← Product Code [EFQ](/productcode/EFQ) · K970455

# DUKAL NON ADHERENT DRESSING (K970455)

_Dukal Corp. · EFQ · May 8, 1997 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K970455

## Device Facts

- **Applicant:** Dukal Corp.
- **Product Code:** [EFQ](/productcode/EFQ.md)
- **Decision Date:** May 8, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

Absorbent non adherent dressing is used as a topical dressing to cover a wound or operative site while providing absorbency.

## Device Story

Dukal Non Adherent Pad; topical wound dressing; provides wound coverage and absorbency; used in clinical or home settings; applied by healthcare providers or patients; physical barrier function; prevents adherence to wound bed.

## Clinical Evidence

Bench testing only; physical property comparison provided.

## Technological Characteristics

Non-adherent pad; thickness 47 mil; Delnet fracture 5lb/in; T-Peel 700 lb/in; absorptive capacity 38 oz/yd²; MD tensile 6lb/in; basis weight 5.0 oz/yd².

## Predicate Devices

- Kendell Non Adherent Dressing

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

4F-1997 15:26 RMS 813 942 3828 P.04
K970455

# 510(K) SUMMARY
(as required by 807.92(c))

MAY - 8 1997

Submitter of 510(k):
Regulatory &amp; Marketing Services, Inc. (RMS)
40178 U.S. 19 North
Tarpon Springs, FL 34689
Phone: 813-942-3908
Fax: 813-942-3828

Contact Person: Ed Ransom

Date of Summary: February 3, 1997

Trade Name: Dukal Non Adherent Pad

Classification Name: Dressing

Predicate Device: Kendell Non Adherent Dressing

Device Description/
Comparison:

|  Material | Dukal | Kendall  |
| --- | --- | --- |
|  Thickness | 47 mil | same  |
|  Delnet Fracture | 5lb/in | same  |
|  T-Peel | 700 lb/in | same  |
|  Absorptive Capacity | 38 oz/yd² | same  |
|  MD Tensile | 6lb/in | same  |
|  Basis weight | 5.0 oz/yd² | same  |
|  Intended Use | Absorbent non adherent dressing is used as a topical dressing to cover a wound or operative site while providing absorbency. | same  |

Intended Use: Absorbent non adherent dressing is used as a topical dressing to cover a wound or operative site while providing absorbency.

---

**Source:** [https://fda-staging.innolitics.com/device/K970455](https://fda-staging.innolitics.com/device/K970455)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com