← Product Code [IYE](/productcode/IYE) · K970564

# DYNAMIC WEDGE (K970564)

_Mitsubishi Electronics America, Inc. · IYE · May 12, 1997 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K970564

## Device Facts

- **Applicant:** Mitsubishi Electronics America, Inc.
- **Product Code:** [IYE](/productcode/IYE.md)
- **Decision Date:** May 12, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The Dynamic Wedge system is intended to be used with MLC and Mitsubishi linear accelerators with computerized consoles to provide the effective isodose curve and replace the need for traditional hard wedges.

## Device Story

Dynamic Wedge system; add-on for Mitsubishi linear accelerators with multileaf collimator (MLC). Creates effective isodose curves via collimator leaf drive during irradiation; replaces traditional hard wedges. Components: collimator, motor drive unit (MDU), local control equipment (LCE), control console, data management equipment (DME). Upper jaws define field length; lower jaws consist of 31 pairs of leaves (62 total) independently driven by stepping motors for irregular field shaping. Used in radiotherapy clinics by radiation oncologists/technicians. DME transfers treatment data to control leaf movement. Benefits: precise dose distribution, elimination of manual wedge handling.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

System includes collimator, stepping motor drive units, local control equipment, and data management equipment. Collimator features 31 pairs of independently driven leaves. Energy source: linear accelerator. Connectivity: data management equipment for treatment data transfer. Software-controlled leaf positioning.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- EXL series of Mitsubishi Linear Accelerators

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K970564

# 510 (k) SUMMARY

## MITSUBISHI LINEAR ACCELERATORS
### EXL SERIES

MAY 12 1997

1. **Submitter:**
The 510 (k) is submitted by Mr. Hideyuki Kawakami, Sales Manager, Medical Systems Division, Mitsubishi Electronics America, Inc., Two Tower Bridge, 1 Fayette Street, Suite 100, Conshohocken, PA 19428. This 510 (k) summary was prepared on February 6, 1997.

2. **Device Name:**
The 510 (k) submission is for Mitsubishi Dynamic Wedge system, an additional function to Mitsubishi multileaf collimator (MLC) as well as the conventional monoblock collimator.

The predicate device is legally marketed, having been found to be substantially equivalent through the 510 (k) premarket notification process.

3. **Predicate device:**
The devices described in this submission are considered to be substantially equivalent to the EXL series of Mitsubishi Linear Accelerators utilizing klystron and magnetron power supplies and conventional hard-wired control consoles. The predicate device are legally marketed, having been found to be substantially equivalent through the 510 (k) premarket notification process.

4. **Device description:**
This submission is intended to be applicable to the Dynamic Wedge system which can be used with multileaf collimator system on Mitsubishi linear accelerators for external beam radiotherapy. The Dynamic Wedge system is designed to create effective isodose curve by using collimator leaf drive during irradiation. The Dynamic Wedge system consists of the collimator, motor drive unit (MDU), local control equipment (LCE) located in the treatment room, control console, and data management equipment (DME).

The collimator consists of two sets of jaws which determines the field size and shape. The upper jaws consist of a pair of monoblocks just like conventional accelerator therapy collimators. The upper jaws move in tandem and determine the field length.

The lower set of jaws consist of 31 pairs of leaves (62 leaves).

Each pair of leaves is independently driven by a stepping motor. Each pair of leaves can be independently set to provide irregular-shaped field settings.

5. **Intended use:**
The Dynamic Wedge system is intended to be used with MLC and Mitsubishi linear accelerators with computerized consoles to provide the effective isodose curve and replace the need for traditional hard wedges.

{1}

6. Comparison of technological characteristics:

This submission describes the modification of existing Mitsubishi MLC system by adding data management equipment (DME) to transfer treatment data for Dynamic Wedge. Monoblock type collimator can also be used by simply deleting functions to control lower leaves for MLC system.

---

**Source:** [https://fda-staging.innolitics.com/device/K970564](https://fda-staging.innolitics.com/device/K970564)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).