← Product Code [MSC](/productcode/MSC) · K970577

# GLYDE DAM LOLLYES (K970577)

_Glyde USA, Inc. · MSC · Jan 8, 1998 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K970577

## Device Facts

- **Applicant:** Glyde USA, Inc.
- **Product Code:** [MSC](/productcode/MSC.md)
- **Decision Date:** Jan 8, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

Glyde Dam Lollyes, when properly used, may help reduce the risk of catching or spreading many Sexually Transmitted Diseases (“STDs”) such as syphilis, gonorrhea, chlamydia infections, genital herpes, and AIDS; however, they cannot eliminate the risk. For maximum benefits, it is important to follow the instructions for use printed on the packaging. Failure to do so may result in the loss of the benefits of the Glyde Dam Lollyes. During intimate contact, lesions and various bodily fluids can transmit STDs. Therefore, the Glyde Dam Lollyes should be applied each and every time before any such contact occurs. The Glyde Dam Lollyes is removed from the wrapper and laid flat. The Glyde Dam Lollyes is placed over the entire vulva, covering both the vaginal opening and the clitoris, while holding the ends of the Glyde Dam Lollyes. Optional: one side may be moistened with a commercially available water-based lubricant. The lubricated side of Dam Lollyes is then placed over the entire vulva, covering both the vaginal opening and the clitoris, while holding the ends of the Glyde Dam Lollyes. The Glyde Dam Lollyes is intended for single use only. Users are instructed not to re-use Glyde Dam Lollyes because of possible cross-contamination. Users are further instructed not to stretch the Glyde Dam Lollyes, and instructed to store them in a cool dry place at room temperature (59 degrees to 86 degrees Fahrenheit) and away from direct exposure to sunlight.

## Device Story

Glyde Dam Lollyes are natural rubber latex sheets (10" x 6" x 0.08mm) used as a physical barrier during cunnilingus. Device is applied by the user over the vulva, covering the vaginal opening and clitoris, to reduce transmission of STDs via bodily fluids or lesions. Single-use only; non-sterile; intended for over-the-counter use. Benefits include reduced risk of STD transmission; however, risk is not eliminated. Proper use requires avoiding stretching and following storage instructions.

## Clinical Evidence

Bench testing only. Biocompatibility, tensile strength/elongation, and virus testing performed to support safety and effectiveness claims.

## Technological Characteristics

Natural rubber latex conforming to ASTM 1076-88. Dimensions: 10.0" x 6.0" x 0.08mm. Physical barrier principle. Single-use. Non-sterile.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- Trojan Brand Latex Condoms

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510k Notification
P194
CONFIDENTIAL
K970577
JAN - 8 1998

# APPENDIX E

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS FOR GLYDE DAM LOLLYES

13 February 1997
Glyde USA, Inc. - Glyde Dam Lollyes™
E
20

{1}

510k Notification
p2074
CONFIDENTIAL

## Submitter

Glyde USA, Inc.
1758 Dexter Avenue North, Suite 3
Seattle, WA 98109 U.S.A.

Telephone Number: 206.283.7664
Facsimile: 206.284.7554
Contact Person: Barbara Lippert, President

**Date summary was prepared**

02.13.97

**Name(s) of the device**

Proprietary (Trade) Name: Glyde Dam Lollyes
Common or Usual Name: Latex dam

**Identification of predicate device(s)**

Trojan Brand Latex Condoms
Carter Products, Division of Carter Wallace, Inc.
New York, N. Y.

13 February 1997
Glyde USA, Inc. - Glyde Dam Lollyes™

E 1
20

{2}

510k Notification
P394
CONFIDENTIAL

## Description of the device

Glyde Dam Lollyes are manufactured from good quality rubber latex conforming to ASTM 1076-88. The size of each Glyde Dam Lollyes is 10.0 inches (± 0.1 inch) by 6.0 inches (± 0.1 inch) by 0.08 millimeters thick (± 0.0006 millimeters.) Glyde Dam Lollyes are designed to be used while performing cunnilingus.

Qualification testing (biocompatibility, tensile strength/elongation, and virus) has been performed to satisfactorily conclude that Glyde Dam Lollyes operate as intended, when used properly. Finally, Glyde Dam Lollyes are tested for a smooth surface (i.e.: the presence of cracks and blisters,) and are visually tested for holes, tears, foreign materials and the like, during the manufacturing process of each lot.

## Intended Use

Glyde Dam Lollyes, when properly used, may help reduce the risk of catching or spreading many Sexually Transmitted Diseases (“STDs”) such as syphilis, gonorrhea, chlamydia infections, genital herpes, and AIDS; however, they cannot eliminate the risk. For maximum benefits, it is important to follow the instructions for use printed on the packaging. Failure to do so may result in the loss of the benefits of the Glyde Dam Lollyes. During intimate contact, lesions and various bodily fluids can transmit STDs. Therefore, the Glyde Dam Lollyes should be applied each and every time before any such contact occurs.

The Glyde Dam Lollyes is removed from the wrapper and laid flat. The Glyde Dam Lollyes is placed over the entire vulva, covering both the vaginal opening and the clitoris, while holding the ends of the Glyde Dam Lollyes. Optional: one side may be moistened with a commercially available water-based lubricant. The lubricated side of Dam Lollyes is then placed over the entire vulva, covering both the vaginal opening and the clitoris, while

13 February 1997
Glyde USA, Inc. - Glyde Dam Lollyes™
E 2
205

{3}

510k Notification
p404 CONFIDENTIAL

holding the ends of the Glyde Dam Lollyes. The Glyde Dam Lollyes is intended for single use only. Users are instructed not to re-use Glyde Dam Lollyes because of possible cross-contamination. Users are further instructed not to stretch the Glyde Dam Lollyes, and instructed to store them in a cool dry place at room temperature (59 degrees to 86 degrees Fahrenheit) and away from direct exposure to sunlight.

## Comparison of device characteristics to predicate

A comparison was made between the physical and performance characteristics of the Glyde Dam Lollyes and the physical and performance characteristics of a legally marketed predicate product. This analysis revealed that Glyde Dam Lollyes are substantially equivalent to that predicate product.

## Non clinical testing

Biocompatibility, tensile strength/elongation and virus testing was performed on the Glyde Dam Lollyes. All testing was supportive of the claims (i.e.: indications for use.)

## Conclusion

In conclusion, the basis for substantial equivalence between the Glyde Dam Lollyes and the legally marketed predicate product is that the products are predominantly similar and that the differences between the two products do not raise new issues of safety and effectiveness. Based upon the information provided herein, it is our conclusion that the Glyde Dam Lollyes are substantially equivalent to the Trojan Brand Condoms.

13 February 1997
Glyde USA, Inc. - Glyde Dam Lollyes™
E 3
201

{4}

DEPARTMENT OF HEALTH &amp; HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JAN - 8 1998

Glyde USA, Inc.
c/o Ms. Louise C. Myers
Technical Adviser
Medical Technology Consultants
14808 N.E. 66th Street
Redmond, Washington 98052-4712

Re: K970577
Glyde Dam Lollyes™
Dated: October 21, 1997
Received: October 23, 1997
Regulatory class: II
21 CFR §884.5300/Product code: 85 MSC

Dear Ms. Myers:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

![img-0.jpeg](img-0.jpeg)

Enclosure

{5}

DEC-22-97 21:20 FROM: LULLES C. KIENS
1423867882
PAGE 3/5

510(k) Number
K970977

Device Name
Glyde Dam Lollyes

Indications for Use
Glyde Dam Lollyes™ are a thin 10" x 6" natural rubber latex sheet specially designed as a barrier for use while performing cunnilingus (oral/vaginal sex) ONLY, and are not intended for use during oral/anal sex.

Glyde Dam Lollyes™, when properly used, may help reduce the risk of catching or spreading many sexually transmitted diseases (“STDs”) such as syphilis, gonorrhea, chlamydia infections, genital herpes and AIDS. However, they cannot eliminate the risk.

During intimate contact, lesions and various bodily fluids can transmit STDs. Therefore, the Glyde Dam Lollyes™ should be applied each and every time before any such contact occurs.

WARNING: DO NOT USE DURING PENETRATING PENILE/VAGINAL OR PENILE/ANAL INTERCOURSE.

WARNING: THIS PRODUCT CONTAINS NATURAL RUBBER LATEX WHICH MAY CAUSE ALLERGIC REACTIONS IN SOME INDIVIDUALS. IF YOU EXPERIENCE A REACTION, STOP USING THIS PRODUCT AND CONTACT YOUR PHYSICIAN.

Concurrence of CDRH, Office of Device Evaluation (ODE)

☐ Prescription Use (per 21 CFR 801.109)
☑ Over-the Counter Use

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K970977

---

**Source:** [https://fda-staging.innolitics.com/device/K970577](https://fda-staging.innolitics.com/device/K970577)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).