The Vitalmax 4000 is intended for use in the hospital/clinical environment to measure and monitor the following parameters: ECG waveform NIBP systolic, diastolic, and mean arterial pressure values (MEAN or MAP) Blood oxygen saturation (SpO2 or Pulse oximetry) SpO2 waveform Pulse (SpO2 and NIBP) signal strength Pulse (SpO2 and NIBP) or heart (ECG) rate Temperature End-tidal Co2 concentration (Et Co2) Co2 waveform Minimum inspired Co2 concentration (inCo2) Respiration rate Additional options offered are: Built-in printer with two channels of annotated waveform or text form of vital signs Thoracic impedance respiration Invasive pressure waveforms: systolic, diastolic, and mean values (IP, IP2) Second temperature channel and Δ temp Analog output The Vitalmax 4000 is available in configured and modular form. The Vitalmax 4000 configured monitor is available in various models which offer the options of Co2, thoracic impedance respiration, printer, and invasive pressures. The vitalmax 4000 is not intended to be used as apnea monitor or during MRI (Magnetic Resonance Imaging).
Device Story
Multi-parameter patient monitors (Vitalmax/Minipack series) for hospital/clinical use; measure ECG, NIBP, SpO2, temperature, CO2, respiration, invasive pressure, and anesthetic gases. Inputs: physiological signals via sensors/electrodes. Processing: signal conditioning and parameter calculation; output: real-time waveforms, numeric vital signs, and trends on integrated display or printer. Operated by clinicians; provides continuous monitoring to assist in patient status assessment and clinical decision-making. Benefits: real-time physiological data for patient safety.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Multi-parameter patient monitoring system; includes ECG, NIBP, SpO2, CO2, invasive pressure, and gas analysis modules. Features integrated display, optional thermal printer, and analog output. Connectivity includes standard physiological sensor interfaces. Software-based parameter calculation and waveform processing.
Indications for Use
Indicated for hospital/clinical environment patient monitoring of ECG, NIBP, SpO2, pulse, temperature, CO2, respiration, invasive pressure, and anesthetic gas concentrations. Contraindicated for use as an apnea monitor or during MRI procedures.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Submission Summary (Full Text)
{0}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Mr. Iihan M. Bilgutay
President
Pace Tech Medical Monitors
510 Garden Avenue North
Clearwater, FL 34615-4126
Re: K970445
Vitalmax 4000 Configure/Vitalmax 4000 Modular/Vitalmax 4100
Regulatory Class: II (Two)
Product Code: 74 MHX
Dated: November 5, 1997
Received: November 10, 1997
Dear Mr. Bilgutay:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{1}
Page 2 - Mr. Iihan M. Bilgutay
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}
# INDICATION FOR USE STATEMENT
In accordance with the FDA new requirement, January 1, 1996 the "Indication For Use", attached, please find the indication for use for all of the devices included under the Premarket notification #K970445. These devices are: Vitalmax 4000 Configured / Vitalmax 4000 Modular / Vitalmax 4100 / Minipack 3000 / Minipack 3100 / Minipack 300 / Vitalmax 800 PLUS.
## Vitalmax 4000 Configured / Modular
The Vitalmax 4000 is intended for use in the hospital/clinical environment to measure and monitor the following parameters:
- ECG waveform
- NIBP systolic, diastolic, and mean arterial pressure values (MEAN or MAP)
- Blood oxygen saturation (SpO2 or Pulse oximetry)
- SpO2 waveform
- Pulse (SpO2 and NIBP) signal strength
- Pulse (SpO2 and NIBP) or heart (ECG) rate
- Temperature
- End-tidal Co2 concentration (Et Co2)
- Co2 waveform
- Minimum inspired Co2 concentration (inCo2)
- Respiration rate
## Additional options offered are:
- Built-in printer with two channels of annotated waveform or text form of vital signs
- Thoracic impedance respiration
- Invasive pressure waveforms: systolic, diastolic, and mean values (IP, IP2)
- Second temperature channel and $\Delta$ temp
- Analog output
The Vitalmax 4000 is available in configured and modular form. The Vitalmax 4000 configured monitor is available in various models which offer the options of Co2, thoracic impedance respiration, printer, and invasive pressures.
The vitalmax 4000 is not intended to be used as apnea monitor or during MRI (Magnetic Resonance Imaging).
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K970445
{3}
# Vitalmax 4100
The Vitalmax 4100 CN is intended for use in the hospital/clinical environment to measure and monitor the following parameters:
- ECG waveform
- NIBP systolic, diastolic, and mean arterial pressure values (MEAN or MAP)
- Blood oxygen saturation (SpO2 or Pulse oximetry)
- SpO2 waveform
- Pulse (SpO2 and NIBP) signal strength
- Pulse (SpO2 and NIBP) or heart (ECG) rate
- Temperature
- End-tidal Co2 concentration (Et Co2)
- Co2 waveform
- Minimum inspired Co2 concentration (inCo2)
- Respiration rate
- Nitrous oxide concentration (N2O)
The Vitalmax 4100 G and Vitalmax 4100 A include these additional parameters:
- End-tidal expired Halothane, Enflurane, or Isoflurane agent gas concentration (etHAL, etENF, etISO)
- Inspired Halothane, Enflurane, or Isoflurane agent gas concentration (inHAL, inENF, inISO)
- Fractional inspired oxygen in gas concentration (Fio2 or %o2)
Additional options offered for both models are:
- Thoracic impedance respiration
- Invasive pressure waveform, systolic, diastolic and mean values (IP1, IP2)
- Second temperature channel and temperature difference [Δ temp]
- Built-in strip chart recorder
- Sevoflurane, and Desflurane (etSev, inSev, etDes, inDes)
- Analog Output
The Vitalmax 4100 is not intended to be used as apnea monitor or during MRI (Magnetic Resonance Imaging).
# Minipack 3000/3100
The Minipack 3000 and Minipack 3100 are intended for use in the hospital/clinical environment to measure and monitor the following parameters:
- ECG waveform
- NIBP systolic, diastolic, and mean arterial pressure values (MEAN or MAP)
- Blood oxygen saturation (SpO2 or Pulse oximetry)
- SpO2 waveform
- Pulse (SpO2 and NIBP) signal strength
{4}
- Pulse (SpO2 and NIBP) or heart (ECG) rate
- Temperature
- End-tidal Co2 concentration (Et Co2)
- Co2 waveform
- Minimum inspired Co2 concentration (inCo2)
- Respiration rate
Additional options offered are:
- Built-in printer with two channels of annotated waveform or text form of vital signs
- Thoracic impedance respiration
- Invasive pressure waveforms: systolic, diastolic, and mean values (IP1, IP2)
- Second temperature channel and Δ temp
- Analog output
The vitalmax 3000 and Vitalmax 3100 are not intended to be used as apnea monitors or during MRI (Magnetic Resonance Imaging).
## Minipack 300/ Vitalmax 800 PLUS
The Minipack 300 and Vitalmax 800 PLUS are intended for use in the hospital/clinical environment to measure and monitor the following parameters:
- ECG waveform
- NIBP systolic, diastolic, and mean arterial pressure values (MEAN or MAP)
- Blood oxygen saturation (SpO2 or Pulse oximetry)
- SpO2 waveform
- Pulse (SpO2 or NIBP) signal strength
- Pulse (SpO2 and NIBP) or heart (ECG) rate
- Temperature
- End-tidal Co2 concentration (Et Co2)
- Co2 waveform
- Minimum inspired Co2 concentration (inCo2)
- Respiration rate
Additional options offered are:
- Thoracic impedance respiration
- Add-on 27 ccolumn thermal printer (Minipack 300 Series)
- Built-in printer with two channels of annotated waveform (Vitalmax 800 PLUS)
- Analog output
The Minipack 300 and Vitalmax 800 PLUS are not intended to be used as apnea monitors or during MRI (Magnetic Resonance Imaging).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.