← Product Code [OCW](/productcode/OCW) · K970720

# ENDOPATH EZ45 ENDOSCOPIC LINEAR CUTTER AND THE ENDOPATH EZ45 NO KNIFE ENDOSCOPIC LINEAR STAPLER (K970720)

_Ethicon Endo-Surgery, Inc. · OCW · Jun 27, 1997 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K970720

## Device Facts

- **Applicant:** Ethicon Endo-Surgery, Inc.
- **Product Code:** [OCW](/productcode/OCW.md)
- **Decision Date:** Jun 27, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The ENDOPATH® EZ45 Endoscopic Linear Cutter [and the ENDOPATH® EZ45 No Knife Endoscopic Linear Stapler] has application in multiple open and other minimally invasive surgical procedures for transection, resection, [and/or creation of anastomoses] and can be used with staple line or tissue buttressing materials, such as bovine pericardium.

## Device Story

The ENDOPATH® EZ45 is a sterile, single-patient-use endoscopic linear stapler. It delivers two double-staggered rows of staples while simultaneously dividing tissue (linear cutter version). The device features a safety lock-out to prevent firing an unloaded instrument or reusing a spent reload. It is reloadable up to seven times (eight total firings). The device is used by surgeons in open or minimally invasive procedures. It can be used with staple line or tissue buttressing materials (e.g., bovine pericardium). The device provides mechanical tissue management, facilitating surgical transection and anastomosis.

## Clinical Evidence

Bench testing only. Pre-clinical laboratory evaluations were performed to ensure the devices function as designed for transection, resection, and anastomosis, including use with buttressing materials.

## Technological Characteristics

Sterile, single-patient-use endoscopic linear stapler. Features staple line length of ~45 mm and cut line of ~41 mm. Reloadable (up to 8 firings). Includes safety lock-out mechanism. Compatible with thin (vascular), standard, and thick tissue reloads. Mechanical operation.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

JUN 27 1997

# Appendix A: 510(k) Summary of Safety and Effectiveness

k970720

## Statement

Information supporting claims of substantial equivalence, as defined under the Federal Food, Drug and Cosmetic Act, respecting safety and effectiveness is summarized below.

For the convenience of the Reviewer, this summary is formatted in accordance with the Agency's final rule"...510(k) Summaries and 510(k) Statements..." (21 CFR §807) and can be used to provide a substantial equivalence summary to anyone requesting it from the Agency.

## Device description

The ENDOPATH® EZ45 Endoscopic Linear Cutter is a sterile, single patient use instrument that delivers two, double-staggered rows of staples while simultaneously dividing the tissue between rows. [The ENDOPATH® EZ45 No Knife Endoscopic Linear Stapler is a sterile, single patient use instrument that delivers two, double-staggered rows of staples.]

The instrument is designed for use in procedures that do not require insufflation. The instruments' safety lock-out feature is designed to prevent firing an unloaded instrument or prevent a used reload from being refired. The instruments have a staple line that is approximately 45 mm long and a cut line [linear cutter only] that is approximately 41 mm long. A staple retaining cap on the reload protects the staple leg points during shipping and transportation.

The instrument is reloadable with a thin (vascular), white reload; a standard, blue reload; or a thick tissue, green reload. Do not reload the instrument more than seven times for a maximum of eight firings per instrument. The use of the instrument with staple line buttressing materials may reduce the number of firings.

## Intended use

For transection, resection, [and/or the creation of anastomoses].

## Indications statement

The ENDOPATH® EZ45 Endoscopic Linear Cutter [and the ENDOPATH® EZ45 No Knife Endoscopic Linear Stapler] has application in multiple open and other minimally invasive surgical procedures for transection, resection, [and/or creation of anastomoses] and can be used with staple line or tissue buttressing materials, such as bovine pericardium.

Continued on next page

ETHICON ENDO-SURGERY, INC.

ENDOPATH® EZ45 Endoscopic Linear Cutter and ENDOPATH® EZ45 No Knife Endoscopic Linear Stapler

{1}

# Appendix A: 510(k) Summary of Safety and Effectiveness, Continued

|  Technological characteristics | The technological characteristics of the New Devices are the same as the Predicate Device.  |
| --- | --- |
|  Performance data | Pre-clinical laboratory evaluations were performed to ensure that the devices can be used as designed. The studies demonstrated acceptable performance in transecting, resecting, and/or creation of anastomoses and for use with buttressing materials.  |
|  Conclusion | Based on the 510(k) summaries and 510(k) statements (21 CFR §807) and the information provided herein, we conclude that the New Devices are substantially equivalent to the Predicate Device under the Federal Food, Drug and Cosmetic Act.  |
|  Contact | Lorri Chavez
Ethicon Endo-Surgery, Inc.
4545 Creek Road
Cincinnati, Ohio 45242  |
|  Date | February 25, 1997  |

ETHICON ENDO-SURGERY, INC.
ENDOPATH® EZ45 Endoscopic Linear Cutter and ENDOPATH® EZ45 No Knife Endoscopic Linear Stapler

{2}

DEPARTMENT OF HEALTH &amp; HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850

Ms. Lorri Chavez
Project Manager, Regulatory Affairs
Ethicon Endo-Surgery, Inc.
4545 Creek Road
Cincinnati, Ohio 45242-2839

Re: K970720
Trade Name: ENDOPATH® EZ45 Endoscopic Linear Cutter and ENDOPATH® EZ45 No Knife Endoscopic Linear Stapler
Regulatory Class: II
Product Code: KOG
Dated: May 21, 1997
Received: June 10, 1997

Dear Ms. Chavez:

We have reviewed your Section 510(k) notification of intent to market the devices referenced above and we have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your premarket notification submissions does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}

Page 2 - Ms. Lorri Chavez

This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices to proceed to the market.

If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

![img-0.jpeg](img-0.jpeg)

Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health

Enclosure

{4}

# Appendix B: Indications for Use Statement

## Statement

Indications for Use Statement:

510(k) Number: K 970720

Device Names: The ENDOPATH® EZ45 Endoscopic Linear Cutter and the ENDOPATH® EZ45 No Knife Endoscopic Linear Stapler.

Indications for Use: The ENDOPATH® EZ45 Endoscopic Linear Cutter [and the ENDOPATH® EZ45 No Knife Endoscopic Linear Stapler] has application in multiple open and other minimally invasive surgical procedures for transection, resection, [and/or creation of anastomoses] and can be used with staple line or tissue buttressing materials, such as bovine pericardium.

![img-1.jpeg](img-1.jpeg)

![img-2.jpeg](img-2.jpeg)

ETHICON ENDO-SURGERY, INC.
ENDOPATH® EZ45 Endoscopic Linear Cutter and ENDOPATH® EZ45 No Knife Endoscopic Linear Stapler

---

**Source:** [https://fda-staging.innolitics.com/device/K970720](https://fda-staging.innolitics.com/device/K970720)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).