← Product Code [GCJ](/productcode/GCJ) · K970369

# VISTA MINI CAMERA SYSTEM (K970369)

_Oktas · GCJ · Apr 14, 1997 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K970369

## Device Facts

- **Applicant:** Oktas
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Apr 14, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The device is intended for use in endoscopic procedures and all types of video assisted procedures, including anterior spinal and general thoracic procedures and as an aid in visualization of the evacuated cardiac chamber.

## Device Story

Vista Mini Camera system; video camera for endoscopic procedures; inputs optical images from body cavities/organs via endoscope; transforms optical input to electronic video signal via camera control unit (CCU); outputs video to monitors or head-mounted displays; used in OR by surgeons/physicians; aids visualization during minimally invasive surgery; facilitates surgical navigation and procedure monitoring; benefits patient through improved visualization during endoscopic/video-assisted interventions.

## Clinical Evidence

Bench testing only. Compliance with IEC 601-1 (general safety), IEC 601-1-2 (EMC), IEC 601-2-18 (endoscopic safety), ISO 10993 (biocompatibility), and UL544 (medical equipment safety). Optical performance data provided.

## Technological Characteristics

Video endoscope system; consists of camera head and camera control unit (CCU); compatible with standard light sources and video monitors/head-mounted displays. Complies with IEC 601-1, IEC 601-1-2, IEC 601-2-18, ISO 10993, and UL544 standards.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- 3D Scope Three Dimensional Medical VideoSystem ([K923160](/device/K923160.md), [K961182](/device/K961182.md))
- Oktas Video Endoscope System with Zoom ([K946171](/device/K946171.md))
- Zeiss EndoLive™ Stereo Endoscope ([K950988](/device/K950988.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Vista Medical Technologies Westborough MA
APR 14 1997
510K Notification
Vista Mini Cameras
January 1997
K970369
510(K) SUMMARY
January 1997

## COMPANY NAME AND ADDRESS

Vista Medical Technologies
134 Flanders Road
Westborough, MA. 01581

## CONTACT PERSON

Martin Newman
Director of Regulatory Affairs and Quality Assurance
Telephone (508) 366-3668
Fax: (508) 366-1543

## DEVICE TRADE NAME

Vista Mini Camera(s)

## COMMON NAME

Video Endoscope System

## PREDICATE DEVICE

1. Device Name: 3D Scope Three Dimensional Medical VideoSystem
Classification: Camera, Television, Endoscopic, without audio
Manufacturer: American Surgical Technologies
300 Billerica Rd
Chelmsford, MA 01824
510(k) #: K923160, K961182

2. Device Name: Oktas Video Endoscope System with Zoom
Classification: Endoscopes and Accessories - 21 CFR 876.1500
Manufacturer: Oktas
134 Flanders Rd
Westborough, MA 01581
510(k) #: K946171

Page 1 of 2
510(k)

{1}

Vista Medical Technologies Westborough MA
510K Notification
Vista Mini Cameras January 1997

3. Device Name: Zeiss EndoLive™ Stereo Endoscope
Classification: Endoscopes and Accessories - 21 CFR 876.1500
Manufacturer: Carl Zeiss, Inc.
Thomwood, NY 10594
510(k) #: K950988

## DEVICE DESCRIPTION

The Vista Mini Camera System is a device used to allow observation in body cavities, organs, or canals through manmade or natural orificies. It is designed for use in endoscopic procedures and all types of video assisted procedures, including anterior spinal and general thoracic procedures and as an aid in visualization of the evacuated cardiac chamber. The product is a video camera. The system will be supplied as a Vista Mini Camera, intended to be used with a Vista camera control unit (CCU). The device is designed to work with commercially available light sources and video monitors or head mounted displays.

## INTENDED USE

The device is intended for use in endoscopic procedures and all types of video assisted procedures, including anterior spinal and general thoracic procedures and as an aid in visualization of the evacuated cardiac chamber.

## PERFORMANCE DATA

The Vista Mini Camera System was designed and tested in compliance with the requirements of the following standards:

|  IEC 601-1 | General Safety requirements for Medical Electronic Equipment  |
| --- | --- |
|  IEC 601-1-2 | Electromagnetic Compatibility Requirements and Tests  |
|  IEC 601-2-18 | Safety of Endoscopic Equipment  |
|  ISO 10993 | Biological Evaluation of Medical Devices  |
|  UL544 | Standard for Safety, Medical and Dental Equipment  |
|   | Optical Test Data  |

Page 2 of 2
510(k)

---

**Source:** [https://fda-staging.innolitics.com/device/K970369](https://fda-staging.innolitics.com/device/K970369)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
