K971146 · Rd Chus, Inc. · KNF · Aug 25, 1997 · Obstetrics/Gynecology
Device Facts
Record ID
K971146
Device Name
LAPARETTE
Applicant
Rd Chus, Inc.
Product Code
KNF · Obstetrics/Gynecology
Decision Date
Aug 25, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LAPARETTE is an electrosurgical instrument with integral suction and irrigation capability. It is intended for use in minimally invasive surgery, including laparoscopy, to cut tissue and control bleeding by use of high-frequency electrical current, to irrigate tissue and cavities, as well as to suction fluids from the wound. This device is not intended for use in tubal sterilization procedures.
Device Story
Hybrid electrosurgical device; combines reusable handle, electrode-cannula, and insulated sleeve with single-use cartridge and sheath. Used in minimally invasive surgery/laparoscopy; operated by surgeons. Device delivers high-frequency electrical current for tissue cutting and coagulation; provides integrated suction and irrigation for wound management. Features pistol-grip handle; tip rotation; retractable sheath for tip shielding during insertion. Components supplied non-sterile; requires facility-based steam or ethylene oxide sterilization. Output is mechanical/electrical action on tissue; surgeon observes effect directly via laparoscope. Benefits include multi-functional capability in single instrument, reducing instrument exchanges during surgery.
Clinical Evidence
Bench and animal testing provided to characterize performance. No human clinical trial data reported.
Technological Characteristics
Hybrid reusable/disposable electrosurgical instrument. Materials: medical-grade polymers. Energy: monopolar high-frequency electrical current. Form factor: pistol-grip handle, 5.5 mm trocar compatibility, ~30.5 mm shaft length. Sterilization: steam (reusable components) and ethylene oxide (protective sheath).
Indications for Use
Indicated for patients undergoing minimally invasive surgery, including laparoscopy, requiring tissue cutting, coagulation, irrigation, or fluid suction. Contraindicated for tubal sterilization procedures.
Regulatory Classification
Identification
A unipolar endoscopic coagulator-cutter is a device designed to destroy tissue with high temperatures by directing a high frequency electrical current through the tissue between an energized probe and a grounding plate. It is used in female sterilization and in other operative procedures under endoscopic observation. This generic type of device may include the following accessories: an electrical generator, probes and electrical cables, and a patient grounding plate. This generic type of device does not include devices used to perform female sterilization under hysteroscopic observation.
{0}
K971146
510(k) Summary
RD CHUS, Inc. LAPARETTTE
(per 21 CFR 807.92)
AUG 25 1997
1. DATE OF PREPARATION: March 27, 1997
2. SPONSOR/APPLICANT: RD CHUS Inc.
3001, 12th ave North,
Fleurimont, Quebec, Canada, J1H 5N4
3. CONTACT NAME: Fernand Jalbert
Telephone: 819-563-5555 Ext. 14252
4. DEVICE NAME:
Trade/Proprietary Name: LAPARETTTE
Common/Usual Name: Electrosurgical electrode, suction, irrigation
Classification Name: Accessory to electrosurgical cutting and coagulation device or laparoscopic accessory
5. IDENTIFICATION OF THE PREDICATE OR LEGALLY MARKETED DEVICE(S) TO WHICH EQUIVALENCE IS BEING CLAIMED:
Davis+Geck’s Multi-Function Pistol Grip (K921696)
American Cyanamid Co.
Ethicon Endopath Probe Plus (K912492)
Ethicon, Inc.
6. DEVICE DESCRIPTION:
INTENDED USE: The LAPARETTTE is intended for use in minimally invasive surgery, including laparoscopy, to cut tissue and control bleeding by use of high-frequency electrical current, irrigate tissue and cavities, as well as suction fluids from the wound. This device is not intended for use in tubal sterilization procedures.
RD CHUS, Inc. LAPARETTTE
3/27/97
510(k) Summary - Page 1
{1}
K971146
ρ 2η²
**OVERVIEW:** The LAPARETTE is a hybrid device consisting of three reusable (Handle, Electrode-Cannula, and Insulated Sleeve) and two single use components (Cartridge and Sheath). All LAPARETTE components are supplied non-sterile, requiring sterilization by the health care facility before use. The LAPARETTE is sold both as a set and as individual products. Bench and animal testing provided in the 510(k) was used to characterize the performance of the LAPARETTE.
7. **BASIS FOR SUBSTANTIAL EQUIVALENCE:** RD CHUS Inc. makes the claim of substantial equivalence to the above devices based on intended use, design considerations, and general operating characteristics. All three devices share the same intended use of electrocautery, irrigation, and suction using the same instrument. All three products are designated for monopolar use only. While the D&G Multi-Function and Ethicon Probe Plus are single use only, the LAPARETTE is a hybrid of reusable and single use devices. The recommended sterilization technique for the LAPARETTE is steam for all but the Protective Sheath which requires ethylene oxide sterilization. Although the specific materials of construction are different, all cited devices use polymers common to various medical devices. All three products use a pistol grip handle design. The LAPARETTE uses a separate disposable cartridge, while that function is integral to the handles of the D&G Multi-Function and Ethicon Probe Plus. The LAPARETTE and the D&G Multi-Function share the feature of tip rotation, while the Ethicon Probe Plus requires that the entire device is rotated. While the LAPARETTE and the Ethicon Probe Plus feature the ability to shield the tip during insertion into the wound by providing the ability to retract the sheath over the exposed tip, the D&G Multi-Function does not allow retraction. The dimensions of all three devices are similar with usable shaft lengths of 30.5 mm for the LAPARETTE, 31.75 mm for the D&G Multi-Function, and 35 mm for the Ethicon Probe Plus. Shaft O.D.s range from 5.0 mm to 5.4 mm. Devices are compatible with trocar sizes of 5.5 (LAPARETTE) and 5.0 for the other devices.
RD CHUS, Inc. LAPARETTE 3/27/97 510(k) Summary - Page 2
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
RDCCHUS, Inc.
c/o Rosina Robinson, R.N., RAC
Medical Device Consultants, Inc.
Staff Consultant
49 Plain Street
North Attleboro, Massachusetts 02760
AUG 25 1997
Re: K971146
Laparette
Dated: July 9, 1997
Received: July 10, 1997
Regulatory class: II
21 CFR §884.4160/Product code: 85 KNF
Dear Ms. Robinson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Enclosure
{3}
510(k) Number (if known): New 510(k) 5971146
Device Name: LAPARETE
Indications For Use: The LAPARETE is an electrosurgical instrument with integral suction and irrigation capability. It is intended for use in minimally invasive surgery, including laparoscopy, to cut tissue and control bleeding by use of high-frequency electrical current, to irrigate tissue and cavities, as well as to suction fluids from the wound. This device is not intended for use in tubal sterilization procedures.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Robert D. Lathing
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number 5971146
Prescription Use ☑
(Per 21 CFR 801.109)
OR Over-The-Counter Use ☐
(Optional Format 1-2-96)
RD CHUS Inc.
LAPARETE 510(k)
3/27/97
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.