← Product Code [GEI](/productcode/GEI) · K970541

# AESCULAP MODULAR MONOPOLAR ELECTRODES (K970541)

_Aesculap, Inc. · GEI · Mar 28, 1997 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K970541

## Device Facts

- **Applicant:** Aesculap, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Mar 28, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Modular Monopolar Electrodes presented in this submission are intended for use in endoscopic surgery (laparoscopy, thoracoscopy) for coagulation and dissection of selected tissue.

## Device Story

Modular monopolar electrodes for endoscopic surgery (laparoscopy, thoracoscopy); used for tissue coagulation and dissection. Device consists of three reusable, modular components: insulated handle, insulated shaft, and interchangeable working tips (e.g., J-hook). Operated by surgeons in clinical settings. Device connects to electrosurgical generators (not included) to deliver energy to tissue via the tip. Modular design allows disassembly for cleaning and steam sterilization. Benefits include surgical versatility through interchangeable tips and improved maintenance via modularity.

## Clinical Evidence

Bench testing only; no clinical data provided. Compliance with IEC 601-2-18 (safety of endoscopic equipment) and manufacturing per ISO/German DIN standards.

## Technological Characteristics

Modular design; 5 mm diameter, 450 mm length. Components: insulated handle (GK382R), insulated shaft (GK380R), and interchangeable working tips. Materials: unspecified insulation materials. Energy: monopolar electrosurgical. Connectivity: none. Sterilization: steam. Standards: IEC 601-2-18, ISO 9001, German DIN.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

AESCULAP

MAR 28 1997

K 970541

510(k) Summary of Safety and Effectiveness in Accordance with SMDA of 1990

Aesculap Modular Monopolar Electrodes

February 11, 1997

Submitted by: Aesculap®, Inc.
1000 Gateway Blvd.
So. San Francisco, CA 94080
Contact: Mary Ellen Holden
Phone: (415) 876-7000 x348
FAX: (415) 589-3007

Product: Aesculap Modular Monopolar Electrodes
Common Name: Monopolar Electrodes

## Intended Use

The Modular Monopolar Electrodes presented in this submission are intended for use in endoscopic surgery (laparoscopy, thoracoscopy) for coagulation and dissection of selected tissue.

## Technological Characteristics

Aesculap currently markets a line of non-modular endoscopic monopolar electrodes and a line of modular endoscopic monopolar scissors and forceps which are subject to 510(k) clearance. This submission represents the following design changes:

- Change electrodes from a non-modular design to a modular design.
- Change in tip insulation material.
- Change in shaft insulation material.

The Modular Monopolar Electrodes are made up of 3 separate components: an insulated handle (GK382R), an insulated shaft (GK380R) and a working tip. The working tips come in several different styles (i.e. J-shaped hook electrode GK383R) which can be used interchangeably. All components are reusable and disassemble for cleaning and sterilization.

The assembled electrodes are 5 mm in diameter and 450 mm in length. They have no suction/irrigation feature.

1000 GATEWAY BOULEVARD • SOUTH SAN FRANCISCO, CA 94080-7030 • USA • 415-876-7000 • FAX: 415-876-7028

{1}

005

## Performance Standards

No applicable performance standards have been promulgated under Section 514 of the Food, Drug and Cosmetic Act for these devices. However, Aesculap’s Modular Monopolar Electrodes comply with the requirements of IEC 601-2-18 (Medical electrical equipment, Part 2: Particular requirements for the safety of endoscopic equipment.)

The Modular Monopolar Electrodes will be manufactured in accordance with ISO and German Din Standards. Furthermore, Aesculap AG has received ISO 9001 certification.

## Sterilization

The Modular Monopolar Electrodes are provided non-sterile and must be sterilized prior to use. The devices may be sterilized by steam sterilization.

## Substantial Equivalence

Aesculap’s Modular Monopolar Electrodes shares similar features and function with corresponding devices distributed by Aesculap, Brimfield and Marlow.

---

**Source:** [https://fda-staging.innolitics.com/device/K970541](https://fda-staging.innolitics.com/device/K970541)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com