FOUNDATION INDIRECT COMPOSITE RESIN

K970495 · Sybron Dental Specialties, Inc. · EBF · Apr 16, 1997 · Dental

Device Facts

Record IDK970495
Device NameFOUNDATION INDIRECT COMPOSITE RESIN
ApplicantSybron Dental Specialties, Inc.
Product CodeEBF · Dental
Decision DateApr 16, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3690
Device ClassClass 2

Indications for Use

Tooth Shade Resin Material, Class II, per 21 CFR 872.3690

Device Story

Foundation Indirect Composite Resin is a dual-cured restorative material for dental laboratory fabrication of crowns and bridges. It reformulates existing Opaceous Dentin Pastes by adding VAZO 88 (heat-activated initiator) and replacing Barium Alumnoborosilicate filler with Magnesium Aluminosilicate. The material undergoes light activation followed by a heat and pressure curing cycle. The device is used by dental laboratory technicians to create durable, esthetically acceptable restorations. The resulting material exhibits reduced thermal expansion and lower polymerization shrinkage compared to previous versions, enhancing durability for long-term oral residence.

Clinical Evidence

Bench testing only. Biocompatibility testing conducted per ISO 10993 standards, including Ames Mutagenicity Assay, Cytotoxicity Study (USP 23), Kligman Maximization Study, and Intramuscular Implantation Test. Efficacy demonstrated via side-by-side bench testing comparisons against predicate devices.

Technological Characteristics

Dual-cured composite resin; contains VAZO 88 (Azobiscyclohexane Carbonitrile) heat-activated initiator; filler is 56 weight percent Magnesium Aluminosilicate glass; exhibits reduced thermal expansion and 33-70% reduction in polymerization shrinkage compared to previous formulations.

Indications for Use

Indicated for use as a dual-cured indirect composite restorative material for the fabrication of crowns and bridges in a dental laboratory setting.

Regulatory Classification

Identification

Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} K9 70495 APR 16 1997 Page 1 of 2 # 510 ( k ) Summary ## Statement of Safety and Effectiveness ### Kerr Foundation Indirect Composite Resin #### Submitter Sybron Dental Specialties Inc. 17 17 West Collins Avenue Orange CA 92867 (714) 516-7486 - Phone (714) 516-7488 - FAX William R. Pike - Contact Person ## Device Name Trade Name: Foundation Indirect Composite Resin Common Name: Dual cured indirect composite restorative material Classification Name: Tooth Shade Resin Material, Class II, per 21 CFR 872.3690 ## Devices for which Substantial Equivalence is Claimed belleGlass HP ( Kerr Corp. ), Concept ( Williams Dental ), Art Glass ( Hereas Kulzer GmbH ) ## BACKGROUND Kerr introduced a new composite resin based Crown and Bridge fabrication system called belleGlass HP in May, 1996 ( Reference K955331 ). This system is comprised of all of the components necessary for a dental laboratory to fabricate composite resin-based Crowns and Bridges and cure them using both light activation combined with a final heat and pressure curing cycle in the belleGlass HP automatic curing device. ( Complete existing Instructions for Use of this product is supplied elsewhere in this submission. ) One of the sets of composite materials used in this process is referred to as Opaceous Dentin Pastes. These consist of 16 light cured only pastes which match the Vita Shade Guide used with Vita Porcelain powders. These materials proved adequate, but because of their single cure mode ( visible light activation ) they may not be durable enough to withstand the constant abuse associated with long term residence in the oral environment. Therefore, a more durable, yet esthetically acceptable material was sought. ## FOUNDATION INDIRECT COMPOSITE Kerr’s Foundation consists of a reformulation of the existing Opaceous Dentin Pastes in order to increase durability and material stability in the mouth. This reformulation consists essentially of two significant modifications: 1. The addition of 0.15 weight percent of VAZO 88 ( Azobiscyclohexane Carbonitrile ), a heat activated polymerization initiator. Page 5 {1} Page 2 of 2 2. The replacement by 56 weight percent of the Barium Alumnoborosilicate glass filler with a corresponding amount of Magnesium Aluminosilicate, a low coefficient of thermal expansion glass filler. This combination of initiator and glass filler modifications results in a composite resin-based restorative material that exhibits half the thermal expansion characteristics of existing products and produces reductions in Volume % Polymerization Shrinkage in a range of 33 to 70 %. ## SAFETY The safety of Kerr Foundation Indirect Restorative has been demonstrated by subjecting cured samples of the material to various types of biocompatibility tests as recommended in the ISO 10993 biocompatibility guidance standard. These tests were conducted by an independent laboratory which specializes in safety and toxicity evaluation. The tests include: 1. Ames Mutagenicity Assay 2. Cytotoxicity Study (Elution Test, USP 23) 3. Kligman Maximization Study (Tissue Sensitization) 4. Intramuscular Implantation Test (Tissue Irritation) ## EFFICACY Effectiveness or suitability to the intended purpose of Kerr Foundation has been demonstrated by a combination of in-house testing and side by side test comparisons to predicate devices currently on the market. Results of this bench testing indicates that Kerr Foundation Indirect Composite performs as well or better than three predicate devices currently on the market Results of this testing is included elsewhere in this submission. Page 6
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...