2 exceeded the USP minimum recommended knot-pull strength by more than a factor of four (4).

In summary, American Medical Systems has provided information within the 510(k) Premarket Notification to indicate that the AMS Mainstay™ Urologic Bone Anchor with the proposed modifications is safe and effective for its intended use (Bladder Neck Suspension) as one of the indications cleared for the currently marketed Howmedica® Mainstay™ Soft Tissue Anchor. Additionally, the modified AMS Mainstay™ Urologic Bone Anchor has been shown to be comparable in terms of intended use and technological characteristics to the Howmedica® Mainstay™ Soft Tissue Anchor currently in

{2}------------------------------------------------

commercial distribution. The data and information provided within this 510(k) Premarket Notification adequately support that the AMS Mainstay™ Urologic Bone Anchor with the proposed modifications is substantially equivalent to the Howmedica® Mainstay™ Soft Tissue Anchor that is currently in commercial distribution.

# EQUIVALENCY TABLE

| American Medical Systems<br>Mainstay Urologic Bone<br>Anchor | Howmedica Mainstay Soft<br>Tissue Anchor (K953531) | Zimmer Statak Anchors<br>(K926384) |
|--------------------------------------------------------------|----------------------------------------------------|------------------------------------|
| INTENDED USE                                                 |                                                    |                                    |
| a. Soft Tissue Attachment to<br>Bone                         | X                                                  | X                                  |
| b. Bladder Neck Suspension                                   | X                                                  | X                                  |
| MATERIAL                                                     |                                                    |                                    |
| Ti-6Al-4V                                                    | X                                                  | X                                  |
| DESIGN                                                       |                                                    |                                    |
| a. Cancellous Screw Thread                                   | X                                                  | X                                  |
| b. Self-drilling, Self-tapping                               | X                                                  | X                                  |
| c. Proximal Hole for Suture<br>Attachment                    | X                                                  | X                                  |
| OPERATIONAL<br>PRINCIPLES                                    |                                                    |                                    |
| a. Attachment Site for Suture                                | X                                                  | X                                  |
| b. Anchored by Screw Threads                                 | X                                                  | X                                  |
| c. Removable                                                 | X                                                  | X                                  |

ે. >

{3}------------------------------------------------

SERVICES • USA

DEPARTMENT OF HEALTH & HUMAN

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. John M. Otto Senior Requlatory Affairs Associate American Medical Systems, Inc. 10700 Bren Road West 55343 Minnetonka, Minnesota

Re : K971332 AMS Mainstay™ Urologic Bone Anchor Regulatory Class: II Product Codes: MBI and HWC Dated: June 9, 1997 Received: June 10, 1997

JUL - 9 1997

Dear Mr. Otto:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਰ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{4}------------------------------------------------

Page 2 - Mr. John M. Otto

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. , The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if Known): K971332

Device Name: AMS Mainstay™ Urologic Bone Anchor

Indications For Use: The AMS Mainstay™ Urologic Bone Anchor is intended only for the fixation of surgical suture material to the pelvis for the purpose of bladder neck suspension for female urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency

1. 2012 2011 11 1 112110

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Device Evaluation (ODE)

---

001 ULI 4343 . 4 0

(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K971332

Prescription Use X

OR

Over-The Counter Use

(Per 21 CFR 801.109)

t:\510k\ifuform.doc

---

**Source:** [https://fda-staging.innolitics.com/device/K971332](https://fda-staging.innolitics.com/device/K971332)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).