INNOFLOUR TOPIRAMATE REAGENT SET
Device Facts
| Record ID | K970510 |
|---|---|
| Device Name | INNOFLOUR TOPIRAMATE REAGENT SET |
| Applicant | Oxis Intl., Inc. |
| Product Code | MSL · Clinical Toxicology |
| Decision Date | May 12, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3350 |
| Device Class | Class 2 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K970510 · May 12, 1997 | INNOFLOUR TOPIRAMATE REAGENT SET | Oxis Intl., Inc. | Patient clinical samples | The device performance was validated by comparing results from the INNOFLUOR™ Topiramate Assay System against Topiramate Gas Chromatography (GC) using 117 patient samples from individuals undergoing topiramate therapy. | Method comparison; Patient samples; Clinical validation |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Method Comparison Study; Retrospective method comparison | Patients receiving topiramate therapy; Sample Size: 117 | Topiramate Gas Chromatography (GC) | Correlation of topiramate concentration measurements |
Indications for Use
The INNOFLUOR™ Topiramate Reagent Set, the INNOFLUOR™ Topiramate Calibrator Set and the INNOFLUOR™ Topiramate Control Set are packaged and sold separately, but are referred to collectively, in all associated product labeling, as the INNOFLUOR™ Topiramate Assay System. The INNOFLUOR™ Topiramate Assay System is an in vitro diagnostic device intended for the quantitative determination of total topiramate in serum or heparinized plasma by fluorescence polarization immunoassay. Topiramate serum or heparinized plasma concentrations are measured to aid in achieving appropriate therapy. The assay system is for use on the Abbott TDx® or the TDxFLx® (TDx®/TDxFLx®) analyzer. The INNOFLUOR™ Topiramate Reagent Set is intended for the quantitative determination of total topiramate in serum or heparinized plasma. The reagent set is intended for use in the INNOFLUOR™ Topiramate Assay System. The INNOFLUOR™ Topiramate Calibrator Set is intended for use in the calibration of the INNOFLUOR™ Topiramate Assay System. The INNOFLUOR™ Topiramate Reagent Set and Calibrator Set are used together to generate the calibration curve on the TDx®/TDxFLx® analyzer. The calibration curve must be established prior to assaying unknown samples. Prior to performing the calibration procedure, the correct analyzer operating parameters must be set by following the instructions provided in the Product Insert Supplement, which is included with every INNOFLUOR™ Topiramate Reagent Set. The INNOFLUOR™ Topiramate Control Set is intended for use in the quality control of the INNOFLUOR™ Topiramate Assay System. The concentrations of the three levels of the controls have been selected to best monitor the quality and stability of the calibration curve. The control results must be within established acceptable ranges before patient samples are assayed.
Device Story
In vitro diagnostic assay system; utilizes fluorescence polarization immunoassay (FPIA) technology; measures total topiramate concentrations in serum or heparinized plasma. System components include reagent set, calibrator set, and control set. Operated on Abbott TDx® or TDxFLx® analyzers. Calibration curve generated using calibrator set; quality control performed using control set. Healthcare providers use quantitative results to monitor topiramate levels, aiding in achieving appropriate therapeutic dosing for patients. Intended for clinical laboratory use.
Clinical Evidence
Bench testing only. Comparison of 117 patient samples using INNOFLUOR™ Topiramate Assay System versus Gas Chromatography (GC). Linear regression analysis yielded equation: (INNOFLUOR™) = 0.985 x (GC) - 0.147, with a correlation coefficient of 0.9934.
Technological Characteristics
Fluorescence polarization immunoassay (FPIA) technology. Designed for use on Abbott TDx®/TDxFLx® analyzers. System includes reagents, calibrators, and controls for quantitative analysis of serum or heparinized plasma.
Indications for Use
Indicated for quantitative determination of total topiramate in serum or heparinized plasma to aid in therapeutic monitoring of patients receiving topiramate therapy.
Regulatory Classification
Identification
A diphenylhydantoin test system is a device intended to measure diphenylhydantoin, an antiepileptic drug, in human specimens. Measurements obtained by this device are used in the diagnosis and treatment of diphenylhydantoin overdose and in monitoring levels of diphenylhydantoin to ensure appropriate therapy.
Predicate Devices
- INNOFLUOR™ Phenobarbital Assay System (INNOFLUOR™ Phenobarbital Reagent Set and INNOFLUOR™ Phenobarbital Calibrator Set)
- Abbott Phenobarbital II Assay
Reference Devices
- Topiramate Gas Chromatography