SIREGRAGH T.O.P. 33 AND SIREGRAPH T.O.P. 40 UNIVERSAL FLUOROSCOPIC X-RAY SYSTEMS

K970734 · Siemens Medical Solutions USA, Inc. · OWB · Apr 21, 1997 · Radiology

Device Facts

Record IDK970734
Device NameSIREGRAGH T.O.P. 33 AND SIREGRAPH T.O.P. 40 UNIVERSAL FLUOROSCOPIC X-RAY SYSTEMS
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeOWB · Radiology
Decision DateApr 21, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1650
Device ClassClass 2
AttributesPediatric

Indications for Use

The Siregraph T.O.P. 33 and T.O.P. 40 Universal Fluoroscopic X-ray Systems are devices intended to visualize anatomical structures by converting a pattern of X-ray into a visible image through electronic amplification. Both systems have medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examination of the urogenital tract, lymphography, endoscopy, myelography, venography, pediatrics, arthrography, interventional radiology, digital angiography, and digital subtraction angiography (DSA).

Device Story

Universal fluoroscopic X-ray system with overtable tube assembly; available as T.O.P. 33 (undertable spot film) or T.O.P. 40 (Digital Fluoro Radiography with undertable Image Intensifier). System converts X-ray patterns into visible images via electronic amplification. Operated by clinicians via table-side controls or remote console. Components include patient support table, X-ray generator, tube, image intensifier, TV system, digital imaging system, and monitors. Used in clinical settings for diagnostic and interventional procedures. Updated XCS communication network facilitates component interaction. Output displayed on monitors for real-time visualization and clinical decision-making during procedures.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Universal fluoroscopic X-ray system; overtable X-ray tube; undertable spot film (T.O.P. 33) or Image Intensifier (T.O.P. 40). Includes X-ray generator, TV system, digital imaging system, and monitors. Features XCS communication network for component connectivity. Ergonomic design. Class II device.

Indications for Use

Indicated for visualization of anatomical structures in patients of all ages, including pediatric, requiring gastrointestinal, cranial, skeletal, thoracic, lung, urogenital, lymphographic, endoscopic, myelographic, venographic, arthrographic, interventional, or angiographic (including DSA) examinations.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0} APR 21 1997 K970734 # SIEMENS ## Attachment 12 ### 510(k) Summary **Siregraph T.O.P. 33 and Siregraph T.O.P. 40** Submitted by: Siemens Medical Systems, Inc. 186 Wood Avenue South Iselin, NJ 08830 February 27, 1997 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. 1. Contact Person: Ms. Jing Zhang Phone: (908)321-4927 Fax: (908)321-4841 2. Device Name and Classification: Trade Name: Siregraph T.O.P. 33 and Siregraph T.O.P. 40 Universal Fluoroscopic X-ray Systems Classification Name: Image Intensified Fluoroscopic X-ray System Classification Panel: Radiology CFR Section: 21 CFR §892.1650 Device Class: Class II Device Code: 90JAA ↑ OWB 3. Intended Use: The Siregraph T.O.P. 33 and T.O.P. 40 Universal Fluoroscopic X-ray Systems are devices intended to visualize anatomical structures by converting a pattern of X-ray into a visible image through electronic amplification. Both systems have medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examination of the urogenital tract, lymphography, endoscopy, myelography, venography, pediatrics, arthrography, interventional radiology, digital angiography, and digital subtraction angiography (DSA). 4. Substantial Equivalence: The Siregraph T.O.P. 33 and 40 are substantially equivalent to the following devices in commercial distribution: Simplified 510(k) for Siregraph T.O.P. 33 & 40 Siemens Medical Systems, Inc. Pg. 57 of 58 {1} # SIEMENS - Siregraph D3 and D340, K860913 Siemens Medical Systems, Inc. - Siregraph CF, K960266 Siemens Medical Systems, Inc. ## 5. Device Description: Siregraph T.O.P. is a universal fluoroscopic X-ray diagnostic system with an overtable X-ray tube assembly. Two versions are available: Siregraph T.O.P. 33 for use with an undertable spot film device, and Siregraph T.O.P. 40 for use in Digital Fluoro Radiography (DFR) with an undertable Image Intensifier (I.I.). Both systems are operated either via table side control or the remote control console. ## 6. Summary of Technological Characteristics of the Device Compared to the Predicate Devices: Siregraph T.O.P. has the same technological characteristics as the predicate Siregraph D3/D340. Both systems are remote fluoroscopic X-ray diagnostic systems with an overtable X-ray tube assembly. Like Siregraph D3 and D340, Siregraph T.O.P. consists of the basic system (patient support table), and standard system components: X-ray generator, X-ray tube, Image Intensifier, TV system, digital imaging system, monitors, optional Bucky wall stand and optional ceiling-mounted support for second X-ray tube. The differences between Siregraph D3/D340 and Siregraph T.O.P. are: - The Siregraph T.O.P. basic system is redesigned for better ergonomics. - The communication network among system components (e.g., generator, tube, etc.) has been updated. The XCS communication system is employed. - Siregraph T.O.P. are configured with the latest commercially available system components. ![img-0.jpeg](img-0.jpeg) Simplified 510(k) for Siregraph T.O.P. 33 & 40 Siemens Medical Systems, Inc. Pg. 58 of 58 {2} HUMAN SERVICES, INC. DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 Ms. Kathleen Rutherford Manager, Regulatory Submissions Siemens Medical Systems, Inc. 186 Wood Avenue South ISELIN NJ 08830 FEB 19 2013 Re: K970734 Trade/Device Name: Siregraph T.O.P. 33 and Siregraph T.O.P 40 Universal Fluoroscopic X-ray System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB and JAA Dated: February 27, 1997 Received: February 28, 1997 Dear Ms. Rutherford: This letter corrects our substantially equivalent letter of April 21, 1997. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {3} Page 2 medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of *In Vitro Diagnostic Device Evaluation and Safety* at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, ![img-1.jpeg](img-1.jpeg) Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4} Attachment 1 Indications for Use 510(k) Number (if known): K970734 Device Name: Siregraph T.O.P. 33 and Siregraph T.O.P. 40 Indications for Use: The Siregraph T.O.P. 33 and Siregraph T.O.P. 40 Universal Fluoroscopic X-ray Systems are devices intended to visualize anatomical structures by converting a pattern of X-ray into a visible image through electronic amplification. Both systems have medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examination of the urogenital tract, lymphography, endoscopy, myelography, venography, pediatrics, arthrography, interventional radiology, digital angiography, and digital subtraction angiography (DSA). Concurrence of the CDRH, Office of Device Evaluation (ODE) David C. Segman (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K970734 Prescription Use ☑ OR Over-The-Counter Use ☐ (per 21 CFR 801.109) Simplified 510(k) for Siregraph T.O.P. 33 & 40 Siemens Medical Systems, Inc.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...