SHERWOOD MEDICAL COMPANY , KANGAROO, ENTRISTAR, PERCUTANEOUS GASTROSTOMY KIT
K971230 · Innerdyne, Inc. · KNT · Jun 13, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K971230
Device Name
SHERWOOD MEDICAL COMPANY , KANGAROO, ENTRISTAR, PERCUTANEOUS GASTROSTOMY KIT
Applicant
Innerdyne, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Jun 13, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sherwood Medical Company, Kangaroo®, EntriStar®, "Percutaneous Gastrostomy Kit" is a medical device for the percutaneous placement of a long-term enteral feeding tube.
Device Story
The Kangaroo®, EntriStar® Percutaneous Gastrostomy Kit is a sterile, single-use system for placing enteral feeding tubes. The kit includes a radially expanding access device (an expandable dilator sheath with an access needle), a gastrostomy tube, obturator, insertion/removal device, and various surgical accessories (scalpel, drape, gauze, lubricant). The device is used by clinicians in a clinical setting to establish gastric access via endoscopic or laparoscopic guidance. The access needle/sheath assembly is inserted through the abdominal and stomach walls; the needle is removed, and the dilator is advanced to facilitate tube placement. This system provides a minimally invasive method for long-term nutritional support, benefiting patients unable to ingest food orally.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
The kit includes a radially expanding access device (expandable dilator sheath assembly), gastrostomy tube, and surgical accessories. The dilator sheath is designed for axial compression to reduce diameter during insertion. The system is sterile and intended for single use.
Indications for Use
Indicated for patients requiring percutaneous placement of a long-term enteral feeding tube.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Submission Summary (Full Text)
{0}
JUN 13 1997
K971230
Page 1 of 1
# SUMMARY OF SAFETY & EFFECTIVENESS INFORMATION
Submitter: InnerDyne, Inc.
5060 West Amelia Earhart Drive
Salt Lake City, Utah 84116
Estab. Reg. No: 1721520
Date Prepared: 15 March 1997
Contact: Rick Gaykowski
Vice President, Regulatory Affairs
and Quality Assurance
Sponsor: Sherwood Medical Company
1915 Olive Street
St. Louis, MO 63103
Estab. Reg. No.: 1924954
Classification Name: Gastrointestinal Tube & Accessories
Generic Name: Gastrostomy Tube & Accessories
Trade Name: Sherwood Medical Company, Kangaroo®, EntriStar®,
"Percutaneous Gastrostomy Kit"
The Kangaroo®, EntriStar®, "Percutaneous Gastrostomy Kit" is a sterile, single use system containing a gastrostomy tube for enteral feeding and the components required to insert the tube percutaneously, which may include endoscopic and/or laparoscopic methodology. The kit consists of a radially expanding access device, gastrostomy tube, obturator, insertion/removal device, "Y" connector, luer adapter, external retention disk, retention band, #11 scalpel, drape, gauze pads, lubricant, and Instruction for Use (IFU).
The radially expanding access device is an expandable dilator sheath assembly with an access needle mounted within its lumen. The tubular member of the dilator sheath is configured so as to be axially compressed to reduce the outside diameter of the device prior to insertion. Upon use, the expanding dilator sheath/needle assembly is inserted through the abdominal tissue into the gastric cavity.
If endoscopic technique is utilized, an endoscope is advanced into the stomach using standard techniques, followed by stomach insufflation. An abdominal incision is made and the access needle/sheath assembly is inserted directly through the abdominal and stomach walls into the stomach. The access needle is removed and the cannula/dilator is advanced through the radially expanding sleeve. If laparoscopic technique is utilized, under laparoscopic vision the abdominal cavity is entered and the stomach site is selected for insertion of the gastrostomy tube.
InnerDyne, Inc.
Kangaroo®, EntriStar® Gastrostomy Kit Premarket Notification
Page-12
{1}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUN 13 1997
Sherwood Medical Company
% Mr. Rick Gaykowski
Corporate Vice President, Regulatory Affairs and Quality Assurance
InnerDyne, Inc.
5060 West Amelia Earhart Drive
Salt Lake City, Utah 84116
Re: R971230
Sherwood Medical Company, Kangaroo®, EntriStar®, "Percutaneous Gastrostomy Kit"
Dated: March 31, 1997
Received: April 2, 1997
Regulatory class: II
21 CFR §876.5980/Product code: 78 KNT
Dear Mr. Gaykowski:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act
{2}
Page 2 - Mr. Rick Gaykowski
for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{3}
Page 1 of 1
510(k) Number (if known): K971230
Device Name: Sherwood Medical Company, Kangaroo®, EntriStar®, "Percutaneous Gastrostomy Kit"
Indications for Use: The Sherwood Medical Company, Kangaroo®, EntriStar®, "Percutaneous Gastrostomy Kit" is a medical device for the percutaneous placement of a long-term enteral feeding tube.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑ OR Over-the-Counter Use
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
Cawley Y. Neuland for R. Getting
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K971230
InnerDyne, Inc.
Kangaroo®, EntriStar® Gastrostomy Kit Premarket Notification
Page-10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.