K970512 · Becton Dickinson Microbiology Systems · MDB · Jan 8, 1998 · Microbiology
Device Facts
Record ID
K970512
Device Name
BACTEC MYCO/F LYTIC CULTURE VIALS
Applicant
Becton Dickinson Microbiology Systems
Product Code
MDB · Microbiology
Decision Date
Jan 8, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2560
Device Class
Class 1
Indications for Use
BACTEC MYCO/F LYTIC when used with the BACTEC 9000MB instrument is a non-selective culture medium for the qualitative culture and recovery of mycobacteria from blood specimens.
Device Story
BACTEC MYCO/F LYTIC is a non-selective growth medium vial for mycobacteria recovery from 1-5 mL blood specimens. Used with BACTEC 9000MB instrument in clinical laboratories. Vial contains Middlebrook 7H9 and Brain Heart Infusion broth with fluorescence sensor. Principle of operation: sensor monitors oxygen concentration changes resulting from microbial metabolism; decreasing oxygen increases fluorescence. Instrument continuously monitors vials for fluorescence; positive result indicates presumptive presence of viable mycobacteria. Healthcare providers use output to identify mycobacterial infections in patients, including immunocompromised and transplant populations. Benefits include automated, non-invasive detection of mycobacteria in blood.
Clinical Evidence
Clinical study at two tertiary care hospitals compared BACTEC MYCO/F LYTIC to BACTEC 13A using 284 blood specimens. 39 pathogenic mycobacterial isolates recovered. 5 (13%) recovered in subject medium only; 2 (5%) in predicate only. False positive rate 0.4% (1/284); false negative rate 0% based on terminal subcultures. Contamination rate 0.9%. Internal bench testing evaluated recovery of 14 mycobacteria species at various CFU levels.
Technological Characteristics
Growth medium: Modified Middlebrook 7H9 and Brain Heart Infusion broth. Contains fluorescence sensor (O2-based). Dimensions: 1-5 mL blood capacity. Connectivity: Used with BACTEC 9000MB instrument. Monitoring: Non-invasive, fluorescent detection. Incubation: 37°C ± 1.5°C with internal agitation every 10 minutes.
Indications for Use
Indicated for the qualitative culture and recovery of mycobacteria in human blood specimens from patients suspected of having septicemia.
Regulatory Classification
Identification
A microbial growth monitor is a device intended for medical purposes that measures the concentration of bacteria suspended in a liquid medium by measuring changes in light scattering properties, optical density, electrical impedance, or by making direct bacterial counts. The device aids in the diagnosis of disease caused by pathogenic microorganisms.
Predicate Devices
BACTEC 13A MYCOBACTERIA CULTURE MEDIUM
Reference Devices
BACTEC MYCO/F Sputa culture medium
Submission Summary (Full Text)
{0}
1410512
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
SUBMITTED BY: BECTON DICKINSON MICROBIOLOGY SYSTEMS
7 LOVETON CIRCLE
SPARKS, MD 21152
JAN - 8 1998
CONTACT: Dennis R. Mertz, Manager of Regulatory Affairs
TELEPHONE: (410) 316-4099
FAX: (410) 316-4499
PREPARED: December 23, 1997
DEVICE NAME: BACTEC MYCO/F LYTIC BLOOD CULTURE MEDIUM
DEVICE CLASSIFICATION: Monitor, Microbial Growth, Class I
PREDICATE DEVICE: BACTEC 13A MYCOBACTERIA CULTURE MEDIUM
INTENDED USE: BACTEC MYCO/F LYTIC when used with the BACTEC 9000MB instrument is a non-selective culture medium for the qualitative culture and recovery of mycobacteria from blood specimens.
DEVICE DESCRIPTION:
BACTEC MYCO/F LYTIC blood culture medium is a non-selective growth medium intended for the culture and recovery of mycobacteria and designed for blood volumes of one to five mL. BACTEC MYCO/F LYTIC culture medium is a Middlebrook 7H9 and Brain Heart Infusion broth formulation with specific formulation modifications made to enhance the growth of mycobacteria. It is used specifically with the BACTEC 9000MB instrument in the monitoring of clinical blood specimens for the presence of microorganisms. This medium contains the same fluorescence senor as the BACTEC MYCO/F Sputa culture medium and detection is based on changes in oxygen concentration in the vial resulting from metabolism and growth of microorganisms. The sensor is monitored by the BACTEC 9000MB System for increasing fluorescence which is proportional to the decrease in oxygen. A positive determination indicates the presumptive presence of viable microorganisms in the vial.
SUBSTANTIAL EQUIVALENCE:
Table 1 summarizes the similarities and differences between the BACTEC MYCO/F LYTIC Culture medium and the BACTEC 13A Mycobacteria culture medium.
{1}
# INTERNAL PERFORMANCE:
A study was conducted to evaluate the recovery and time to detection of a variety of mycobacteria species at different CFU levels with the BACTEC MYCO/F LYTIC Culture medium. The following isolates were detected as positive in the BACTEC 9000MB instrument using BACTEC MYCO/F Lytic medium: *M. tuberculosis*, *M. kansasii*, *M. fortuitum*, *M. avium*, *M. intracellulare*, *M. bovis*, *M. terrae*, *M. simiae*, *M. gordonae*, *M. celatum*, *M. abscessus*, *M. malmoense*. During internal studies, *M. xenopi* and *M. szulgai* exhibited unsatisfactory recovery with BACTEC MYCO/F LYTIC culture medium. TABLE 2 describes the results of this study.
# CLINICAL PERFORMANCE:
The BACTEC MYCO/F Lytic culture medium was evaluated with the BACTEC 9000MB instrument at two clinical sites considered large tertiary care teaching hospitals in geographically diverse areas. The site populations included patients suspected of a mycobacterial infection, immunocompromised patients and transplant patients. The BACTEC MYCO/F Lytic culture medium was compared to the BACTEC 13A culture medium for the recovery and detection of mycobacteria from blood specimens. A total of 284 blood specimens were tested during the study. The total number of pathogenic mycobacterial isolates recovered in the study was 39 (See TABLE 1). Of these positives, five (13%) were recovered in the BACTEC MYCO/F Lytic culture medium only and two (5%) were recovered by BACTEC 13A culture medium only. A total of 28 BACTEC MYCO/F LYTIC vials were over filled with specimen (between 6 to 20 mL) during the clinical evaluation and were not included in this study since they were above the maximum fill volume. Of these 28 BACTEC MYCO/F LYTIC vials, 16 (57%) were identified as false positive.
Of the 284 blood specimens tested in the clinical study, one BACTEC MYCO/F LYTIC vial (0.4%) was determined to be false positive (instrument-positive, smear and/or subculture-negative). Of the 38 instrument positive MYCO/F LYTIC vials, 1 (2.6%) was determined to be false positive. The false negative rate (instrument-negative, smear and/or subculture-positive) was determined to be 0% based on terminal subcultures of 50% of negative vials. The contamination rate during this evaluation was determination to be 0.9%.
TABLE 1: SUMMARY OF MYCO/F LYTIC CULTURE MEDIUM ISOLATE RECOVERY DURING CLINICAL TRIALS
| Organism | Total Isolates | Myco/F Lytic Medium Only | 13A Medium Only | Both |
| --- | --- | --- | --- | --- |
| All Pathogenic Mycobacteria: | | | | |
| Mycobacterium avium | 30 | 3 | 1 | 26 |
| Mycobacterium tuberculosis | 6 | 0 | 0 | 6 |
| Mycobacterium kansasii | 3 | 2 | 1 | 0 |
| Total | 39 | 5 | 2 | 32 |
{2}
Table 1. Substantial Equivalence of BACTEC MYCO/F LYTIC Culture Medium to BACTEC 13A
| | BACTEC MYCO/F LYTIC | BACTEC 13A |
| --- | --- | --- |
| Intended Use | Qualitative culture and recovery of mycobacteria | Qualitative culture and recovery of mycobacteria |
| Sample Type | Blood, unprocessed and other sterile body fluids | Blood, unprocessed |
| Sample Volume | 1 - 5 mL | 1 - 5 mL |
| Blood to Broth Ratio | 1 to 8 | 1 to 6 |
| Growth Medium | Modified Middlebrook 7H9 and enriched brain heart infusion broth | Modified Middlebrook 7H9 broth |
| Reactive ingredients: • Process water • Brain heart infusion • Soybean-Casein Digest • 7H9 Broth Base • Inositol • Casein Hydrolysate • Ferric Ammonium Citrate • Glycerol • Sodium Polysulfonate(SPS) • Tween 80(Polysorbate) • Saponin • L-Asparagine¹ • Catalase • Antifoam Agent • Hemin • ¹⁴C Substrate • Potassium Phosphate • Pyridoxal HCL • Ammonium Sulfate | 40mL 0.5%w/v 0.10%w/v 0.12%w/v 0.05%w/v 0.10%w/v 0.006%w/v 0.10%w/v 0.025%w/v 0.0025%w/v 0.24%w/v 0.10%w/v --- 0.01%w/v --- --- 0.024%w/v 0.0001%w/v --- | 30mL --- --- 0.47%w/v --- 0.10%w/v --- 0.025%w/v 0.02%w/v --- --- 1440 units --- --- 5μCi --- --- --- |
| Supplement | None | BACTEC Enrichment |
| Instrument | BACTEC 9000MB | BACTEC 460TB |
| Growth Detection | O₂ metabolism | Palmitate Decarboxylation |
| Incubation T°/mixing | 37°C ± 1.5°C; internal instrument agitation every 10 minutes | 37°C ± 1.5°C; no agitation by instrument |
| Type of Monitoring | Non-invasive, fluorescent detection | Invasive vial headspace sampling |
| Antimicrobial Supplement | None | None |
(Noted as Supplement H)
{3}
Table 2
Detection of Mycobacteria in the Myco/F Lytic Medium.
BACTEC 9000MB
| | strain | cfu/bottle | 1 mL blood | 3 mL blood | 5 mL blood |
| --- | --- | --- | --- | --- | --- |
| *M. tuberculosis* | 582 | 0, 0 | 20.7 | 18.7 | 17.3 |
| Replicate | | | 25.0 | neg | neg |
| Average | | | 22.9 | 18.7 | 17.3 |
| *M. avium* | 2638 | 49, 45 | 7.8 | 8.1 | 8.1 |
| Replicate | | | 8.1 | 8.1 | 8.1 |
| Average | | | 8.0 | 8.1 | 8.1 |
| *M. intracellulare* | 2792 | 80, 44 | 25.8 | 16.7 | 10.5 |
| Replicate | | | 24.3 | 22.5 | 15.0 |
| Average | | | 25.0 | 19.6 | 12.8 |
| *M. fortuitum* | 3072 | 5, 0 | 9.1 | 5.6 | 5.0 |
| Replicate | | | 6.8 | 6.0 | 5.2 |
| Average | | | 8.0 | 5.8 | 5.1 |
| *M. bovis* | 2003 | 12, 13 | 24.4 | 20.0 | 20.0 |
| Replicate | | | 24.7 | 21.0 | 20.4 |
| Average | | | 24.5 | 20.5 | 20.2 |
| *M. kansasii* | 2205 | 7,3 | 12.3 | 14.3 | 14.3 |
| Replicate | | | 13.0 | 13.3 | neg |
| Average | | | 12.7 | 13.8 | 14.3 |
| *M. terrae* | 3001 | 0, 0 | 15.3 | 11.6 | 16.6 |
| Replicate | | | 17.0 | 32.4 | neg |
| Average | | | 16.2 | 22.0 | 16.6 |
| *M. szulgai* | 2353 | 1,2 | 15.7 | neg | neg |
| Replicate | | | neg | 22.7 | neg |
| Average | | | 15.7 | 22.7 | neg |
| *M. simiae* | 2304 | 68, 58 | 7.2 | 7.4 | 7.7 |
| Replicate | | | 7.2 | 7.5 | 7.4 |
| Average | | | 7.2 | 7.5 | 7.6 |
| *M. gordonae* | 2454 | 2, 5 | 26.4 | 28.7 | neg |
| Replicate | | | 24.0 | 32.0 | neg |
| Average | | | 25.2 | 30.4 | neg |
| *M. celatum* | 3661 | 53, 31 | 18.3 | 12.0 | 12.3 |
| Replicate | | | 18.3 | 15.0 | 12.3 |
| Average | | | 18.3 | 13.5 | 12.3 |
| *M. abscessus* | 3370 | 1, 0 | 4.9 | 4.4 | 3.9 |
| Replicate | | | 9.3 | 4.3 | neg |
| Average | | | 7.1 | 4.4 | 3.9 |
| *M. malmoense* | 3472 | 16, 20 | 30.0 | 21.5 | 10.8 |
| Replicate | | | 32.0 | 21.1 | 18.6 |
| Average | | | 31.0 | 21.3 | 14.7 |
| *M. haemophilum* | 5121 | 1, 1 | 36.1 | 23.0 | 18.4 |
| Replicate | | | 39.3 | 23.4 | 24.4 |
| Average | | | 37.7 | 23.2 | 21.4 |
| *M. xenopi* | 3052 | 0, 0 | neg | neg | neg |
| Replicate | | | neg | neg | neg |
| Average | | | neg | neg | neg |
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
JAN - 8 1998
Mr. Dennis R. Mertz
Manager, Regulatory Affairs
Becton Dickinson Microbiology Systems
7 Loveton Circle
Sparks, Maryland 21152-0999
Re: K970512
Trade Name: BACTEC Myco/F Lytic Culture Vials
Regulatory Class: I
Product Code: MDB
Dated: October 20, 1997
Received: October 21, 1997
Dear Mr. Mertz:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{5}
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{6}
Page ___ of ___
510(k) Number (if known): K970512
Device Name: Myco/F Lytic Culture Vials
Indications For Use:
The BACTEC ® MYCO/F LYTIC Culture vial when used with the BACTEC 9000MB System is a qualitative test for the culture and recovery of mycobacteria in human blood specimens from patients who are suspected of having septicemia.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
John Ticohurst MD 1/6/98
(Division Sign-Off) Interim Chief, Microbiology
Division of Clinical Laboratory Devices
510(k) Number K970512
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.