K970380 · Pyng Medical Corp. · FMI · Apr 25, 1997 · General Hospital
Device Facts
Record ID
K970380
Device Name
F.A.S.T. 1 INTRAOSSEOUS INFUSION SYSTEM
Applicant
Pyng Medical Corp.
Product Code
FMI · General Hospital
Decision Date
Apr 25, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The F.A.S.T. 1™ Intraosseous Infusion System is intended for intraosseous infusion as an alternative to intravenous access to facilitate emergency resuscitation through the use of drugs and fluids.
Device Story
Hand-held, manually operated passive device for emergency intraosseous infusion in adults. Input: manual force applied by user to introducer handle. Mechanism: stainless steel needle tip penetrates skin and cortical bone of sternum; internal depth-control mechanism detects cortical bone surface to disconnect handle at correct penetration depth. Output: infusion tube placed in marrow space for fluid/drug delivery. Used in emergency settings by paramedics/clinicians. Includes adhesive target/strain-relief patch for landmarking and site protection, and protector dome. Benefits: facilitates rapid vascular access by relatively untrained users; minimizes leakage; provides strain relief for infusion line; ensures correct depth placement.
Clinical Evidence
No clinical data on patients provided. Evidence consists of non-clinical engineering tests (mass, force, strength), biocompatibility testing (ISO 10993-1), sterilization validation (ANSI/ISO 11137), and mechanical performance testing (ISO 594). Performance evaluated via simulated clinical trials using mechanical simulators (simulated sternum/infusion system) and human cadaver/excised sternum studies conducted by civilian and military paramedics.
Technological Characteristics
Passive, manually operated infusion system. Materials: stainless steel, medical grade polyurethane, PVC, polyetherimide (radiation sterilizable), and medical-grade adhesives. Features: depth-control mechanism, target/strain-relief patch, protector dome. Connectivity: none. Sterilization: radiation. Standards: ISO 10993-1 (biocompatibility), ANSI/ISO 11137 (sterilization), ISO 594 (Luer connectors).
Indications for Use
Indicated for emergency resuscitation in adults requiring intraosseous infusion of drugs and fluids as an alternative to intravenous access.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}
PYNG MEDICAL CORP.
COPY # 4/12
K970380
# 11.0 510(K) SUMMARY
## 510(K) SUMMARY
APR 25 1997
This summary of safety and effectiveness was prepared on January 31, 1997 and is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92
### Submitted by:
Pyng Medical Corp.
2660 Oak Street,
Vancouver, B.C. V6H 3Z6
Canada.
Phone: (604) 875 4526 Fax: (604) 875 5390
Contact person: David L. Johnson, Ph.D.,
Vice-President, Research and Development
### Names of Device:
Trade Name: F.A.S.T. 1™ Intraosseous Infusion System
Common Name: Intraosseous Infusion System (Hypodermic Needle)
Classification Name: Hypodermic Single Lumen Needle (880.5570)
Class: II
Panel: 80
Procode: FMI - Hypodermic Single Lumen Needle
### Predicate Devices:
The legally marketed devices (predicate devices) upon which substantial equivalence to the F.A.S.T. 1™ Intraosseous Infusion System is based are:
1. "COOK Intraosseous Infusion Needles". The Cook, Inc. SUR-FAST® Intraosseous Infusion Needle Set and Components, and the Cook Disposable Intraosseous Infusion Needles. The 510(k) numbers for these products are K913258 and K915409.
2. "SHERWOOD ILLINOIS needle". Sherwood Medical Co. Monoject® 16 gage Illinois Sternal / Iliac Aspiration Needle, reorder number 8881-245164. The 510(k) number for this product is K883014.
97-01-31
PYNG MEDICAL CORP.
Page 11 - 1
File: SEC11.doc
{1}
PYNG MEDICAL CORP.
COPY # 4712
# Description of Device:
The F.A.S.T. 1™ Intraosseous Infusion System has been designed to provide a means, as an alternative to intravenous infusion, for facilitating emergency resuscitation of adults through the administration of fluids and drugs. The F.A.S.T. 1™ Intraosseous Infusion System has been designed for use on the adult sternum. It consists of a primary component - the infusion tube, and secondary components - the introducer for placing it through the skin into the sternum of the patient, the target / strain-relief patch, and a protector dome. There is also an infusion tube removal tool.
# Function, concepts, performance characteristics:
The concept of the device is that it will allow relatively untrained and unpractised users to safely and effectively insert a flexible infusion needle into the adult sternum.
The F.A.S.T. 1™ Intraosseous Infusion System is a hand-held, manually operated passive device, (it has no internal energy source). The infusion needle has a sharp stainless steel tip, a flexible shaft and a standard Luer connector. The user grasps an introducer handle attached to the infusion needle, and pushes the needle tip through the skin and overlying tissue into the sternum of the subject. Fluids or drugs are delivered through the needle to the marrow space of the sternum, and flow out the emissary veins of the sternum into the central circulation.
When the user has pushed the infusion needle to the correct penetration depth in the sternum, a mechanism inside the introducer handle disconnects the handle from the infusion needle, preventing the user from pushing the needle any further. The mechanism is fast acting - the user can push the needle in, have it released at the correct depth, and remove the introducer handle, all in the space of 1-2 seconds. The principle of the mechanism is that it detects the position of the anterior surface of the cortical bone of the sternum, and relates the position of the tip of the infusion needle to the surface of the cortical bone.
Another feature of the device is an adhesive patch which is applied to the top of the patient's chest. During application the user aligns the patch with the supra-sternal notch. The position of the recommended insertion site is marked on the patch to aid in landmarking. Also, the patch and an associated cover provide protection and strain-relief to the site and the installed infusion needle.
The materials used in the F.A.S.T. 1™ Intraosseous Infusion System are materials commonly used in other infusion devices; stainless steels, medical
97-01-31
PYNG MEDICAL CORP.
Page 11 - 2
File: SEC11.doc
{2}
PYNG MEDICAL CORP.
COPY # 4/12
grade polyurethane and PVC, radiation sterilizable medical grade plastic (polyetherimide), and patch and adhesive materials commonly used in mounting transdermal drug delivery devices to human skin for extended periods.
## Intended Use of the Device:
The F.A.S.T. 1™ Intraosseous Infusion System is intended for intraosseous infusion as an alternative to intravenous access to facilitate emergency resuscitation through the use of drugs and fluids.
## Technological Characteristics compared with Predicate Devices:
The F.A.S.T. 1™ Intraosseous Infusion System and the COOK needles and the SHERWOOD needle have common technological characteristics in that they:
- are passive, hollow entry ports for the delivery of fluids to a marrow space;
- have a stainless steel tip which penetrates through skin and overlying tissue into a marrow space;
- are manually inserted using forces applied by the practitioner to a plastic handle, at least part of which is removable;
- penetrate the cortical bone by the user pushing on the stainless steel tip, either without twisting (Pyng) or with twisting (Cook, Sherwood);
- have a lumen which may be blocked with a removable stylet during insertion;
- have a lumen which conveys drugs or fluids to a tip positioned in a marrow space;
- include a distal connector which is a Luer taper connector and which connects to the source of drugs or fluids;
- are supplied as biocompatible, sterile, single-use products;
- insert a needle or portal which is removed by attaching a remover and/or twisting and pulling out.
Each of the F.A.S.T. 1™ Intraosseous Infusion System, the COOK needles and the SHERWOOD needle consists of an apparatus and method for placing, into the marrow space of a bone, a hollow tube with a Luer connector on the distal end, for the purpose of delivering drugs and fluids. Each is manually operated. Each is made of common biocompatible steels and plastics.
Not all the technological characteristics of the F.A.S.T. 1™ Intraosseous Infusion System are the same as the predicate devices. The new technological characteristics of the F.A.S.T. 1™ Intraosseous Infusion System all arise from
97-01-31
PYNG MEDICAL CORP.
Page 11 - 3
File: SEC11.doc
{3}
PYNG MEDICAL CORP.
COPY # 4/12
the fundamental design objective of producing a system which is fully compatible with the environment in which it will be used. The F.A.S.T. 1™ Intraosseous Infusion System has several new technological characteristics.
## Tests conducted:
Data have been obtained from tests conducted by or for Pyng Medical Corp., and from published literature. These tests relate to the safety and effectiveness of the new technological characteristics, and of technological characteristics shared with the predicate devices. The data include material on biocompatibility, material relating to technological characteristics shared with the predicate devices, and material relating to new technological characteristics.
These data support the assertion that the F.A.S.T. 1™ Intraosseous Infusion System is as safe and as effective as the predicate devices.
## Non-clinical data, simulated clinical trials, simulations and simulators - general description:
The data include the results of common non-clinical engineering tests — masses, forces, strengths, etc. In addition, much data was acquired during testing with excised human sternums, with human cadavers, and with live human subjects in a variety of simulations and simulated clinical trials.
In order to appropriately evaluate and develop the F.A.S.T. 1™ Intraosseous Infusion System, Pyng developed two different mechanical devices as simulators - a simulated Intraosseous Infusion System, and a simulated sternum. These devices allowed the developers, and potential users, to experience the use of the system on live human volunteers without injury to the volunteers, and to experience insertion and placement of infusion portals at the correct depth through overlying layers of tissue and skin. Using these devices, simulated clinical trials were conducted by civilian and military paramedics.
Further data on the performance of the complete F.A.S.T. 1™ Intraosseous Infusion System were gathered from testing with human cadavers and excised sternums.
## How the test results support substantial equivalence:
The F.A.S.T. 1™ Intraosseous Infusion System and predicate devices meet the relevant voluntary and mandatory standards including those for biocompatibility
97-01-31
PYNG MEDICAL CORP.
Page 11 - 4
File: SEC11.doc
{4}
PYNG MEDICAL CORP.
COPY # 4/12
(ISO 10993-1 as modified in FDA General Program memorandum #G95-1), sterilization (ANSI/ISO 11137 (R2-94)), and mechanical performance of Luer connectors (ISO 594).
The F.A.S.T. 1™ Intraosseous Infusion System has passed evaluation of all the internal specifications developed by Pyng Medical Corp. to ensure that the infusion system meets or exceeds the performance of the predicate devices, and meets or exceeds the safety and effectiveness requirements for a successful adult emergency intraosseous infusion system. The F.A.S.T. 1™ Intraosseous Infusion System has been evaluated by both civilian and military paramedics, and given highly positive ratings. To achieve this level of performance certain technological features not common to the predicate devices had to be developed and incorporated.
**Safety.** The quick push-pull insertion method, the protector dome, and the strain-relief and target functions of the target patch ensure that safe use of the system requires only the skills possessed by the intended user. The protector dome keeps the infusion site clean, and allows the user to see the site during infusion. The depth control mechanism, the bone probe and the target function of the patch avoid excess penetration. The flexible needle and subcutaneous portal, and the strain relief function of the patch minimize leakage. Data included in the submission support these assertions and the conclusion that the F.A.S.T. 1™ Intraosseous Infusion System is as safe as the predicate devices.
**Effectiveness.** The target function of the patch facilitates correct placement at the recommended insertion site. The depth control mechanism and the bone probe ensure that the system penetrates into the marrow space. These features together with the push-pull insertion method ensure that the system meets user criteria for vascular access time and success rate. The strain-relief function of the patch resists pull-out resulting from tension on the infusion line. The flexible needle and subcutaneous portal, and the protector dome, ensure that the system will be unaffected by large external forces and be compatible with other procedures performed on the patient. Data included in the submission support these assertions and the conclusion that the F.A.S.T. 1™ Intraosseous Infusion System is as effective as the predicate devices.
## Conclusions:
The F.A.S.T. 1™ Intraosseous Infusion System and the predicate devices are all indicated for use in intraosseous infusion.
The F.A.S.T. 1™ Intraosseous Infusion System and one or both of its predicate devices are similar in that they are passive, needle-like devices intended for
97-01-31
PYNG MEDICAL CORP.
Page 11 - 5
File: SEC11.doc
{5}
PYNG MEDICAL CORP.
COPY # 4/12
emergency intraosseous infusion. They have a handle, at least part of which is removable, which the care-giver uses to place a steel tip through overlying tissue into the marrow space of a bone, using pushing and/or twisting motions. They have a lumen with a distal Luer taper connector which connects to a source of drugs or fluids to be infused. They are removed by attaching a remover and/or twisting and pulling out. They are supplied as biocompatible, sterile, single use products.
The above data and conclusions demonstrate that the F.A.S.T. 1™ Intraosseous Infusion System is as safe and effective as the predicate devices.
97-01-31
PYNG MEDICAL CORP.
Page 11 - 6
File: SEC11.doc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.