← Product Code [LBI](/productcode/LBI) · K971256

# DUO-LITE PHOTOTHERAPY SYSTEM (K971256)

_Air-Shields, Inc. · LBI · Sep 24, 1997 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K971256

## Device Facts

- **Applicant:** Air-Shields, Inc.
- **Product Code:** [LBI](/productcode/LBI.md)
- **Decision Date:** Sep 24, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5700
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The DUO-LITE™ Phototherapy System is designed to provide treatment in the form of Phototherapy to those infants that have been identified as having elevated bilirubin levels (hyperbilirubinemia). Phototherapy as a means for reducing bilirubin levels is a well established treatment modality. The output of the light has been optimized for maximum absorption by the skin in order to reduce the bilirubin levels as quickly as possible. The DUO-LITE™ Phototherapy System provides the option to select white or blue light via a selector switch. This provides the caregiver with the option to give high output phototherapy for some infants (or early in the treatment cycle) followed by more moderate phototherapy for other infants (or later in the treatment cycle). By contrast, other phototherapy lamp products give the option of either White or Blue light which needs to be set up by an engineer in advance of the treatment. The DUO-LITE™ has combined these two light sources into one housing for the convenience of the caregiver. The DUO-LITE™ Phototherapy System can be placed on the hood of an infant incubator, mounted on the Air-Shields Rail System, or positioned over an open warming bed or crib.

## Device Story

Duo-Lite Phototherapy System treats neonatal hyperbilirubinemia by emitting light optimized for bilirubin absorption. Device consists of three 75W quartz halogen bulbs with IR/UV and dichroic filters. Caregiver selects white or blue light via housing switch to adjust therapeutic output intensity. Device mounts on incubator hoods, rail systems, or mobile stands over open warmers/bassinets. Includes digital counter for exposure time and bulb life tracking. Quartz halogen technology maintains constant irradiance over bulb life, unlike degrading fluorescent alternatives. Used by clinicians in hospital settings to reduce bilirubin levels; output intensity is regulated by internal power supply.

## Clinical Evidence

Bench testing only. Spectral irradiance measured from 250 to 1100nm across various filter, voltage, and distance configurations. In-house mattress mapping supported peak irradiance specifications. No clinical data provided.

## Technological Characteristics

Quartz halogen bulbs (3 x 75W); IR/UV and dichroic filters; 100/120/220/240V 50/60Hz power supply; dimensions 23"x10"x5"; weight ~15 lbs; digital timer for exposure/bulb life; manual selector switch for light color (white/blue).

## Regulatory Identification

A neonatal phototherapy unit is a device used to treat or prevent hyperbilirubinemia (elevated serum bilirubin level). The device consists of one or more lamps that emit a specific spectral band of light, under which an infant is placed for therapy. This generic type of device may include supports for the patient and equipment and component parts.

## Predicate Devices

- Air-Shields DUO-LITE™ Phototherapy System
- Air-Shields Micro-Lite™ Phototherapy System
- Air-Shields Fluoro-Lite® Phototherapy Unit
- Ohmeda Spot Phototherapy Lamp
- Olympic 66
- Olympic 33
- Drager PT4000

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SEP 24 1997
K971256

April 2, 1997

1. Submitter: Air-Shields, Inc.
330 Jacksonville Road
Hatboro, PA 19040
215-675-5200
Contact: Marci L. Goldfinger

2. Device Trade Name: Duo-Lite™ Phototherapy System
Device Classification Name: Neonatal Phototherapy Unit
Device Common Name: Phototherapy Unit

3. Predicate Devices:
The Air-Shields DUO-LITE™ Phototherapy System, Air-Shields Micro-Lite™ Phototherapy System, Air-Shields Fluoro-Lite® Phototherapy Unit, Ohmeda Spot Phototherapy Lamp, Olympic 66 and 33, and Drager PT4000.

4. Device Description:
The DUO-LITE™ Phototherapy System provides light output for maximum absorption by the infant's skin in order to reduce bilirubin levels. The DUO-LITE™ is designed for use in conjunction with a variety of other products. Perhaps the most common application is when placed on the hood of an incubator, or fitted to a rail system and positioned over the hood of the incubator. The DUO-LITE™ can also be mounted to a mobile stand, thus extending its applications to use with Open Warmers and Open Bassinets.

The product is L 23 x W 10 x H 5 (inches) and weighs approximately 15 Lbs. It is available in 100, 120 and 220/240V 50/60 Hz. The DUO-LITE™ is comprised of three 75W quartz halogen bulbs. The light emitted from these bulbs is filtered through IR/UV and Dichroic filters. The output of the bulbs is dependent on the driving voltage which is regulated by the internal power supply.

The user can select the use of White light or Blue light using the selector switch on the front of the lamp housing. The different color lights give different therapeutic outputs depending on the individual needs of the patient. Adjacent to the On/Off switch is a digital counter which can be set to display exposure time and bulb life remaining.

The DUO-LITE™ uses quartz halogen technology since the irradiance remains constant over the life of the bulb. By way of comparison, fluorescent bulbs degrade rapidly over the course of time and as such need to be continually monitored to ensure the output is therapeutically acceptable.

{1}

Page 2
Air-Shields Duo-Lite™ Phototherapy System
510(k) Summary

## 5. Statement of Intended Use:

The DUO-LITE™ Phototherapy System is designed to provide treatment in the form of Phototherapy to those infants that have been identified as having elevated bilirubin levels (hyperbilirubinemia). Phototherapy as a means for reducing bilirubin levels is a well established treatment modality. The output of the light has been optimized for maximum absorption by the skin in order to reduce the bilirubin levels as quickly as possible.

The DUO-LITE™ Phototherapy System provides the option to select white or blue light via a selector switch. This provides the caregiver with the option to give high output phototherapy for some infants (or early in the treatment cycle) followed by more moderate phototherapy for other infants (or later in the treatment cycle). By contrast, other phototherapy lamp products give the option of either White or Blue light which needs to be set up by an engineer in advance of the treatment. The DUO-LITE™ has combined these two light sources into one housing for the convenience of the caregiver.

The DUO-LITE™ Phototherapy System can be placed on the hood of an infant incubator, mounted on the Air-Shields Rail System, or positioned over an open warming bed or crib.

## 6. Comparison Matrix: Refer to Figure #1

## 7. Discussion of Non-clinical tests submitted:

Submitted in the 510(k) is a report on the spectral irradiance measurements conducted by an outside test house on the Duo-Lite™ Phototherapy System. The spectral irradiance was measured and recorded from 250 to 1100nm for the following configurations:

|  Filter Color | Voltage | Distance  |
| --- | --- | --- |
|  Blue | 120V | 17”  |
|  Blue | 132V | 17”  |
|  White | 120V | 17”  |
|  White | 132V | 17”  |
|  Blue | 120V | 15.75”  |
|  Blue | 132V | 15.75”  |
|  White | 120V | 15.75”  |
|  White | 132V | 15.75”  |

The results were as expected for a device of this type across this bandwidth and also indicate low UV and IR output. Mattress mapping done in our in-house laboratory support our peak irradiance specifications. The data supports our claim of substantial equivalence to similar phototherapy devices already on the market.

{2}

Air-Shields Duo-Lite™ Phototherapy System

510(k) Summary

Product Comparison Matrix

|  Product Feature | Air-Shields DUO-LITE™ | Air-Shields Micro-Lite® | Air-Shields Fluoro-Lite® | Ohmeda Spot PT Lamp | Olympic Model 66 | Olympic Model 33 | Drager PT 4000  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Type | Freestanding | Freestanding | Freestanding | Freestanding | Freestanding | Freestanding | Freestanding  |
|  |   |   |   |   |   |   |   |
|  Size (inches) |  |  |  |  |  |  |   |
|  Height on Stand | 55 - 65 | 55 - 65 | 38.5 - 56 | 57 - 91 | 49 - 65 | 42 - 61 | 42 - 64  |
|  L x W | 21 x 17 | 21 x 17 | 27 x 13 | 12 x 6.25 | 26 x 11 | 25 x 19 | 21.3 x 11.2  |
|  |   |   |   |   |   |   |   |
|  Weight (on stand) | 66 Lbs | 56.5 Lbs | 86 Lbs | 40 Lbs | 54 Lbs | 52 Lbs | 45.7 Lbs  |
|  |   |   |   |   |   |   |   |
|  Angle Adjustment | 0 - 45° | 0 - 45° | +/-90° | Multi Directional | 0 - 90° | 0 - 60° | 0 - 30°  |
|  |   |   |   |   |   |   |   |
|  Irradiance (400 - 520 Nm) | ≥10 (White) ≥20 (Blue) μW/cm²/nm @ 43 cm | >9μW/cm²/nm @ 43 cm | >12μW/cm²/nm @ 39 cm | 4 - 20 μW/cm²/nm | 4 - 9 (White) 4 - 20 (Blue) | 4 - 9 (White) 4 - 20 (Blue) | >13μW/cm²/nm @ 40 cm  |
|  |   |   |   |   |   |   |   |
|  Bulb |  |  |  |  |  |  |   |
|  Type | Quartz Halogen | Quartz Halogen | Fluorescent | Quartz Halogen | Fluorescent | Fluorescent | Fluorescent  |
|  Number | 3 x 75W | 3 x 50W | 4 x 20W | 1 x 250W | 4 x 20W | 8 x 20W | 6 x 20W  |
|  Life | 1,000 Hrs | 1,000 Hrs | 500 Hrs | 500 Hrs | 200 Hrs | 200 Hrs | 1,000Hrs  |
|  Color | Blue or White | White | Blue and/or White | White | Blue and/or White | Blue and/or White | Blue  |
|  |   |   |   |   |   |   |   |
|  Procedure Timer | Yes | Yes | Optional | No | Optional | Optional | Yes  |
|  |   |   |   |   |   |   |   |
|  Line Voltage | 120 Vac | 120 Vac | 120 Vac | 115 Vac | 115 Vac | 120 Vac | 110 - 115 Vac  |
|  |   |   |   |   |   |   |   |
|  Power Consumption | 350W | 300W | 140W | 80 - 100W | 160 - 180W | 160 - 180W | 120W  |
|  |   |   |   |   |   |   |   |
|  Leakage Current | <300μA | <100μA | <100μA | <100μA | <100μA | <100μA | <300μA  |
|  |   |   |   |   |   |   |   |
|  Communication Port | Yes | No | No | No | No | No | No  |

FIGURE #1

{3}

DEPARTMENT OF HEALTH &amp; HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850

SEP 24 1997

Ms. Marci L. Goldfinger
Director, Quality Assurance and Regulatory Affairs
Air Shields, Incorporated
330 Jacksonville Road
Hatboro, Pennsylvania 19040-2211

Re: K971256
Trade Name: Duo-Lite Phototherapy System
Regulatory Class: II
Product Code: BLI
Dated: July 10, 1997
Received: July 11, 1997

Dear Ms. Goldfinger:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{4}

Page 2 - Ms. Goldfinger

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

![img-0.jpeg](img-0.jpeg)

Timothy A. Ulatowski
Director
Division of Dental, Infection Control and General Hospital Devices
Office of Device Evaluation
Center for Devices and Radiological Health

Enclosure

{5}

Page ___ of ___

510(k) Number (if known): K971256

Duo-Lite™ Phototherapy System

Device Name:

## Indications For Use:

The DUO-LITE™ Phototherapy System is designed to provide treatment in the form of Phototherapy to those infants that have been identified as having elevated bilirubin levels (hyperbilirubinemia).

Phototherapy as a means for reducing bilirubin levels is a well established treatment modality. The output of the light has been optimized for maximum absorption by the skin in order to reduce the bilirubin levels as quickly as possible.

The DUO-LITE™ Phototherapy System provides the option to select white or blue light via a selector switch. This provides the caregiver with the option to give high output phototherapy for some infants (or early in the treatment cycle) followed by more moderate phototherapy for other infants (or later in the treatment cycle). By contrast, other phototherapy lamp products give the option of either White or Blue light which needs to be set up by an engineer in advance of the treatment. The DUO-LITE™ has combined these two light sources into one housing for the convenience of the caregiver.

The DUO-LITE™ Phototherapy System can be placed on the hood of an infant incubator, mounted on the Air-Shields Rail System, or positioned over an open warming bed or crib.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Dental, Infection Control, and General Hospital Devices
510(k) Number ________

Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ________

(Optional Format 1-2-96)

![img-1.jpeg](img-1.jpeg)

---

**Source:** [https://fda-staging.innolitics.com/device/K971256](https://fda-staging.innolitics.com/device/K971256)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).