K971272 · Trend Medical, Inc. · KTW · Jun 5, 1997 · Orthopedic
Device Facts
Record ID
K971272
Device Name
TRAUMA-FIX EXTERNAL FIXATOR
Applicant
Trend Medical, Inc.
Product Code
KTW · Orthopedic
Decision Date
Jun 5, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
External fixation of long bone fractures. External fixation and stabilization of surgical osteotomy sites. Leg lengthening procedures. Knee, hip and ankle arthrodesis. External fixation of septic and aseptic non-union of long bones Pelvic fixation
Device Story
Trauma-Fix® External Fixator is a metallic bone fixation system available in unilateral (uni-bar) and quadrilateral (multi-plane) configurations. System components include pin holder clamps, articulation couplings, connecting rods (including telescopic), half-pins, full/transfixing pins, KTB wires, and various surgical instruments (drills, wrenches, drivers). Device is used by surgeons in clinical/OR settings to immobilize bone segments, manage open wounds, and facilitate fracture healing or limb lengthening. Unilateral frames provide simple, rapid fixation; quadrilateral frames offer multi-plane stability for complex reductions. The system allows for compression and distraction at the fracture site. The device is supplied non-sterile and is intended for single-patient use (disposable). Healthcare providers assemble the frame intraoperatively to provide mechanical support to the skeletal system, enabling weight-bearing or limb salvage as clinically indicated.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, material, and intended use comparisons to legally marketed predicate devices. Bench testing and adherence to ASTM/ISO standards are cited as the basis for safety and effectiveness.
Technological Characteristics
Metallic bone fixation system consisting of aluminum and stainless steel components. Includes clamps, rods, and pins. Employs mechanical fixation via half-pins or transfixing pins. Designed for unilateral or quadrilateral configurations. Supplied non-sterile; validated for gravity steam sterilization at 134°C for 15 minutes (SAL 10^-6). Complies with ASTM and ISO standards.
Indications for Use
Indicated for patients requiring external fixation of long bone fractures, stabilization of surgical osteotomy sites, leg lengthening, joint arthrodesis (knee, hip, ankle), treatment of septic/aseptic non-union of long bones, and pelvic fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
OrthoFrame™ by Thera-Kinetics, Inc.
Hoffman-Type External Fixation by Zimmer, Inc.
ACE-Fischer™ Fixator by DePuyACE
ACE-Align Fixator™ by DePuyACE
Submission Summary (Full Text)
{0}
JUN - 5 1997
K971272
| 510(k) Premarket Notification | Trauma-Fix® |
| --- | --- |
| Summary of Safety and Effectiveness Information | External Fixator |
Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92
1. Device Name:
Trade Name: Trauma-Fix® External Fixator
Common Name: External Fixator
Classification Name: Single/multiple component metallic bone fixation appliances and accessories
3. Establishment Name & Registration Number:
Name: Trend Medical, Inc.
Number: Pending
4. Classification:
accessories. (a) Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system. (b) Classification. Class II.
Product Code: Unknown
Device Class: Class II
Classification Panel: Orthopaedic and Rehabilitation Devices
5. Contact Person:
Mr. Gordon Tao
Trend Medical, Inc.
14433 Catalina Street
San Leandro, CA 94577
6. Special Controls:
Special controls have not been established for this device.
exfix.doc
18
{1}
exfix.doc
19
# 7. Description of the Device:
**Background:** The Trauma-Fix® External Fixator is offered as both a unilateral (uni-bar) and quadrilateral style (also called multi-plane) external fixation frame. External fixators known as uni-bar and multi-plane are typical of present industry standard designs. The uni-bar fixator concept is an outgrowth of the multi-plane fixator varying chiefly in the use of half-pins rather than the transfixing pins typical of multi-plane fixation, though occasionally half-pins are used with multi-plane fixators.
Unilateral frames using half pins and multi-plane fixators using half pins and transfixing pins were developed and popularized by Dr. Hoffmann near the turn of the century with the devices’ seeing much use in Germany during the First World War. At the time the external fixation concept was not much appreciated in this country. Doctor Hoffmann used his device in the management of gun shot wounds involving significant tissue and bone loss of the extremities. In the era prior to antibiotic therapy, such injuries frequently led to sepsis, amputation or death if primary wound closure or casting alone were attempted. Limb salvage under these conditions required the ability to immobilize the bony and soft injury and to irrigate or debride the soft tissue injury daily for an extended period of time. Casting and wound closure could be eliminated or delayed as needed to manage infection, granulation healing and fracture healing. An added benefit was the fact that some weight bearing was possible and limb shortening could be reduced or eliminated.
Uni-bar fixation using half-pins for the treatment of various long bone fractures and trauma was an outgrowth of the multi-plane transfixing pin concept. The chief benefit of uni-bar fixation is that it is generally a less complex method of fracture fixation and open wound management than is treatment with the multi-plane fixator. As a result, the uni-bar technique is usually easier. The limitation of the uni-bar technique is that it provides less adjustment or tolerance for significant rotational deformities or weight bearing. Also, when large axial compressive or distractive forces are encountered there can be a tendency for the axis of the long bone to deflect due to the lack of bilateral or multi-plane support provided by the additional connection rods of a multi-plane fixator construct.
The functional basis for the Trauma-Fix® External Fixator line is to provide fixators that allow the practitioner the option to use the appropriate fixator for the presenting fracture injury type. Open wound orthopaedic management without significant weight bearing or high fracture site compression requirements may be addressed in a simple cost effective manner.
The Trauma-Fix® External Fixator line is made up of the following items:
## Fixator
- Pin Holder Clamps, 5 & 8 mm diameter
- Articulation Coupling, 5 & 8 mm diameter
- Connecting Rods, 5 & 8 mm diameter
- Adjustable Telescopic Connecting Rods, 5 & 8 mm diameter
- Small Lengthening Device for Forearm
- Colles Forearm Frame
- Half-Pins
- Full-Pins/Transfixing Pins
- KTB Wires
- Uni-Bar Pin Clamp
- Uni-Bar Smooth Rods
- Uni-Bar Compression & Distraction Rod
- Uni-Bar Connector
{2}
# Instruments
Drill Brace
Chuck, 3, 4, 5, 6, mm
T-Wrench
Open Wrench
Hex Screw Driver
Drill Bit
Trochar & Sleeve
In summary, uni-bar-type fixators offer simple design, ease of use and rapid stable fixation. Quadrilateral frames provide a more complex structure but also offer maximum flexibility and will allow for complex fracture reduction in all three planes, anterior, posterior and axial rotation. Both frame types are designed to be disposable and are not recommended for reuse.
## 8. Comparison to Predicate Device(s):
Four currently available external fixation devices have been selected for comparison.
1. OrthoFrame™ by Thera-Kinetics, Inc. Comparison based on design and indications for use.
2. Hoffman-Type External Fixation by Zimmer, Inc. Comparison based on design, and indications for use.
3. ACE-Fischer™ Fixator by DePuyACE. Comparison based on design, materials and indications for use.
4. ACE-Align Fixator™ by DePuyACE. Comparison based on design, materials and indications for use.
The OrthoFrame™ is an example of a unilateral external fixator. This particular device is made from composite materials.
The Zimmer brand Hoffman-Type External Fixation is an example of an essentially unchanged Hoffmann-type external fixator. This device is essentially identical to Dr. Hoffmanns original devices. Made from certified stainless steel.
The ACE-Fischer™ Fixator is an example of a more modern design multi-plane fixator. The multi-plane fixator is analogous to the quadrilateral multi-plane fixator. The device is made from titanium alloy (6Al, 4V, ti) and aluminum.
The ACE-Align Fixator™ is also an example of a unilateral external fixator. This particular device is made from titanium alloy (6Al, 4V, ti) and aluminum.
## 9. Packaging:
Plastic bags, peel pouches, and clear tubes are used to contain the individual device components and or pins. The packaging selected for use is sufficient to identify, protect and transport the devices safely. Shipper materials are standard, paper fiber industry typical bulk box-type shipper packaging.
exfix.doc
{3}
# 10. Sterilization/Re-sterilization:
All fixation pins, instruments and external fixation compression and distraction components are supplied Non-Sterile. Non-Sterile implants are packaged in "clean only" condition but are free of manufacturing debris and residue. Each component is inspected after processing to evaluate and document the removal of manufacturing residue and debris. However, it is recommended that all pins, frames and clamps be removed from their shipping and packing materials and washed and rinsed thoroughly before storage and or sterilization.
The recommended sterilization method, time and temperature for the implants is gravity steam sterilization for 15 minutes at $134^{\circ}$ C ($270^{\circ}$ F). The Sterility Assurance Level (SAL) of this recommended sterilization cycle is $10^{-6}$ (SAL $10^{-6}$). Validation of the recommended cycle has been conducted by a qualified commercial laboratory. The validation method used is known as the overkill method. Other steam sterilization cycle time and temperature conditions may also render the device sterile.
The recommended sterilization method is based on Health Industry Manufacturers Association (HIMA) & the Association of Operating Room Nurses (AORN) protocols. Some health care facilities may wish to use other variations of the HIMA & AORN sterilization recommendations. However, individuals or hospitals not using the recommended method, temperature and time should validate any alternative sterility processing method using appropriate laboratory techniques.
# 12. Conclusion:
Based on the materials, intended uses, design, long standing safety and effectiveness use of the concept, the Uni-bar™ External Fixator unit is equivalent to the referenced legally marketed comparison external fixation devices. The feature comparison chart below graphically demonstrates this equivalence.
# 13. Comparison Table:
| FEATURE | Trauma-Fix®
External Fixator | Zimmer
Hoffmann | OrthoFrame | Ace | SE? |
| --- | --- | --- | --- | --- | --- |
| Intended Use: | External fixation of long bone fractures. External fixation and stabilization of surgical osteotomy sites. Leg lengthening procedures. Knee, hip and ankle arthrodesis. External fixation of septic and aseptic non-union of long bones Pelvic fixation | Same | Same | Same | Yes |
| Distraction | Yes | Yes | Yes | Yes | Yes |
| Compression: | Yes | Yes | Yes | Yes | Yes |
| Materials: | Aluminum & Stainless | Stainless steel | Composite | Al, Ti & Stainless | Yes |
| Design: | Uni-lateral & Quadrilateral | Quadrilateral | Uni-lateral | Uni-lateral & Multi-plane | Yes |
| Half Pins: | Yes | Yes | Yes | Yes | Yes |
| Transfix Pins: | Yes | Yes | Yes | Yes | Yes |
| Instruments: | Yes | Yes | Yes | Yes | Yes |
| Per. Standards: | ASTM - ISO | ASTM - ISO | ASTM - ISO | ASTM - ISO | Yes |
| Disposable: | Yes | No | No | No | No |
| Sterilization: | Steam | Steam | Steam | Steam | Yes |
| Manufacturer: | Trend Medical | Zimmer | Thera-Kenitics | DePuy/ACE | Yes |
exfix.doc
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUN - 5 1997
Mr. David W. Schlerf
Buckman Company, Inc.
Representing Trend Medical, Inc.
1000 Burnett Avenue, Suite 450
Concord, California 94520
Re: K971272
Trauma-Fix® External Fixator
Regulatory Class: II
Product Code: KTW
Dated: March 17, 1997
Received: April 4, 1997
Dear Mr. Schlerf:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}
Page 2 - Mr. David W. Schlerf
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Division of General and Restorative Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{6}
Page 1 of 1
510(k) Number (if known): K971272
Device Name: Trauma-Fix® External Fixator
## Indications For Use:
1. External fixation of long bone fractures.
2. External fixation and stabilization of surgical osteotomy sites.
3. Leg lengthening procedures.
4. Knee, hip and ankle arthrodesis.
5. External fixation of septic and aseptic non-union of long bones
6. Pelvic fixation
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY
Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ☑
OR
Over-The-Counter Use ☐
(Per 21 CFR 801.109)
(Optional format 1-2-96)
exfix.doc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.