K970782 · General Robotic Devices, Inc. · JAB · Nov 13, 1997 · Radiology
Device Facts
Record ID
K970782
Device Name
IMAGING JET SYSTEM
Applicant
General Robotic Devices, Inc.
Product Code
JAB · Radiology
Decision Date
Nov 13, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1660
Device Class
Class 2
Indications for Use
The Image Jet System provides physicians with a three dimensional image of an endoscope inside a patient.
Device Story
Imaging Jet is an endoscopic accessory providing 3D visualization of an endoscope's shape within a patient. The system comprises a sensor catheter, transport unit, signal processing unit, and computer. The sensor catheter, containing embedded sensors, is inserted into the endoscope's biopsy channel via an automated transport unit. As the catheter traverses the channel, it detects curvature and transmits electrical signals to the processing unit. The computer calculates the 3D spatial image of the endoscope and displays it on a screen for the physician. Used in clinical settings, the device assists physicians in determining scope position, loop direction, and lesion location. It serves as a radiation-free alternative to fluoroscopy, allowing image manipulation to facilitate clinical understanding and decision-making during endoscopic procedures.
Clinical Evidence
Bench testing only. Laboratory studies compared the Imaging Jet's calculated 3D images against actual colonoscope shapes in various configurations. Results demonstrated the system's capability to generate images with sufficient accuracy (±4% shape measurement accuracy) for its intended use. No human clinical trials were conducted.
Technological Characteristics
System includes a sensor catheter (rod-like), signal processing unit, and computer. Sensors embedded in the catheter tip measure curvature along the biopsy channel. Connectivity is via electrical signal transmission to a processing unit. Biocompatible materials used for patient-contacting components. Operates on electricity. No radiation source. Software performs 3D image reconstruction based on sensor input.
Indications for Use
Indicated for patients with difficult colons requiring colonoscopy and patients requiring small bowel endoscopy. Contraindications not specified.
Regulatory Classification
Identification
A non-image-intensified fluoroscopic x-ray system is a device intended to be used to visualize anatomical structures by using a fluorescent screen to convert a pattern of x-radiation into a visible image. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
X-ray fluoroscopy instrument by Picker International, Inc.
Endoscope and Accessories: Measuring Device by Olympus, Inc.
Submission Summary (Full Text)
{0}
GRD
General Robotic Devices, Inc.
2069 Golfside Drive • Ypsilanti, MI 48197 • Tel: 313-572-1102 • Fax: 313-572-1132
P1 of 16
Nov 13 1997
# 510(K) SUMMARY
Company:
General Robotic Device, Inc.
2069 Golfside Drive
Ypsilanti, Michigan 48197
K970782
Contact Person:
Yansong Shan, Ph.D.
Phone: 313-572-1102
Fax: 313-572-1132
Date of Summary Preparation: February 28, 1997
{1}
3 2 4 1 6
K970 782
# 1. DEVICE IDENTIFICATION
Device Name: Spatial Imaging Sensor for Endoscopy
Trade Name: Imaging Jet
Proprietary Name: Imaging Jet
Classification Name: Endoscope and accessories
21 CFR 876.1500
Classification: Class II
1-1
{2}
P3 of 16
K970782
## 2. Predicate Devices
1. X-ray fluoroscopy instrument by Picker International, Inc. (Cleveland, OH)
2. Endoscope and Accessories: Measuring Device by Olympus, Inc.
2-1
{3}
P40816
K970782
# 3. Description of the Device
The Imaging Jet is an endoscopic accessory that provides a spatial view of an endoscope inside a patient. When in use, the sensor catheter is inserted into the biopsy channel of an endoscope inside a patient. The sensor catheter measures the shape of a scope and sends sensory signals to a computer for calculating and displaying the image on the computer screen.
A complete Imaging Jet system consists of the following major components as shown in: (1). a computer which governs the operation of the whole system, calculates the scope spatial image and displays the image; (2). a signal processing unit which acquires and processes the sensory signals and supplies them to the computer for image calculation; (3). a sensor catheter which is a long flexible rod with sensors embedded inside in its head portion; (4). a sensor catheter transport unit which is used to insert and withdraw sensor catheter into and from the biopsy channel of a scope; and (5). a mobile stand for transporting the whole system.

Figure 3-1. System overview
{4}
P5 of 16
K970782
The signal processing unit houses all the electronics and power supplies to the system. The unit measures $400 \, \text{mm} \times 300 \, \text{mm} \times 100 \, \text{mm}$ ($L \times W \times H$) and weighs about 3 kg. The signal processing unit is shown in Figure 3-2.

Figure 3-2. Signal processing unit
The sensor catheter consists of two segments. At the tip of the catheter, there is a sensor head segment where sensors are embedded. The rest is used to provide the necessary length. The sensor catheter is shown in Figure 3-3. During application, only the sensor head and part of the insertion tube are inside the endoscope, the rest of the catheter including the rear part of the tube and sensor catheter connector are outside the scope (They remained inside the transport unit).

Figure 3-3. The sensor catheter
{5}
P 6 a 16 K970782
The sensor catheter transport unit is used to automatically insert and withdraw sensor catheter into and from the biopsy channel of a colonoscope. This will reduce the work load on the physician, meanwhile maintains a smooth sensor motion that will result in good quality of sensor signals. The sensor catheter should be loaded into the transport unit prior to an operation by doctor's assistant. This unit also generates feedback to the control unit so that the actual insertion motion of the catheter can be monitored and controlled. The transport unit is shown in Figure 3-4.

Figure 3-4. The sensor catheter transport unit
The sensor catheter transport unit delivers the sensor catheter through the biopsy channel of an endoscope. During insertion, the catheter detects curvature of the path of the biopsy channel and sends electrical signals to the signal processing unit. The computer interfaces with the signal processing unit and calculates the image of the endoscope. After the computer calculates the image of the endoscope, it displays the image onto the computer screen in a three dimensional manner.
{6}
P 7 of 16
K970782
# 4. Statement of Intended Uses
The Imaging Jet System provides physicians with a three dimensional image of an endoscope inside a patient.
{7}
P8 8 16 K970782
# 5. Substantial Equivalence Comparison
Table 5-1. Substantial Equivalence Comparison Table
| | Imaging Jet | Predicate device Flu-roscopy | Predicate device Measuring Device |
| --- | --- | --- | --- |
| Intended use | determine (1) overall image of a scope, (2) existence and direction of loops, (3) number of loops, and (4) location of lesion. | determine (1) overall image of a scope, (2) existence and direction of loops, (3) number of loops, and (4) location of lesion. | Measure the size of the polyp. |
| Technology characteristics | This device uses sensors to measure the curvature of a scope along the insertion length while the catheter traveling through the endoscope biopsy channel. The 3D position estimation is obtained from the sensor signals of the catheter. Then the 3D scope image is displayed on a computer screen using computer graphics techniques. | The x-ray cameras are built based on the penetration of the x-ray into a patient. The scope inside the patient blocks the x-rays that directly fall to the scope. Therefore, there is a difference on radiation intensity on the receiver. The x-ray image can be formed by displaying the image on to a screen. The screen can directly show the image of the scope inside the patient. | This device allows doctors to measure the size of a polyp by compare it with the measuring marks on the device. |
| Method of application | The sensor catheter is inserted through the biopsy channel of the endoscope in concern and reaches the distal end of the endoscope. | An x-ray camera is out-side the patient. | The measuring device is inserted through the biopsy channel of the endoscope in concern and reaches the distal end of the endoscope. |
| Target population | Patients with “difficult” colon while needing colonoscopy and most patients who need small bowel endoscopy. | Patients with “difficult” colon while needing colonoscopy and most patients who need small bowel endoscopy. | Patients with polyps and the size of the polyps is in concern. |
| Biocompatibility | All exterior materials are biocompatible. | Not applicable. | All exterior materials are biocompatible. |
| Performance | (1) The shape meas-urement accuracy of this device is ±4%. This accuracy is suffi-cient for its intended use. | (1) The accuracy of po-sition estimation is at best 10 mm, but in fact much worse. (2) Multiple pictures need to be taken to pro- | Provide quantitative measure of the size of a polyp. |
{8}
P 9 of 16
K970782
| | (2) Provide a three-dimensional image in one operation.
(3) User can manipulate the image to facilitate understanding. | vide three dimensional information.
(3) In general, while using fluoroscopy to determine the location of an endoscope, a physician uses a single two-dimensional image and attempts to extrapolate a three dimensional image. The error in this extrapolation is great, several centimeters at least. This error is increased by changing rate of magnification as the x-ray beam travels from the x-ray emitter to the x-ray receiver. | |
| --- | --- | --- | --- |
| **Radiation safety** | No radiation hazard. | Radiation health hazard to patients, operators, and other personnel in the room. Possible damage to the scope overtime. | No radiation hazard. |
| **Sterility and cleaning** | Cleaning and disinfection are needed after each use. | Cleaning and sterilization are not needed. | Cleaning and disinfection are needed after each use. |
| **Energy used** | Electricity | Electricity | No energy source needed |
| **Where used** | Any doctor’s office | Hospital or major clinics | Any doctor’s office |
| **Physical appearance** | The appearance of the sensor catheter can be characterized as a “rod”. | An x-ray emitter and a receiver. | The appearance of same this device can be characterized as a “rod”. |
In summary, the Imaging Jet has the same intended use as that of the x-ray fluoroscopy instrument for endoscopy procedures. The effectiveness of this device is established through thorough mathematical analysis. The safety is established through material selection, system design, manufacturing quality control, and mathematical proof. This system provides similar information as an x-ray fluoroscopy instrument and offers several advantages over an x-ray fluoroscopy instrument: (1) There is no radiation hazard to patients and operators, (2) Low cost in both facility and usage, (3) Easy to use, no need to transport patients back and forth between x-ray unit and endoscopy unit, (4) It provides a full three-dimensional image, and (5) Users can manipulate the image to facilitate understanding. As a result of the above analysis, we consider the Imaging Jet is substan-
{9}
F 10 of 16
K970782
tially equivalent to the x-ray fluoroscopy instrument for its intended use and the Image Jet is at least as safe and effective as the x-ray fluoroscopy.
The common technical characteristics of both Imaging Jet device and the Endoscope Measuring Device by Olympus can be summarized as: (1) When in use, the catheters of both the Endoscope Measuring Device by Olympus and the Imaging Jet are inserted into the biopsy channel of an endoscope. (2) Both devices contain no invasive tools, e.g., tools for biopsy forceps, at their distal ends. Their physical appearances can both be characterized as “rod”. (3) During use, both devices may contact the interior tissue of a patient. The safety of both devices is ensured through the biocompatibility of the materials selected to construct the devices.
Therefore it is concluded that the Imaging Jet device achieves the same intended use as the predicate device (x-ray fluoroscopy instrument by Picker) and it is as safe and effective as the predicate devices (x-ray fluoroscopy instrument by Picker and Endoscope and Accessories: Measuring Device by Olympus, Inc.).
5-3
{10}
P11 of 16
K970782
# 6. Performance
Laboratory tests were conducted to test the performance of the Imaging Jet system in terms of the overall image calculation capability, we used a colonoscope to form different shapes. Then we used the Imaging Jet system to measure these shapes. The following figures illustrate the comparison between the actual scope images and the measured images using Imaging Jet.

(a1) Actual scope image

(a2) Measured image using Imaging Jet
6-1
{11}
P17 of 16
K970782

(b1) Actual scope image

(b2) Measured image using Imaging Jet

(c1) Actual scope image

(c2) Measured image using Imaging Jet
6-2
{12}
P13 of 16
K976782

(d1) Actual scope image

(d2) Measured image using Imaging Jet

(e1) Actual scope image

(e2) Measured image using Imaging Jet
6-3
{13}
P 14 of 16
1<970>

(f1) Top view of actual scope image

(f2) Top view of measured image using Imaging Jet
6-4
{14}
P15 of 16
K970782

(f3) Side view of actual scope image

(f4) Side view of measured image using Imaging Jet
Figure 6-1. Image comparison
From the above figures, we conclude that the Imaging Jet system is capable of generating images with sufficient accuracy for its intended use.
{15}
P16 of 16
K970782
# 7. Conclusions
In summary, the Imaging Jet has the same intended use as that of the x-ray fluoroscopy instrument for endoscopy procedures. The effectiveness of this device is established through thorough mathematical analysis and laboratory tests. The safety is established through material selection, system design, manufacturing quality control, and mathematical proof.
Therefore it is concluded that the Imaging Jet device achieves the same intended use as the predicate device (x-ray fluoroscopy instrument by Picker) and it is as safe and effective as the predicate devices (x-ray fluoroscopy instrument by Picker and Endoscope and Accessories: Measuring Device by Olympus, Inc.).
7-1
{16}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Yansong Shan, Ph.D.
President
General Robotic Devices, Inc.
2069 Golfside Drive
Ypsilanti, Michigan 48197
Re: K970782
Imaging Jet System
Dated: October 10, 1997
Received: October 20, 1997
Regulatory class: II
21 CFR §876.1500/Product code: 78 KOG
21 CFR §892.1600/Product code: 78 JAB
NOV 13 1997
Dear Dr. Shan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{17}
Page 1 of 1
510(k) Number (if known): K970782
Device Name: Imaging Jet System
Indications For Use:
The Image Jet System provides physicians with a three dimensional image of an endoscope inside a patient.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Robert P. Stilling
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K970782
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.