K970545 · Marquette Medical Systems, Inc. · DSI · May 8, 1997 · Cardiovascular
Device Facts
Record ID
K970545
Device Name
DASH 1000 PATIENT MONITOR
Applicant
Marquette Medical Systems, Inc.
Product Code
DSI · Cardiovascular
Decision Date
May 8, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Marquette Eagle 1000 (DASH 1000) Patient Monitor is a patient monitor that is designed to be used to monitor a patient’s basic physiological parameters including: electrocardiography (ECG), invasive blood pressure, non-invasive blood pressure, oxygen saturation, and temperature. Optionally, the printing of information by a paper recorder may be added to the basic monitor configuration. Use of this device is intended for patient populations including: adult, pediatric, and/or neonatal.
Device Story
Microprocessor-based patient monitor; acquires physiological signals including ECG, invasive/non-invasive blood pressure, SpO2, and temperature. Processes signals to display real-time data and trigger audible/visual alarms for parameter violations. Includes optional paper recorder. Used in hospital environments (OR, ICU, recovery) and outpatient clinics by trained clinicians (physicians, nurses, anesthetists). Output allows clinicians to monitor patient status and make clinical decisions based on displayed vitals and alarm notifications. Benefits include continuous patient surveillance and timely notification of physiological changes.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included precision, accuracy, and safety verification compared to predicate devices to ensure performance meets specified requirements.
Technological Characteristics
Microprocessor-based, software-driven patient monitor. Parameters: ECG, invasive/non-invasive blood pressure, SpO2, temperature. Connectivity: optional paper recorder. Power: battery-operated. Software: reused from former devices. Sterilization: not applicable (non-invasive monitoring).
Indications for Use
Indicated for monitoring basic physiological parameters (ECG, invasive/non-invasive blood pressure, SpO2, temperature) in adult, pediatric, and neonatal patient populations within hospital settings (OR, post-anesthesia recovery, critical/intensive care) or outpatient clinics/surgical centers. Not for home use, patient transport outside the hospital, or without physician order.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Marquette Eagle 3000 Patient Monitor
Submission Summary (Full Text)
{0}
K970545
# 510(k) Summary of Safety and Effectiveness
MAY 8 1997
Date: February 7, 1997
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
## Submitter:
Marquette Medical Systems
8200 West Tower Avenue
Milwaukee, WI 53223 U.S.A.
## Contact Person:
Dianne Schmitz
Corporate Regulatory Affairs
Marquette Medical Systems
Phone: (414) 362-3230
Fax: (414) 355-3790
## General Information:
### Common/Usual Name
This device is commonly known as a patient monitoring system.
### Trade/Proprietary Name
Marquette's trade/ proprietary name for this device domestically within the United States is the DASH 1000 Patient Monitor. Marquette Hellige's trade/ proprietary name internationally is the Eagle 1000 Patient Monitor.
### Classification Name(s)
The Marquette Eagle 1000 Patient Monitor's classification names, classification panels, and regulation citations include:
* 21 CFR 870.1110 Monitor, Blood Pressure, Indwelling 74CAA
* 21 CFR 870.1130 Monitor, Blood Pressure, Non-Indwelling 74BXD
* 21 CFR 880.2910 Monitor, Temperature (with probe) 80BWX
* 21 CFR 870.2300 Monitor, Cardiac (incl. cardiotachometer & rate alarm) 74DRT
* 21 CFR 870.2700 Oximeter, Pulse 74BWS
* 21 CFR 870.1025 Detector and Alarm, Arrhythmia 74DSI
### Device Classification
Devices monitoring similar parameters have been determined to be Class II devices according to the Cardiovascular Device Classification Panel. Therefore, it is believed that the system will be considered a Class II device.
However, due to ongoing discussions between the Agency and Industry on the definition of 21 CFR 870.1025, Arrhythmia detector and alarm, which is a Class III parameter, Marquette Medical Systems has included the additional information that would be required for a Class III device within the premarket notification submission. This is being provided should the Agency determine the Eagle 1000 Patient Monitor is a Class III device.
Klaus Rudolf /des Jan. 97
00190
{1}
Performance Standards
Performance standards (Section 514 of the Act) have not yet been established for the device that is the subject of this premarket notification submission.
## Predicate Device:
Marquette Eagle 3000 Patient Monitor
## Device Description:
The Marquette Eagle 1000 Patient Monitor is a patient monitoring system that is designed to be used to monitor a patient's basic physiological parameters including: electrocardiography (ECG), invasive blood pressure, non-invasive blood pressure, oxygen saturation, and temperature.
## Intended Use:
The Marquette Eagle 1000 Patient Monitor is designed to monitor and display patient data. Its design allows the operator to adjust parameter alarm settings that would audibly and visually notify the operator when a violation occurs. The option is provided for printing of information by a paper recorder.
Use of the Marquette Eagle 1000 Patient Monitor is intended for patient populations including: adult, pediatric, and/or neonatal.
Use of the Marquette Eagle 1000 Patient Monitor is not recommended for use in patient's home or residence, during patient transport outside the hospital, or when it has not been ordered by a physician or other qualified medical personnel.
Use of the Marquette Eagle 1000 Patient Monitor is intended for operating room (OR), post anesthesia recovery, critical care and intensive care. These departments are typically located in hospitals or may be located in outpatient clinics or free standing surgical centers. It is intended for use by physicians, physician assistants, registered nurses, certified registered nurse anesthetists, or other hospital personnel trained in the use of the equipment.
## Conclusions:
The Eagle 1000 Patient Monitor is a microprocessor-based, software-driven device. The signal-acquisition and -processing technologies and the basic parts of the device software were re-used from former devices.
Testing was performed on the Eagle 1000 Patient Monitor and its predicate devices. Precision, accuracy, as well as safety testing was performed. Test results indicate that the Eagle 1000 Patient Monitor provides an equivalent level or better in performance, when compared to the legally marketed predicate device(s) when tested to the accuracy requirements as specified in the contents of the premarket notification submission.
The Eagle 1000 Patient Monitor passed the EC type-examination, and thus bears the CE mark.
Marquette Medical Systems has demonstrated that the Eagle 1000 (DASH 1000) Patient Monitor is as safe and effective, and performs substantially equivalent to the predicate device.
Klaus Rudolf /des Jan. 97
00191
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Ms. Dianne Schmitz
Marquette Medical Systems
8200 West Tower Avenue
Milwaukee, Wisconsin 53223
MAY - 8 1997
Re: K970545
Eagle 1000 (DASH 1000) Series Patient Monitor with Model numbers:
Eagle Model 1001 (Software Version 1.0); Eagle Model 1002 (Software Version 1.0); Eagle Model 1003 (Software Version 1.0); Eagle Model 1004 (Software Version 1.0); Eagle Model 1005 (Software Version 1.0); Eagle Model 1006 (Software Version 1.0); Eagle Model 1007 (Software Version 1.0); Eagle Model 1008 (Software Version 1.0); Eagle Model 1009 (Software Version 1.0); Eagle Model 1010 (Software Version 2.0); Eagle Model 1011 (Software Version 2.0); Eagle Model 1014 (Software Version 2.0); and Eagle Model 1015 (Software Version 2.0)
Regulatory Class: III (three)
Product Code: 74 DSI
Dated: February 11, 1997
Received: February 12, 1997
Dear Ms. Schmitz:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP 2 -
{3}
Page 2 - Ms. Dianne Schmitz
inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the *Federal Register*. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}
510(k) Number (if known): Unknown; 510(k) filed February 11, 1997
Device Name: Eagle 1000 (DASH 1000) Patient Monitor
Indications For Use: MAY - 8 1997
The Marquette Eagle 1000 (DASH 1000) Patient Monitor is a patient monitor that is designed to be used to monitor a patient’s basic physiological parameters including: electrocardiography (ECG), invasive blood pressure, non-invasive blood pressure, oxygen saturation, and temperature.
Optionally, the printing of information by a paper recorder may be added to the basic monitor configuration.
Use of this device is intended for patient populations including: adult, pediatric, and/or neonatal.
| EAGLE 1002
(101 123 02) | Battery | ECG | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| EAGLE 1003
(101 124 02) | Battery | ECG | Recorder | | | |
| EAGLE 1004
(101 123 03) | Battery | ECG | SpO_{2} | | | |
| EAGLE 1005
(101 124 03) | Battery | ECG | SpO_{2} | Recorder | | |
| EAGLE 1006
(101 123 04) | Battery | ECG | | NBP | | |
| EAGLE 1007
(101 124 04) | Battery | ECG | | NBP | Recorder | |
| EAGLE 1008
(101 123 05) | Battery | ECG | SpO_{2} | NBP | | |
| EAGLE 1009
(101 124 05) | Battery | ECG | SpO_{2} | NBP | Recorder | |
| EAGLE 1010
(101 123 06) | Battery | ECG | SpO_{2} | NBP | TEMP | |
| EAGLE 1011
(101 124 06) | Battery | ECG | SpO_{2} | NBP | TEMP | Recorder |
| EAGLE 1014
(101 123 07) | Battery | ECG | SpO_{2} | NBP | TEMP | 2 x IBP |
| EAGLE 1015
(101 124 07) | Battery | ECG | SpO_{2} | NBP | TEMP | 2 x IBP |
EAGLE-Varianten NBP = Non-invasive Blood Pressure IBP = Invasive Blood Pressure
Software Version 1.0 for Eagle Models 1002, 1003, 1004, 1005, 1006, 1007, 1008, +1009
Software Version 2.0 for Eagle Models 1010, 1014, 1011, +1015.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K970545
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.