ACCU-PREP is intended to be used to abrade and enhance the bond between the prepared tooth and cast dental metals for Maryland bridges, for cementation of cast metal restorations, for porcelain repair, as a porcelain etch substitute, to prepare indirect composite restorations for cementation, for repair of composite or resin restorations, repair of acrylic facings on cast restorations, endodontic post preparation, denture repairs, and orthodontic applications such as: preparation of brackets to enhance bonding, cement removal from brackets, and enhancement of adhesion to tooth structure.
Device Story
ACCU-PREP is a handheld, compressed-air-driven dental sandblasting instrument. It utilizes a venturi principle to draw abrasive powder (typically 50μm aluminum oxide) from an internal reservoir and propel it at high velocity toward dental hard tissues or restorative materials. The device is operated by dental professionals in a clinical setting. By roughening surfaces, it enhances micromechanical bonding for various dental procedures, including bridge cementation, porcelain/composite repairs, and orthodontic bracket preparation. The device functions similarly to a perfume atomizer, where air flow creates a vacuum to entrain abrasive particles. Compared to the predicate, the device offers improved aesthetics, a more ergonomic reservoir placement (Type II), and operates at a lower minimum air pressure (40 psig). The output is a stream of abrasive particles used to clean or abrade surfaces, facilitating stronger adhesion of dental materials and improving clinical outcomes for restorative and orthodontic treatments.
Clinical Evidence
Bench testing only. Performance compared to predicate using shear bond strength tests on human extracted teeth and Scanning Electron Microscopy (SEM) of abraded surfaces. Results showed no statistically significant difference in bond strength (p ≥ 0.05) and virtually identical surface abrasion patterns.
Technological Characteristics
Handheld dental sandblasting instrument; compressed air energy source (40-100 psig); venturi-based abrasive delivery; uses 50μm aluminum oxide (range 0.05-90μm); Type II model is fully autoclavable; mechanical operation; no software or electronic components.
Indications for Use
Indicated for dental professionals performing restorative, endodontic, and orthodontic procedures requiring surface abrasion to enhance bonding. Applications include Maryland bridges, cast metal restoration cementation, porcelain repair, porcelain etching, composite/polymeric restoration repair, acrylic facing repair, endodontic post preparation, denture repair, and orthodontic bracket bonding/cleaning.
Regulatory Classification
Identification
An airbrush is an AC-powered device intended for use in conjunction with articulation paper. The device uses air-driven particles to roughen the surfaces of dental restorations. Uneven areas of the restorations are then identified by use of articulation paper.
Special Controls
The special control for this device is International Electrotechnical Commission's IEC 60601-1-AM2 (1995-03), Amendment 2, “Medical Electrical Equipment—Part 1: General Requirements for Safety.”
Predicate Devices
MICROETCHER™
Submission Summary (Full Text)
{0}
BISCO, INC., 1500 W. Thorndale Ave., Itasca, IL 60143
510(k) submission for ACCU-PREP™
JUL 1 1997
K970484
30
# 510(k) SUMMARY
As Required by the Safe Medical Devices Act of 1990
# IDENTIFICATION OF THE LEGALLY MARKETED PREDICATE DEVICE
## PREDICATE DEVICE
### MICROETCHER™
MICROETCHER is a compressed air driven dental sandblasting device intended for intra-oral use to roughen dental hard tissues or restorative materials to enhance bonding of resin materials. The device is a hand held instrument resembling a dental handpiece in shape and function. The unit is operated at 60-100 psig of compressed air. A venturi principle forces an abrasive, aluminum oxide (50μm particle size), toward the object to be abraded. The high velocity abrasive imparts sufficient energy to abrade and remove material.
The device can be used extra-orally for similar purposes, that is, to roughen dental restorations to enhance bonding or to behave as a cleaning device.
## DESCRIPTION OF APPLICANT DEVICE
### ACCU-PREP™
ACCU-PREP is virtually identical to the MICROETCHER in design and function. The two models of the device, Type I and Type II are identical in design and function but differ cosmetically. That is Type I is nearly identical to the MICROETCHER and Type II is esthetically more pleasing. In addition, Type II has the abrasive reservoir in the center of the handpiece as opposed to the back end as in the MICROETCHER and it is completely autoclavable. The ACCU-PREP operates at somewhat lower minimum pressure than the MICROETCHER, 40 psig as opposed to 60.
## INTENDED USES OF APPLICANT DEVICE
ACCU-PREP is intended to be used to abrade and enhance the bond between the prepared tooth and cast dental metals for Maryland bridges, for cementation of cast metal restorations, for porcelain repair, as a porcelain etch substitute, to prepare indirect composite restorations for cementation, for repair of composite or resin restorations, repair of acrylic facings on cast restorations, endodontic post preparation, denture repairs, and orthodontic applications such as: preparation of brackets to enhance bonding, cement removal from brackets, and enhancement of adhesion to tooth structure.
{1}
BISCO, INC., 1500 W. Thorndale Ave., Itasca, IL 60143
510(k) submission for ACCU-PREP™
31
# SCIENTIFIC CONCEPTS and SIGNIFICANT PERFORMANCE CHARACTERISTICS
The scientific concepts of this device are basic. A relatively high velocity stream of compressed gas (air), when blown across the opening of a tube set at 90 degrees to the air stream creates a low pressure at the opposite end of the perpendicular tube. This low pressure (vacuum) sucks up an abrasive powder from a reservoir and propels the abrasive thereby abrading the surface of an object in its path. The phenomenon is analogous to a perfume bottle. The abrasive powder behaves as the perfume, that is, as air is blown across the opening the abrasive powder is blown into the air stream.
Comparison of ACCU-PREP to the predicate device shows both are very similar with the major difference being esthetics. The proposed new products, Types I and II, are functionally identical to the predicate device but are more esthetically acceptable top the patient as well as to the dentistry and dental auxiliary. One notable difference between the tow is the applicant device operates at lower air pressure.
Both employ 50 µm aluminum oxide as the primary abrasive but other grits can be used where appropriate. The range of normal grit sizes is 0.05 to 90 µm
Since both devices are appear essentially identical their performances were compared by testing in actual use situations. Shear bond tests were performed on human extracted teeth using identical conditions with the only difference being the device abrading the substrate. No statistically significant difference was found in this test at p ≥ 0.05. Examination of abraded surfaces using Scanning Electron Microscopy showed virtually identical results.
James L. Sandrik, Ph.D.
BISCO, Inc.
Itasca, IL 60143
February 7, 1997
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Dr. James L. Sandrik, Ph.D.
Director of Technical Affairs
Bisco, Incorporated
1500 West Thorndale Avenue
Itasca, Illinois 60143
JUL 1 1997
Re: K970484
Trade Name: Accu-Prep
Regulatory Class: III
Product Code: KOJ
Dated: February 7, 1997
Received: February 10, 1997
Dear Dr. Sandrik:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices. General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}
Page 2 - Dr. Sandrik
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Timothy A. Ulatowski
Director
Division of Dental, Infection Control and General Hospital Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{4}
BISCO, INC., 1500 W. Thorndale Ave., Itasca, IL 60143
510(k) submission for ACCU-PREP™
6
# INDICATIONS for USE
510(k) Number (if known): --
Device Name: **ACCU-PREP**
## Indications for Use:
1. **Maryland Bridges (resin bonded bridges)**: for preparation of cast metal framework of both non-precious (Ni-Cr) and noble (Au, Pt, Pd) alloys.
2. **Cast restoration cementation**: for preparation of intaglio surface of cast metal restorations to enhance micromechanical bonding especially of traditional cements (zinc phosphate).
3. **Porcelain repair**: for preparation of exposed metal (and porcelain) to repair failed porcelain fused to metal restoration intra-orally.
4. **Porcelain etch substitute**: for preparation of intaglio surface of porcelain restorations in place of etching with hydrofluoric acid in the mouth.
5. **Composite surface preparation substitute**: for preparation of intaglio surface of composite or polymeric restoration in place of surface active agents (i.e. methyl methacrylate).
6. **Composite/polymeric restoration repair**: for preparation of exposed failed restoration as in 5 above.
7. **Repair of acrylic facings on cast restorations**: for preparation of cast metal surface to enhance bond of self-cure acrylic facing to restoration.
8. **Endodontic post preparation**: for preparation of endo post to enhance micromechanical bond of post to prepared tooth.
9. **Denture repairs**: for preparation of denture base to facilitate repairs.
10. **Orthodontic applications**: for preparation of ortho brackets to enhance bonding to the bracket. For cement removal from brackets before re-bonding. For enhancement of adhesive to tooth structure.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRL, Office of Device Evaluation (ODE)

Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐ (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.