← Product Code [DKB](/productcode/DKB) · K971287

# SYNCHRON SYSTEMS DRUG CALIBRATION 3 PLUS (K971287)

_Beckman Instruments, Inc. · DKB · May 8, 1997 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K971287

## Device Facts

- **Applicant:** Beckman Instruments, Inc.
- **Product Code:** [DKB](/productcode/DKB.md)
- **Decision Date:** May 8, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3200
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

The SYNCHRON® Systems DRUG CALIBRATOR 3 PLUS, when used in conjunction with SYNCHRON Gentamicin Reagent and Tobramycin Reagent, is intended for use on the SYNCHRON Systems and for the calibration of gentamicin and tobramycin.

## Device Story

Liquid human serum matrix calibrator; contains gentamicin and tobramycin; used on Beckman SYNCHRON CX and LX clinical chemistry systems. Replaces previous single-analyte calibrator (gentamicin only) with combined six-level set. Used by laboratory personnel to calibrate clinical chemistry analyzers; ensures accurate recovery of analyte levels via anchor method correlation. Benefits patient by providing standardized, accurate drug concentration measurements for therapeutic monitoring.

## Clinical Evidence

Bench testing only. Stability testing performed at 25°C, 37°C, and 45°C; supports 24-month expiration claim. Value assignment process validated ability to recover weighed-in analyte levels correlated to known standards via anchor method.

## Technological Characteristics

Liquid human serum matrix; contains gentamicin and tobramycin; six-level calibrator set; designed for use with SYNCHRON CX and LX clinical chemistry analyzers.

## Regulatory Identification

A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- SYNCHRON Systems DRUG CALIBRATOR 3 ([K955644](/device/K955644.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K971287
MAY - 8 1997

# Summary of Safety &amp; Effectiveness
## SYNCHRON® Systems DRUG CALIBRATOR 3 PLUS

### 1.0 Submitted By:

Sheri Hall
Product Submissions Manager
Beckman Instruments, Inc.
200 S. Kraemer Blvd., W-337
Brea, California 92822-8000
Telephone: (714) 993-8961
FAX: (714) 961-4457

### 2.0 Date Submitted:

1 April 1997

### 3.0 Device Name(s):

#### 3.1 Proprietary Names

SYNCHRON® Systems DRUG CALIBRATOR 3 PLUS

#### 3.2 Classification Name

(21 CFR § 862.1150) Calibrator

### 4.0 Predicate Device(s):

|  SYNCHRON Systems
CAL 5 Plus | Predicate | Predicate Company | Docket Number  |
| --- | --- | --- | --- |
|  SYNCHRON Systems
DRUG CALIBRATOR 3
PLUS | SYNCHRON Systems
DRUG CALIBRATOR 3 | Beckman
Instruments, Inc. | K955644  |

{1}

Beckman Instruments, Inc., Section 510(k) Notification
SYNCHRON® Systems DRUG CALIBRATOR 3 PLUS
Summary of Safety &amp; Effectiveness

## 5.0 Description:

The SYNCHRON Systems DRUG CALIBRATOR 3 PLUS is designed for use with clinical chemistry analyzers from Beckman Instruments, such as the SYNCHRON CX® and LX™ Clinical Systems. The SYNCHRON Systems DRUG CALIBRATOR 3 PLUS is for use in the calibration of SYNCHRON Systems Gentamicin and Tobramycin chemistries.

## 6.0 Intended Use:

The SYNCHRON® Systems DRUG CALIBRATOR 3 PLUS, when used in conjunction with SYNCHRON Gentamicin Reagent and Tobramycin Reagent, is intended for use on the SYNCHRON Systems and for the calibration of gentamicin and tobramycin.

## 7.0 Comparison to Predicate(s):

The SYNCHRON Systems DRUG CALIBRATOR 3 PLUS is a liquid human serum matrix to which gentamicin and tobramycin are added. The SYNCHRON Systems DRUG CALIBRATOR 3 (six level set) contained only gentamicin. The SYNCHRON Systems DRUG CALIBRATOR 3 PLUS will contain both gentamicin and tobramycin into a single, six level, calibrator set. The SYNCHRON Systems DRUG CALIBRATOR 3 PLUS is intended to be used on Beckman's SYNCHRON Systems.

## 8.0 Summary of Performance Data:

The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence of the SYNCHRON Systems DRUG CALIBRATOR 3 PLUS to the SYNCHRON Systems DRUG CALIBRATOR 3 and is stable for gentamicin and tobramycin. The value assignment process for each analyte demonstrates the ability to recover weighed in analyte levels which can be correlated to a known standard via the anchor method.

### SYNCHRON Systems DRUG CALIBRATOR 3 PLUS
#### Stability Testing Summary

|  Stress Temperature | Duration of Incubation | Predicted Stability | Beckman Stability Claim*  |
| --- | --- | --- | --- |
|  25°C | 60 Days | 30 Months | 24 Months  |
|  37°C | 21 Days | 30 Months | 24 Months  |
|  45°C | 10 Days | 30 Months | 24 Months  |

* Expiration dating placed on the package based on date of manufacture

This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.

file: CAL3510K.DOC

---

**Source:** [https://fda-staging.innolitics.com/device/K971287](https://fda-staging.innolitics.com/device/K971287)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
