K970865 · I.S.G. Technologies, Inc. · HAW · Jun 4, 1997 · Neurology
Device Facts
Record ID
K970865
Device Name
ISG FAMILY OF VIEWING WANDS
Applicant
I.S.G. Technologies, Inc.
Product Code
HAW · Neurology
Decision Date
Jun 4, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The Family of Viewing Wands, (including cranial and spinal applications) is comprised of a medical workstation and an integrated position sensing probe, is intended to be used pre-operatively and intra-operatively to perform the following functions: - process and display pre-operatively radiographic images on a monitor; - provide intra-operative image information based on the position and orientation of a user directed probe; - allow for the optical integration and usage of different position sensing technologies (e.g. articulated arm, infrared freehand) and surgical probes (handheld pointers); - store/retrieve image data on computer access media (e.g. hard disks, archive media); and - transmit data over local area and wide area networks.
Device Story
Medical imaging workstation integrated with position-sensing user-directed probe; correlates pre-operative CT/MR images with intra-operative patient anatomy. System displays probe tip position and orientation on images in real-time. Used by surgeons pre-operatively for surgical planning (trajectory, entry point, instrumentation sizing) and intra-operatively for navigation in regions with complex anatomy, distorted landmarks, or proximity to critical structures. Benefits include improved surgical guidance in subcortical brain or spinal procedures. Supports various sensing technologies (articulated arm, infrared freehand) and handheld pointers. Data stored on computer media and transmitted via networks.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use.
Technological Characteristics
Medical workstation with integrated position-sensing probe. Supports multiple sensing technologies including articulated arms and infrared freehand systems. Connectivity via local and wide area networks. Data storage on hard disks/archive media. Class II stereotactic instrument.
Indications for Use
Indicated for patients with space-occupying lesions or malformations (soft tissue/osseous) in the head, or patients requiring decompressive/reconstructive spinal surgery or having imaged space-occupying lesions/malformations in the spine. Contraindicated for patients suspected of Creutzfeld-Jakob’s disease if adequate instrument sterilization cannot be assured.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}
ISC
K970865
JUN - 4 1997
# 510(K) SUMMARY
## I. COMPANY & CONTACT PERSON
ISG TECHNOLOGIES, INC.
6509 Airport Road
Mississauga L4v-1S7
Ontario, Canada
## II. DEVICE NAME AND CLASSIFICATION
Family of Viewing Wands
Classification: Stereotactic Instrument (21 CFR 882.4560)
## III. DESCRIPTION OF DEVICE
The ISG Family of Viewing Wands is comprised of a medical imaging workstation that is integrated with a position-sensing user directed probe. This allows for the provision of information pertaining to the current position of the probe related to the surrounding anatomy by correlating pre-operative CT and/or MR imaging data with the intraoperative situation. Once a patient’s CT or MR images have been transferred to the workstation, the patient is correlated to that image dataset. The system then displays the orientation of the probe and the position of the tip on the corresponding image, updating the display in real-time as the probe is moved.
## IV. INTENDED USE
The Family of Viewing Wands, (including cranial and spinal applications) is comprised of a medical workstation and an integrated position sensing probe, is intended to be used pre-operatively and intra-operatively.
## V. SUBSTANTIAL EQUIVALENCE
The spinal application as described in this submission is a line (application) extension to the Family of Viewing Wands (K960714) and is substantially equivalent to other devices (Stealth Station - K954276) within the classification of diagnostic devices (21CFR 882.4560) that have been reviewed under the provisions of the Medical Device Amendments and are being marketed in interstate commerce.
COPYRIGHT 1997 ISG TECHNOLOGIES, INC.
ISG Technologies, Inc. 6509 Airport Road, Mississauga, Ontario, Canada M4V 1S7 • (905) 672-2100 1-800-724-5970 Fax (905) 672-2307
{1}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUN - 4 1997
Mr. Charles Hall
Director, Clinical Development
ISG Technologies, Inc.
6509 Airport Road
Mississauga, Ontario
CANADA L4V 1S7
Re: K970865
Trade Name: ISG Family of Viewing Wands
Regulatory Class: II
Product Code: 84HAW
Dated: March 4, 1997
Received: March 10, 1997
Dear Mr. Hall:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}
Page 2 - Mr. Charles Hall
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{3}
100 29 197 11:24
510(k) Number (if known): K970865
Device Name: ISG Family of Viewing Wands
## Indications For Use:
The Family of Viewing Wands, (including cranial and spinal applications) is comprised of a medical workstation and an integrated position sensing probe, is intended to be used pre-operatively and intra-operatively to perform the following functions:
- process and display pre-operatively radiographic images on a monitor;
- provide intra-operative image information based on the position and orientation of a user directed probe;
- allow for the optical integration and usage of different position sensing technologies (e.g. articulated arm, infrared freehand) and surgical probes (handheld pointers);
- store/retrieve image data on computer access media (e.g. hard disks, archive media); and
- transmit data over local area and wide area networks.
## Circumstances:
The Family of Viewing Wands is intended to be used pre-operatively:
- when more than one approach (e.g. entry point, trajectory, craniotomy size, screw placement) is being considered;
- when determining the size of any instrumentation;
continued...
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
[Signature]
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K970865
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
{4}
--/35 --2/3 --2/2/2.2 --29 37 11:24
...Circumstances, continued:
and intra-operatively, when:
- the target, trajectory, or approach is in a region with few anatomical landmarks (e.g. subcortical brain) or with complex anatomy (e.g. scoliosis, fractures, deformity);
- the target, trajectory, or approach is not entirely exposed (e.g. posterior approach for screw placement);
- the target, trajectory, or approach is in a region where the normal landmarks have been distorted either by disease (e.g. degenerative disease, osteophytes) or by previous surgery;
- the target, trajectory, or approach is in close proximity to critical structures (e.g. venous sinuses, air sinuses, spinal nerve roots or blood vessels) which must be avoided or negotiated; or
- the target’s delineation is important and will not move significantly during the approach (e.g. skull based tumors, sinus diseases, corpus callosotomies, spinal cord tumors).
**Disease or Conditions:**
The Family of Viewing Wands is indicated for patients who have space-occupying lesions or malformations (both soft tissue and osseous) in the head.
It is also indicated for patients who require decompressive or reconstructive surgery of the spine, or who have imaged space-occupying lesions or malformations in the spine.
The Family of Viewing Wands is contraindicated for patients suspected of having Creutzfeld-Jakob’s disease if adequate sterilization of the instrument cannot be assured.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.