COMPUTED TOMOGRAPHY X-RAY SYSTEM

K970606 · GE Medical Systems · JAK · Apr 4, 1997 · Radiology

Device Facts

Record IDK970606
Device NameCOMPUTED TOMOGRAPHY X-RAY SYSTEM
ApplicantGE Medical Systems
Product CodeJAK · Radiology
Decision DateApr 4, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2

Indications for Use

The CT ProSpeed Family of CT Systems with ProSpeed Renaissance Options is indicated for head and whole body x-ray computed tomography applications. These options are modifications which can be added to the existing family of ProSpeed Systems. The options include Warp Scan Option which is intended to provide 0.8 second 360-degree helical or axial scans. Fluoro Scan Option is an almost "real time" reconstruction (Smart Recon) with a display latency of 0.8 to 1.6 seconds. An in-room monitor provides an image display with either a 3 frame/sec or 6 frame/sec display of 90 one second rotation scans. Controls are provided for the control of the table and gantry by a physician or their assistant during procedures that require imaging. SmartPrep Option is a tool to aid in triggering a helical or axial series of scans after contrast is injected. It allows the operator to monitor contrast as it enhances. A series of images, acquired immediately after injection, is automatically displayed along with pre-selected regions of interest. A graph plots the increase in CT numbers. When the region of interest begins approaching peak enhancement the operator can see that happen on the graph and initiate the CT series.

Device Story

CT ProSpeed Family with Renaissance options are x-ray computed tomography scanners; consist of gantry, patient table, console, and computer. Warp Scan provides 0.8s helical/axial scans. Fluoro Scan provides near real-time reconstruction (Smart Recon) with 0.8-1.6s latency; includes in-room monitor and control pedestal for physician/assistant operation of gantry/table during procedures. SmartPrep monitors contrast enhancement; displays images and ROI graphs of CT numbers to assist operator in triggering diagnostic scan series at peak enhancement. Used in clinical settings by physicians/technicians. Output is CT imagery and enhancement graphs to guide clinical decision-making and scan timing.

Clinical Evidence

No clinical data. Safety and effectiveness demonstrated through bench testing, hazard analysis, and system evaluation against established performance specifications and federal/international standards (UL 187, IEC 601-1, 21 CFR Subchapter J).

Technological Characteristics

X-ray CT scanner; gantry, table, console, computer. Materials compliant with UL 187, IEC 601-1, 21 CFR Subchapter J. Features: Warp Scan (0.8s rotation), Fluoro Scan (Smart Recon, 3-6 fps display), SmartPrep (contrast monitoring/triggering).

Indications for Use

Indicated for head and whole body x-ray computed tomography applications in patients requiring diagnostic imaging.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0} APR - 4 1997 GE Medical Systems General Electric Company P.O. Box 414 Milwaukee, WI 53201 K970606 # Summary of Safety & Effectiveness This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h) **Submitter:** Larry A. Kroger, Ph.D. Regulatory Programs Manager Who may be contacted by telephone at 414-544-3894 or by FAX at 414-544-3863. Summary prepared 14 February 1997 ## Product Identification **Name:** CT ProSpeed Family of CT Systems with ProSpeed Renaissance Options **Manufacturer:** GE-YMS 7-127 Asahigaoka 4-Chome Hino-Shi, Tokyo, Japan 191 **Distributor:** GE Medical Systems 3000 Grandview Waukesha, Wi 53188 ## Marketed Devices: The CT ProSpeed Family of CT Systems with ProSpeed Renaissance Options is of comparable type and substantially equivalent to currently marketed Computed Tomography X-ray Systems that comply with the same or equivalent standards and have the same intended uses. ## Device Description: The ProSpeed Family of Systems are x-ray computed tomography scanners based on the ProSpeed Plus platform consisting of a gantry, patient table, console, computer and associated accessories. **Materials:** Materials and construction are equivalent to the CT ProSpeed Plus and are compliant with UL 187, IEC 601-1, and 21 CFR Subchapter J. **Design:** The design is essentially the same as the CT ProSpeed with HiLight detector (K944013) the difference being that the Fluoro Scan option includes an in-room monitor and a control pedestal. The control pedestal allows for gantry tilt and table positioning as well as initiation and termination of scanning. **Energy Source and Exposure Levels:** The energy source and exposure levels are the same as those previously submitted for ProSpeed with HiLight Detector. **Principals of Operation:** The same as ProSpeed with HiLight Detector. 100 Years of Radiology 1895 - 1995 {1} # Indications for Use: The CT ProSpeed Family of CT Systems with ProSpeed Renaissance Options is indicated for head and whole body x-ray computed tomography applications. These options are modifications which can be added to the existing family of ProSpeed Systems. The options include Warp Scan Option which is intended to provide 0.8 second 360-degree helical or axial scans. Fluoro Scan Option is an almost "real time" reconstruction (Smart Recon) with a display latency of 0.8 to 1.6 seconds. An in-room monitor provides an image display with either a 3 frame/sec or 6 frame/sec display of 90 one second rotation scans. Controls are provided for the control of the table and gantry by a physician or their assistant during procedures that require imaging. SmartPrep Option is a tool to aid in triggering a helical or axial series of scans after contrast is injected. It allows the operator to monitor contrast as it enhances. A series of images, acquired immediately after injection, is automatically displayed along with pre-selected regions of interest. A graph plots the increase in CT numbers. When the region of interest begins approaching peak enhancement the operator can see that happen on the graph and initiate the CT series. # Comparison with Predicate: It is the opinion of GE Medical Systems that the CT ProSpeed Family of CT Systems with ProSpeed Renaissance Options is of a comparable type and substantially equivalent to currently marketed head and whole body x-ray computed tomography systems with respect to design, material composition, energy source, and radiation characteristics. # Adverse Effects on Health: Potential electrical, mechanical and radiation hazards are identified in a hazard analysis and controlled by: - Fault Tree Analysis to demonstrate the non-existence or extremely low probability of unwanted events. - System evaluation to insure performance to specifications and Federal Regulations. - Adherence to Industry and International Standards. (UL and IEC) # Conclusions: Use of the CT ProSpeed Family of CT Systems with ProSpeed Renaissance Options does not result in any new potential safety risks and performs as well as or better than devices currently on the market.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...