← Product Code [GCT](/productcode/GCT) · K971057

# COGENLIGHT TECHNOLOGIES XLS ILLUMINATOR SYSTEM (K971057)

_Cogent Light Technologies, Inc. · GCT · Mar 28, 1997 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K971057

## Device Facts

- **Applicant:** Cogent Light Technologies, Inc.
- **Product Code:** [GCT](/productcode/GCT.md)
- **Decision Date:** Mar 28, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Both devices are intended for the same use, that of providing visible light for the illumination of surgical headlights and various rigid or flexible endoscopes that contain fiber bundles for illumination. Applications for both systems include, but are not limited to, externally illuminated endoscopes used in arthroscopy, bronchoscopy, gynecology, laparoscopy, obstetrics, oto-rhino-laryngoscopy, urology, and vascular endoscopy.

## Device Story

XLS Illuminator System provides high-intensity visible light for surgical visualization. Device consists of a control box housing a Xenon arc lamp and power supply; connects to proprietary single quartz fiberoptic cable or standard fiber bundles via adapter. Used in OR/clinical settings by surgeons/staff to illuminate surgical sites via endoscopes or headlamps. Safety features include lamp access interlock, forced-air cooling, internal thermal protection, and system timer limiting lamp use to 650 hours (below 1000-hour rating). Housing includes internal baffles to contain debris in event of non-passive lamp failure. Single fiber technology provides uniform illumination compared to traditional fiber bundles. Device benefits include consistent, uniform light output and reduced risk of lamp failure during procedures.

## Clinical Evidence

No clinical data. Bench testing only. Safety and effectiveness supported by design specifications, thermal protection mechanisms, lamp life management, and comparative analysis of spectral output and illumination uniformity against predicate technology.

## Technological Characteristics

Xenon arc lamp light source; forced-air cooling; internal thermal protection; safety interlock; system timer for lamp life management; internal baffles for debris containment. Proprietary single quartz fiberoptic cable or fiber bundle adapter. Standalone control box unit.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Cogent Light Technologies Illuminator System ([K971057](/device/K971057.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

From: Keith Lowrey To: FDA Donald St. Pierre Date: 3/27/97 Time: 09:21:04 Page 1 of 2 MAR 28 1997 K971057

Cogent Light Technologies 510(k) Premarket Notification for the XLS Illuminator System
510(k) Number: K 971057

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS:

A. Safety:

The Cogent Light XLS Illuminator System is designed with a safety interlock on the lamp (XLF) access cover. If the cover is opened, the power to the illuminator will be interrupted. The instrument is cooled with a forced air fan to prevent the temperature of the unit from exceeding a safe level. If in the event that the fan malfunctions and fails to operate properly and the internal temperature of the illuminator exceeds the engineered specifications, the device is designed with an internal thermal protection mechanism that automatically shuts off the power to the unit.

The lamp (XLF) with fixture and power source may fail passively (i.e., just not turn on, or just go out). Typically, lamp and power source defects do not result in lamps going out during a surgical case. It is more likely that a lamp cannot be turned on during preparation for a surgical case. The consequences is that this would cause an inconvenience and a time delay. However, it is expected to result in no adverse medical condition in the patient.

To minimize such inconveniences, the XLS Illuminator System contains a user replaceable lamp (XLF) and would encourage the operator to have a spare lamp available on site.

The lamp (XLF) may fail in a non-passive mode, resulting in an audible sound which may startle the user. Due to the inherent nature of all short arc Xenon lamps, including those found in the predicate device, it is impossible to guarantee against all such failures, although they are rare.

To minimize such events, the XLS system has incorporated several safety features into the product design. These safety features include:

(1) The lamp is never operated beyond its specified power range.

(2) The lamp is limited to 650 operating hours by a system timer. This limit is significantly below the safe rated life expectancy of the lamp (approximately 1000 hours) as stated by the lamp manufacturer.

(3) The housing of the system has interior structures (baffles), which will dampen the sound level and confine any debris which might result from the non-passive lamp failure.

(4) The user replaceable Lamp Housing Module (XLF) would minimize the inconvenience.

B. Effectiveness:

{1}

From: Keith Lowrey To: FDA Donald St. Pierre
Date: 3/27/97 Time: 09:22:08
Page 2 of 2
K971057

# Cogent Light Technologies 510(k) Premarket Notification for the XLS Illuminator System
510(k) Number: **K 971057**

All Xenon arc lamps produce light that is spectrally similar to sunlight. What is actually delivered to an endoscope, surgical headlight, or surgical instrument, is determined by optical coatings and the transmissive properties of the light delivery system (single or bundle fiber).

The XLS Illuminator uses a proprietary single fiberoptic cable. A single multimode fiber typically produces a uniform image at the output end independent of the intensity profile of the light on the input end. By contrast, fiber bundles produce non-uniform images because they consist of many fibers bonded together, producing spaces between the fibers and generally low light intensity at the very center. Any application involving direct illumination (e.g., surgical headlamp) requires preferably uniform illumination.

Fiber bundles have been the standard light delivery system in use. Liquid bundles are somewhat more transmissive, but substantially more expensive. Light delivered through a fiber bundle is attenuated more in the blue than the red,

producing a more yellow light than that from a liquid bundle. Cogent Light Technologies’ single quartz fiber, used as part of the subject device, as well as the predicate device, transmits white light uniformly throughout the visible spectrum.

## C. SUMMARY:

Cogent Light Technologies believes that the Cogent Light XLS Illuminator System is substantially equivalent in design, component materials, function, labeling, and intended use to the Cogent Light Technologies Illuminator System.

Both devices are intended for the same use, that of providing visible light for the illumination of surgical headlights and various rigid or flexible endoscopes that contain fiber bundles for illumination. Applications for both systems include, but are not limited to, externally illuminated endoscopes used in arthroscopy, bronchoscopy, gynecology, laparoscopy, obstetrics, oto-rhino-laryngoscopy, urology, and vascular endoscopy.

Both devices consist of a light source control box which houses a Xenon lamp, power supply, and connects to a proprietary fiberoptic cable. With a fiber bundle adapter, this illuminator can also be used with regular fiber bundle optic cables and headlamps.

510(k) Number: **K 971057**

---

**Source:** [https://fda-staging.innolitics.com/device/K971057](https://fda-staging.innolitics.com/device/K971057)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
