K971039 · Hemagen Diagnostics, Inc. · MID · Jun 3, 1997 · Immunology
Device Facts
Record ID
K971039
Device Name
HEMAGEN ANTI CARDIOLIPIN SCREEN
Applicant
Hemagen Diagnostics, Inc.
Product Code
MID · Immunology
Decision Date
Jun 3, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5660
Device Class
Class 2
Indications for Use
The device is intended for the qualitative detection of IgG, IgM, and IgA autoantibodies to cardiolipin in human serum.
Device Story
ELISA-based in vitro diagnostic assay; detects IgG, IgM, and IgA autoantibodies to cardiolipin in human serum. Bovine cardiolipin antigen immobilized on microplate wells; patient serum added; anticardiolipin antibodies bind to antigen. Trivalent peroxidase-conjugated goat anti-human IgG, IgM, and IgA cocktail added; TMB substrate produces color change proportional to bound antibody concentration. Reaction stopped with sulfuric acid; optical density (OD) measured via ELISA plate reader. ODs compared to reference serum. Used in clinical laboratory settings by trained personnel. Results aid in diagnosis of antiphospholipid syndrome.
Clinical Evidence
Bench testing only. Comparative study of 117 serum samples (38 positive, 79 normal) using proposed device vs. predicate devices. Results: relative sensitivity 100% (95% CI: 91.0%-100%), relative specificity 92.3% (95% CI: 84.2%-96.4%). Precision studies (inter-assay and intra-assay) performed with negative, low positive, and high positive samples; CVs ranged from 6.5% to 15.0%.
Technological Characteristics
ELISA-based immunological test system. Bovine cardiolipin antigen coated on microplate wells. Trivalent HRP-conjugated goat anti-human IgG, IgM, and IgA detection antibody. TMB substrate. Optical density measurement via ELISA plate reader. Standalone diagnostic kit.
Indications for Use
Indicated for screening patient serum for IgG, IgM, and IgA autoantibodies to cardiolipin. Used as an aid in the diagnosis of antiphospholipid syndrome in patients with systemic lupus erythematosus and other connective tissue diseases.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Predicate Devices
Hemagen ® Anti-Cardiolipin IgG and IgM {EIA method} (K932373)
Hemagen ® Anti-Cardiolipin IgA Calibrator Reagents {EIA method} (K941840)
Submission Summary (Full Text)
{0}
PAGE 1
K971039
510(k) Summary
1. Submitter's Name/Contact Person
JUN - 3 1997
Joseph M. Califano, Regulatory Affairs Manager
Address
Hemagen Diagnostics, Inc.
34-40 Bear Hill Road
Waltham, MA, 02154
Phone No: (617) 890-3766
Fax No: (617) 890-3748
e-mail: jcalifano@hemagen.com
**Date Prepared**
7 March 1997
2. **Device Names**
Trade Name: Hemagen ® aCL Screen
Common Name: anti Cardiolipin screening assay
Classification Name: Immunological test system, multiple autoantibodies {21 CFR 866.5660}
3. **Predicate Device(s)**
a. Hemagen ® Anti-Cardiolipin IgG and IgM {EIA method}
Reference 510 (k) Docket No. K 932373
b. Hemagen ® Anti-Cardiolipin IgA Calibrator Reagents {EIA method}
Reference 510 (k) Docket No. K 941840
510(k) Summary - Page 1
{1}
PAGE 2
## 4 Description of Device
The ELISA methodology is commonly used for serum antibody evaluations. In this assay, purified bovine cardiolipin has been attached to the inner surfaces of microplate test wells by the manufacturer. The user adds diluted patient samples to the wells. Anticardiolipin antibodies in patient serum bind specifically to the cardiolipin antigen attached to the plate, and remain in place after a wash step.
A trivalent second antibody cocktail, consisting of peroxidase-conjugated goat anti-human IgG (Fc), IgM (μ), and IgA (α), is added to the wells. After an incubation/wash step, TMB (tetramethylbenzidine) substrate is added. In the wells containing bound cardiolipin-anticardiolipin complexes, the peroxidase enzyme catalyzes a color change in the TMB substrate. After the reaction is stopped with dilute sulfuric acid, the color is read in an ELISA plate reader. Optical densities (ODs) of patient samples are compared to the OD of a reference serum included in the kit.
## 5 Intended Use of Device
The device is intended for the qualitative detection of IgG, IgM, and IgA autoantibodies to cardiolipin in human serum.
## 6. Performance Data
### I. Trivalent HRP Conjugate Standardization
IgG, IgM, and IgA standard preparations (Louisville APL Diagnostics, Inc.) were diluted in Serum Diluent and tested with the predicate devices to establish standard curves. Optimal second antibody concentrations were established by titration for each monovalent HRP conjugate. The three conjugates were then blended, verified against the standard calibration curves, and freeze-dried.
### II. Cutoff Serum Standardization
117 serum specimens from both disease state and normal patient panels were screened concurrently with both the proposed and predicate devices. Receiver Operating Characteristic (ROC) techniques were used to derive an optimized cutoff OD to be utilized in the preparation of the Cutoff Serum. Based upon this analysis and comparison testing, the optimized cutoff OD has demonstrated a relative analytical sensitivity of ≥ 97.0 %.
510(k) Summary - Page 2
{2}
PAGE 3
# III. Comparison Testing
The Hemagen® antiCardiolipin Screen, the Hemagen® Anti-Cardiolipin IgG/ IgM assays, and the Hemagen® Anti-Cardiolipin IgA Calibrator Reagents were used to concurrently assay serum specimens from patients positive for aCl, and from normal blood donors. A total of 117 samples were evaluated.
Table 1: Positive Panel, N = 38
| | Predicate Devices | | |
| --- | --- | --- | --- |
| | POS | NEG | TOTAL |
| Proposed Device | POS | 38 | 0 |
| | NEG | 0 | 0 |
| | TOTAL | 38 | 0 |
| | Relative analytical sensitivity = 100 % {38/38}, 0.95 INTERVAL 90.8% to 100% | | |
Table 2: Normal Panels, N=79
| | Predicate Devices | | |
| --- | --- | --- | --- |
| | POS | NEG | TOTAL |
| Proposed Device | POS | 1 | 6 |
| | NEG | 0 | 72 |
| | TOTAL | 1 | 78 |
| | The indeterminate samples were retested with the proposed device. | | |
| | The relative analytical specificity 92.3 % {72/78}, 0.95 INTERVAL 84.2 % to 96.4 % | | |
Table 3: Combined Panels, N = 117
| | Predicate Devices | | |
| --- | --- | --- | --- |
| | POS | NEG | TOTAL |
| Proposed Device | POS | 39 | 6 |
| | NEG | 0 | 72 |
| | TOTAL | 39 | 78 |
| | Relative sensitivity = 100 % {39/39}, 0.95 INTERVAL 91.0 % to 100 % | | |
| | Relative specificity 92.3 % {72/78}, 0.95 INTERVAL 84.2 % to 96.4 % | | |
510(k) Summary - Page 3
{3}
PAGE 4
# IV. A. Precision
To evaluate anti Cardiolipin Screen precision, inter-assay and intra-assay studies were conducted.
## Inter-assay precision
Three serum samples with various levels of anticardiolipin (negative, low positive, and high positive), the Cutoff Serum, and the Positive Control were assayed five times each, twice a day, on five different days:
| Sample | Mean OD | Std. Dev. | CV |
| --- | --- | --- | --- |
| Negative | 0.059 | 0.007 | 12.3 % |
| Low Positive | 0.197 | 0.023 | 11.7 % |
| High Positive | 0.967 | 0.067 | 6.9 % |
| Cutoff Serum | 0.115 | 0.015 | 13.0 % |
| Positive Control | 0.697 | 0.074 | 10.7 % |
## Intra-assay precision
Three serum samples with various levels of anticardiolipin (negative, low positive, and high positive) were assayed 16 consecutive times in a single run:
| Sample | Mean OD | Std. Dev. | CV |
| --- | --- | --- | --- |
| Negative | 0.069 | 0.010 | 15.0 % |
| Low Positive | 0.306 | 0.022 | 7.1 % |
| High Positive | 1.112 | 0.072 | 6.5 % |
## Conclusions
The results of the comparative studies support the claim that the Hemagen® anti Cardiolipin Screen is substantially equivalent to the comparative devices, and performs as an effective screening assay for the detection of autoantibodies to anticardiolipin IgA, IgG, and IgM in human serum.
510(k) Summary - Page 4
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
Mr. Joseph M. Califano
Manager, Regulatory Affairs
Hemagen Diagnostics, Inc.
34-40 Bear Hill Road
Waltham, Massachusetts 02154
Re: K971039
Trade Name: Hemagen® aCL Screen
Regulatory Class: II
Product Code: MID
Dated: March 20, 1997
Received: March 21, 1997
Dear Mr. Califano:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
{5}
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{6}
COPY
Device Name: Hemagen ® anti Cardiolipin Screen
Indication(s) For Use
The anti Cardiolipin Screen is indicated for the screening of patient serum for the presence of IgG, IgM, and IgA autoantibodies to cardiolipin. Autoantibodies to aCI are typically observed in patients with systemic lupus erythematosus and other connective tissue diseases. This assay is intended to be utilized as an aid in the diagnosis of antiphospholipid syndrome.
(PLEASE DO NOT WRITE BELOW THIS LINE)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter-Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.