K970478 · Marquest Medical Products, Inc. · BTT · Jul 16, 1997 · Anesthesiology
Device Facts
Record ID
K970478
Device Name
SCT 3000 HEATED HUMIDIFIER
Applicant
Marquest Medical Products, Inc.
Product Code
BTT · Anesthesiology
Decision Date
Jul 16, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5450
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The SCT 3000 Heated Humidifier is indicated for use with breathing circuits when healthcare professionals require humidified and/or heated gas being delivered to patients.
Device Story
SCT 3000 Heated Humidifier provides heated, humidified gas to patients with compromised upper airway humidification. Device uses breathing circuits and humidification chambers to deliver gas; prevents airway/lung damage from dry gas; maintains patient body temperature. Operated by healthcare professionals in hospitals, ICU, anesthesia, or home settings. System employs feedback loop using two thermistors for analog temperature input; Intel P80C32 microprocessor regulates heating elements via servo-controlled mechanism. Microprocessor manages heating element duty cycles (on/off) based on thermistor feedback. Device features diagnostic routines, safety alarms (high/low temperature), and visual displays. Output is controlled gas temperature and humidity. Benefits include stable patient temperature and protection of respiratory mucosa.
Clinical Evidence
Bench testing only. Device performance validated against safety standards including UL 544 (Medical and Dental Equipment), CSA C22.2 No. 125-M1984 (Electromedical Equipment), and ANSI Z79.9-1979 (Humidifiers and Nebulizers).
Technological Characteristics
Microprocessor-controlled (Intel P80C32) heated humidifier. Uses pass-over wick humidification. Feedback loop utilizes two thermistors for analog temperature sensing to control heating elements. Supports heated wire and no-wire breathing circuits. Safety features include high/low temperature alarms and diagnostic routines. Meets UL 544, CSA C22.2 No. 125-M1984, and ANSI Z79.9-1979 standards. Reusable device, supplied clean non-sterile.
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring heated and/or humidified gas delivery via breathing circuits during anesthesia or respiratory therapy. Contraindications not specified.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}
K970478
# Non-Confidential Summary of Safety and Effectiveness
## February 5, 1997
JUL 16 1997
Page 1 of 5
Marquest Medical Products, Inc.
11039 East Lansing Circle
Englewood, CO 80112
Tel - (303) 790-4835
Fax - (303) 799-0210
**Official Contact:**
Thomas W. Dielmann,
Vice President, Regulatory Affairs / Quality Assurance
**Proprietary or Trade Name:**
SCT 3000 Heated Humidifier
**Common / Usual Name:**
Heated Humidifier
**Classification Name:**
Respiratory Gas Humidifier
Subsection 868.5450
**Intended Device:**
SCT 3000 uses breathing circuits and humidification chambers to precisely control the delivery of gas temperature and humidity to the patient.
**Predicate Devices:**
Marquest SCT 3000 - K903138
Fisher & Paykel 730 Respiratory Humidifier - K913368
**Device Description / Indicated Use:**
Servo-Controlled Tracking SCT 3000 heated humidifier is designed to provide heated, humidified gas to a patient whose own humidification system - the nose and upper airway - is compromised. The unit is used primarily during anesthesia or prolonged respiratory therapy to prevent damage to the upper airway and lungs caused by dry gas while helping to maintain the patient's body temperature stable.
page 4
{1}
page 5
# Non-Confidential Summary of Safety and Effectiveness (continued)
## February 5, 1997
page 2 of 5
### Device Description / Indicated Use (continued):
The whole system, including disposable humidification chambers, heated wire breathing circuits and no-wire (vent) circuits, achieves precise control of delivered gas temperature and humidity by a servo controlled mechanism.
The SCT 3000 employs a feedback loop where the temperature input from the thermistors controls the heating elements. The SCT 3000 has two thermistors that provide analog input. This information, in turn, is used to control the amount of time that the two heating elements are cycled on/off. The SCT's microprocessor - the Intel P80C32, regulates this servo mechanism. the advanced design allows simple operation, sophisticated diagnostic routines and superior safety features to be built into the instrument, resulting in better performance and safer operation.
### Targeted Population:
Adult, neonate, and pediatric
### Environment of Use:
Hospital, anesthesia, ICU, respiratory therapy, home healthcare
{2}
page 3 of 5
# Non-Confidential Summary of Safety and Effectiveness (continued)
## February 5, 1997
### Comparison to Predicate Devices:
| Attribute | Marquest SCT 3000 | Fisher & Paykel MR 730 |
| --- | --- | --- |
| Use | | |
| Indicated for anesthesia | Yes | Yes |
| Indicated for respiratory therapy | Yes | Yes |
| Population - adult, pediatric, neonate | Yes | Yes |
| Indicated for use only with specific manufacturers heated circuits | Yes | Yes |
| Use with heated wire circuits | Yes | Yes |
| Use with no-wire circuits | Yes | Yes |
| Single patient use applications | Yes | Yes |
| Reusable circuits application | Yes | Yes |
| Heated wire operation | Yes | Yes |
| No-wire operation | Yes | Yes |
| For use with low flow chambers | Yes | Yes |
| For use with high flow chambers | Yes | Yes |
{3}
page 4 of 5
# Non-Confidential Summary of Safety and Effectiveness (continued)
## February 5, 1997
### Comparison to Predicate Devices:
| Attribute | Marquest SCT 3000 | Fisher & Paykel MR 730 |
| --- | --- | --- |
| **Use (continued)**
For use with ventilators - with or without exhalation valves | Yes | Yes |
| For use with anesthesia systems | Yes | Yes |
| For use with various heated wire circuit configurations | Yes | Yes |
| For use with various no-wire circuit configurations | Yes | Yes |
| **Design**
Microprocessor controlled | Yes | Yes |
| High Temperature Alarm | Yes - Distal Temperature > 1 degree C of Set Temp. | Yes - Distal Temperature > 2 degrees C of Set Temp. |
| Low Temperature Alarm | Yes - Distal Temperature < 4 degrees C of Set Temp. | Yes - Distal Temperature < 4 degrees C of Set Temp. |
{4}
page 5 of 5
Non-Confidential Summary of Safety and Effectiveness (continued)
February 5, 1997
| Attribute | Marquest SCT 3000 | Fisher & Paykel MR 730 |
| --- | --- | --- |
| Design (continued)
Maximum Heater Plate Temperature | 98 degrees C | 110 degrees C |
| Humidification Method | Pass Over Wick | Pass Over Wick |
| Alarm Mute | 1 Minute | 3 Minutes |
| Other Features | Pause Mode | Standby Mode |
| Safety Alarms | Yes - Visual Display of All Alarms | Yes - Visual Display or Error Code Display |
| Performance Testing
Meets UL 544, Medical and Dental Equipment / Patient Care Equipment | Yes | Yes |
| Meets CSA C22.2 No. 125-M1984, Electromedical Equipment | Yes | Yes |
| Meets ANSI Z79.9-1979, Humidifiers and Nebulizers for Medical Use | Yes | Yes |
## Differences
Any differences that do exist would not have an effect on the safety or effectiveness of the device.
page 8
{5}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Mr. Thomas W. Dielmann
Marquest Medical Products, Inc.
11039 East Lansing Circle
Englewood, Colorado 80112
JUL 16 1997
Re: K970478
SCT 3000 Heated Humidifier
Regulatory Class: II (two)
Product Code: 73 BTT
Dated: May 28, 1997
Received: June 2, 1997
Dear Mr. Dielmann:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}
Page 2 - Mr. Thomas W. Dielmann
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{7}
The New Marq of Excellence
Marquest Medical Products, Inc.
11039 East Lansing Circle
Englewood, Colorado 80112 U.S.A.
303 790 4835 phone
303 799 0210 fax
Section # 3
Labeling (continued)
## C. Indications for Use Statement
Pursuant to the Notice of 2/6/96 regarding listing of Indications for Use on a separate sheet, the following is per that request.
510(k) Number K970478 (to be assigned)
Device Name: SCT 3000 Heated Humidifier
Indications for Use: The SCT 3000 Heated Humidifier is indicated for use with breathing circuits when healthcare professionals require humidified and/or heated gas being delivered to patients.
Targeted Population: Adult, neonate, and pediatric
Environment of Use: Hospital, anesthesia, ICU, respiratory therapy, home healthcare
Disposable / Reusable: Reusable - Supplied Clean Non-Sterile
Concurrence of CDRH, Office of Device Evaluation (ODE)
Thomas J. Callahan
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K970478
Prescription Use ☑ (per 21 CFR 801.109)
or
Over - The - Counter Use
page 31
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.